Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 621

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 33
8-7-13 Verifications - Softwa re Version v7.x/v6.x/v5.x/v4.x
NOTE: Software Option Strings are software versio n dependant. If a new software version has been 
installed, please follow installa tion instructions for that softwar e. Usually, new software option 
strings are needed.
8-7-13-1 Verify SW Option Key Settings 1.) Press CONFIG
 (F2) and log on as  ADM, see  4-2-4 Log On to the...

Page 622

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 34 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-13-2 Verify Video and VCR Settings  - Software Version v7.x/v6.x/v5.x/v4.x
1.) Select  Service (lower, right corner)
.
6.) From Video Settings pull down menu, verify  Format as recorded in  Table 8-7 on page 8-22 . 
Correct the selection if needed.
7.) From the VCR pull down menu, verify  VCR type as recorded in  Table 8-7 on page 8-22...

Page 623

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 35
8-7-13-3 Verify TCP/IP Settings - So ftware Version v7.x/v6.x/v5.x/v4.x
1.) From the top of the Connec tivity screen, select the Tcpip tab.
2.) Verify that all settings on the TCPIP tab matches those recorded in  Table 8-6 Record settings from 
Tcpip screen - software version v7.x/v6.x/v5.x/v4.x on page 8-21 .
3.) Select  Save settings .
4.) Select  Ok to continue.
8-7-13-4 DC Offset...

Page 624

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 36 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-14 Restore Patient Archive/System  Config. - Software v7.x/v6.x/v5.x/v4.x
1.) Press Config (F2) . 
2.) Select the category  Admin. 
3.) Select the  Restore tab. 
4.) In the  Restore tab select as needed: 
- Patient archive  to restore the patient archive. 
- System configuration  to restore all system settings and user presets. OR 
-...

Page 625

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 37
8-7-14 Restore Patient Archi ve/System Config. - Software v7.x/v6.x/v5.x/v4.x (cont’d)
8.) Select Restore Now . Depending on the selection, one or two  Restore confirmation windows are 
displayed: 
9.) Ensure that the correct so urce is selected an select OK. The selected items are copied to the 
systems. If items from the  System Configuration are restored the system needs to be rebooted....

Page 626

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 38 Section 8-8 - Software Loading Procedure - Software Version v3.x
Section 8-8
Software Loading Procedur e - Software Version v3.x
8-8-1 Introduction
This procedure describes how to perform a complete  installation of the software on the Vivid 7 hard disk. 
8-8-2 Preparations Before  Software Re-installation
• Export Patient Images from the internal hard disk to removable  removable media . 
• Make backup of Customer Presets,...

Page 627

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 39
8-8-3 Customer Provided Prerequisite
• Formatted and labelled  removable media  for Images storage
• Formatted and labelled MO Disk for Patient Archive and User Defined Settings.
8-8-4 Tools Provided With Un it or After an Upgrade
8-8-5 Preparations
1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see  4-2-2 Power ON/ Boot UP 
on page 4-3 .
2.) Store images to  removable...

Page 628

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 40 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-6 Backup - Software Version v3.x
The Backup function enables the user to:
• Make backup of the patient arch ive (patient data and reports).
• Move images from the local hard drive to MO disks.
• Make backup of the system configuration.
8-8-6-1 Move Images From the Local Hard Drive to MO disks. The Move Images  function is used to move archived images  from...

Page 629

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 41
8-8-6-1      Move Images From the Local Hard Drive to MO disks. (cont’d) 7.) Press Backup Now.- The system is checking that the MO disk is inserted. If not, a dialogue window is displayed  prompting the user to insert a MO  disk. Select OK to resume backup.
- The system is checking that the MO disk is fo rmatted. If not, a dialogue window is displayed 
prompting the user to format the...

Page 630

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 42 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-6-2      Backup of Patient Archive and System Configurations to MO Disks (cont’d) 7.) Select Backup Now .
a.) The system will first check if the MO disk is  inserted. If not, a dialogue window is displayed 
prompting you to insert an MO disk. Select  OK
 to resume backup.
b.) The system will then check if the MO disk is fo rmatted. If not, a dialogue window...
Start reading GE Vivid 7 Service Manual

Related Manuals for GE Vivid 7 Service Manual

All GE manuals