Page 621
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 33 8-7-13 Verifications - Softwa re Version v7.x/v6.x/v5.x/v4.x NOTE: Software Option Strings are software versio n dependant. If a new software version has been installed, please follow installa tion instructions for that softwar e. Usually, new software option strings are needed. 8-7-13-1 Verify SW Option Key Settings 1.) Press CONFIG (F2) and log on as ADM, see 4-2-4 Log On to the...
Page 622
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 34 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-13-2 Verify Video and VCR Settings - Software Version v7.x/v6.x/v5.x/v4.x 1.) Select Service (lower, right corner) . 6.) From Video Settings pull down menu, verify Format as recorded in Table 8-7 on page 8-22 . Correct the selection if needed. 7.) From the VCR pull down menu, verify VCR type as recorded in Table 8-7 on page 8-22...
Page 623
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 35 8-7-13-3 Verify TCP/IP Settings - So ftware Version v7.x/v6.x/v5.x/v4.x 1.) From the top of the Connec tivity screen, select the Tcpip tab. 2.) Verify that all settings on the TCPIP tab matches those recorded in Table 8-6 Record settings from Tcpip screen - software version v7.x/v6.x/v5.x/v4.x on page 8-21 . 3.) Select Save settings . 4.) Select Ok to continue. 8-7-13-4 DC Offset...
Page 624
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 36 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-14 Restore Patient Archive/System Config. - Software v7.x/v6.x/v5.x/v4.x 1.) Press Config (F2) . 2.) Select the category Admin. 3.) Select the Restore tab. 4.) In the Restore tab select as needed: - Patient archive to restore the patient archive. - System configuration to restore all system settings and user presets. OR -...
Page 625
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 37 8-7-14 Restore Patient Archi ve/System Config. - Software v7.x/v6.x/v5.x/v4.x (cont’d) 8.) Select Restore Now . Depending on the selection, one or two Restore confirmation windows are displayed: 9.) Ensure that the correct so urce is selected an select OK. The selected items are copied to the systems. If items from the System Configuration are restored the system needs to be rebooted....
Page 626
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 38 Section 8-8 - Software Loading Procedure - Software Version v3.x Section 8-8 Software Loading Procedur e - Software Version v3.x 8-8-1 Introduction This procedure describes how to perform a complete installation of the software on the Vivid 7 hard disk. 8-8-2 Preparations Before Software Re-installation • Export Patient Images from the internal hard disk to removable removable media . • Make backup of Customer Presets,...
Page 627
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 39 8-8-3 Customer Provided Prerequisite • Formatted and labelled removable media for Images storage • Formatted and labelled MO Disk for Patient Archive and User Defined Settings. 8-8-4 Tools Provided With Un it or After an Upgrade 8-8-5 Preparations 1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see 4-2-2 Power ON/ Boot UP on page 4-3 . 2.) Store images to removable...
Page 628
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 40 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-6 Backup - Software Version v3.x The Backup function enables the user to: • Make backup of the patient arch ive (patient data and reports). • Move images from the local hard drive to MO disks. • Make backup of the system configuration. 8-8-6-1 Move Images From the Local Hard Drive to MO disks. The Move Images function is used to move archived images from...
Page 629
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 41 8-8-6-1 Move Images From the Local Hard Drive to MO disks. (cont’d) 7.) Press Backup Now.- The system is checking that the MO disk is inserted. If not, a dialogue window is displayed prompting the user to insert a MO disk. Select OK to resume backup. - The system is checking that the MO disk is fo rmatted. If not, a dialogue window is displayed prompting the user to format the...
Page 630
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 42 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-6-2 Backup of Patient Archive and System Configurations to MO Disks (cont’d) 7.) Select Backup Now . a.) The system will first check if the MO disk is inserted. If not, a dialogue window is displayed prompting you to insert an MO disk. Select OK to resume backup. b.) The system will then check if the MO disk is fo rmatted. If not, a dialogue window...