Page 611
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 23 8-7-11 Record Software Versions - Software Version v7.x/v6.x/v5.x/v4.x 1.) Select the About tab. 2.) Record software and hardware versions in Table 8-8 on page 8-23. Figure 8-20 Information available on the About tab Table 8-8 Record Software versions - Software Version v7.x/v6.x/v5.x/v4.x DESCRIPTIONRECORD VERSIONS APPLICATION SW SYSTEM SW PROCESSOR (BEP) MODEL
Page 612
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 24 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12 Software Loading - Software Version v7.x/v6.x/v5.x/v4.x 8-7-12-1 Overview This software loading procedure covers: • Re-installation of BT’08 softwa re (System Software v3.0.x and Application Software v7.x) on a Vivid 7 unit with BEP4.3 • Re-installation of BT’08 softwa re (System Software v2.1.1 and Application Software v7.x) on a...
Page 613
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 25 8-7-12-3 Loading System Software v3.0.x on Vivid 7 with BEP4.3 NOTE: This procedure describes how to reload software that already has been stored as compressed files on the BEP4.3’s hard disk drive. A CD is not used in this procedure. When using this procedure, you can choose to reload partition C: only, or to erase all partitions. The selection is done later in this procedure. 1.)...
Page 614
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 26 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-3 Loading System Software v3.0.x on Vivid 7 with BEP4.3 (cont’d) 4.) In this step, you can choose if you want to do a co mplete Vivid 7 software installation that overwrites all data, or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “ B” to update the C:...
Page 615
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 27 8-7-12-4 Loading Application Software v7.x on Vivid 7 with BEP4.3 NOTE: This procedure describes how to reload software that already has been stored as compressed files on the BEP4.3’s hard disk drive. A CD is not used in this procedure. To reload application softwa re, follow this procedure: 1.) Reboot the Vivid 7, and press the ESC-key each second. The boot se quence will be...
Page 616
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 28 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-4 Loading Application Soft ware v7.x on Vivid 7 with BEP4.3 (cont’d) 4.) Select “ C. Reload Application SW 5.) Confirm the selection by pressing Y. The system reboots and the Application soft ware installation starts automatically. When completed, the Vivid 7 reboots and starts up normally. The User Presets, the System...
Page 617
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 29 8-7-12-5 Vivid 7 System Software Loading - App lication Software Version v7.x/v6.x/v5.x/v4.x NOTE: This procedure covers installation of software from CDs. 1.) Insert the applicable Vivid 7 System Software CD into the CD/DVD drive. 2.) Power down the System, see 4-2-3 Power Shut Down on page 4-7 . 3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch...
Page 618
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 30 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-5 Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x (cont’d) 5.) In this step, you can choose if you want to do a co mplete Vivid 7 installation or if you only want to update or reinstall the software on the C: partition (C: drive). Usually it should be OK to select “B” to update the...
Page 619
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 31 8-7-12-5 Vivid 7 System Software Loading - Application Software Version v7.x/v6.x/v5.x/v4.x (cont’d) 6.) The Software loader program starts and the below Time Bar Info Screen appears. 7.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 8.) You will now be prompted to remove the first CD and...
Page 620
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 32 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-12-6 Application SW Loading - Soft ware Version v7.x/v6.x/v5.x/v4.x 1.) Insert the applicable CD labelled Vivid 7 & EchoPAC PC Application SW . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until the LCD’s on the Oper ator Panel is...