Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 611

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 23
8-7-11 Record Software Versions  - Software Version v7.x/v6.x/v5.x/v4.x
1.) Select the About tab.
2.) Record software and hardware versions in  Table 8-8 on page 8-23.
 
Figure 8-20   Information available on the About tab
Table 8-8    Record Software versions  - Software Version v7.x/v6.x/v5.x/v4.x
DESCRIPTIONRECORD VERSIONS
APPLICATION SW
SYSTEM SW
PROCESSOR (BEP) MODEL 

Page 612

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 24 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12 Software Loading - Software  Version v7.x/v6.x/v5.x/v4.x
8-7-12-1 Overview
This software loading procedure covers: 
• Re-installation of BT’08 softwa re (System Software v3.0.x and Application Software v7.x) on a 
Vivid 7 unit with BEP4.3
• Re-installation of BT’08 softwa re (System Software v2.1.1 and Application Software v7.x) on a...

Page 613

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 25
8-7-12-3 Loading System Software  v3.0.x on Vivid 7 with BEP4.3
NOTE: This procedure describes how to reload software that already has been stored as compressed  files on the BEP4.3’s hard disk drive. A  CD is not used in this procedure.
When using this procedure, you can choose to reload  partition C: only, or to erase all partitions. The 
selection is done later in this procedure.
1.)...

Page 614

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 26 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-3      Loading System Software v3.0.x on Vivid 7 with BEP4.3 (cont’d) 4.) In this step, you can choose if you want to do a co mplete Vivid 7 software installation that overwrites 
all data, or if you only want to  update or reinstall the software on the C: partition (C: drive). 
Usually it should be OK to select “ B” to update the C:...

Page 615

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 27
8-7-12-4 Loading Application Software  v7.x on Vivid 7 with BEP4.3
NOTE: This procedure describes how to reload software that already has been stored as compressed  files on the BEP4.3’s hard disk drive. A  CD is not used in this procedure.
To reload application softwa re, follow this procedure:
1.) Reboot the Vivid 7, and press the  ESC-key each second. The boot se quence will be...

Page 616

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 28 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-4      Loading Application Soft ware v7.x on Vivid 7 with BEP4.3 (cont’d)
4.) Select “ C. Reload Application SW  
5.) Confirm the selection by pressing  Y.
The system reboots and the Application soft ware installation starts automatically.
When completed, the Vivid 7 reboots and starts  up normally. The User Presets, the System...

Page 617

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 29
8-7-12-5 Vivid 7 System Software Loading - App lication Software Version v7.x/v6.x/v5.x/v4.x
NOTE: This procedure covers installation of software from CDs. 1.) Insert the applicable  Vivid 7 System Software CD  into the CD/DVD drive.
2.) Power down the System, see  4-2-3 Power Shut Down on page 4-7 .
3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch...

Page 618

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 30 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-5              Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x  (cont’d)
5.) In this step, you can choose if you want to do a co mplete Vivid 7 installation or if you only want to 
update or reinstall the software on the C: partition (C: drive). 
Usually it should be OK to select  “B” to update the...

Page 619

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 31
8-7-12-5              Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x  (cont’d)
6.) The Software loader program starts and the below Time Bar Info Screen appears.
  
7.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the 
screen indicates process.
8.) You will now be prompted to  remove the first CD and...

Page 620

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 32 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-12-6 Application SW Loading - Soft ware Version v7.x/v6.x/v5.x/v4.x
1.) Insert the applicable CD labelled  Vivid 7 & EchoPAC PC Application SW .
2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the  scanner. (This is called Forced Shut Down).
3.) Wait until the LCD’s on the Oper ator Panel is...
Start reading GE Vivid 7 Service Manual

Related Manuals for GE Vivid 7 Service Manual

All GE manuals