GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 73 8-9-14-2 Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d) 7.) The Software loader program starts and the below Time Bar Info Screen appears. 8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 9.) You will now be prompted to remove the first CD and restart th e unit when the...
Page 662
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 74 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14-3 Application SW Installation - Software Version v2.x 1.) Insert the applicable CD labelled Vivid7 & EchoPAC PC Application SW , see Table 8-14 CDs with BT’02 Software, Version v2.x on page 8-55 . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until...
Page 663
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 75 8-9-15 Verifications - Software Version v2.x NOTE: Software Option Strings are software versio n dependant. If a new software version has been installed, please follow installa tion instructions for that softwar e. Usually, new software option strings are needed. 8-9-15-1 Verify Software Option Key Settings - Software Version v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log...
Page 664
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 76 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-15-2 Verify Video and VCR Settings - Software Version v2.x 1.) Select Service (lower, right corner) . 2.) From Video Settings pull down menu, verify Format as recorded in Table 8-19 Video Settings and VCR Type on page 8-68 . Correct the selection if needed. 3.) From the VCR pull down menu, verify VCR type as recorded in Table 8-19 on page 8-68 ....
Page 665
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 77 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x 1.) Select Connectivity . 2.) Select Service. 3.) Select Standard Print . 4.) Select Open. Figure 8-74 Choose Standard Printer Connectivity Standard Print Open Services Scroll down to “Standard Print”
Page 666
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 78 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 5.) Select Printer. 6.) Select Properties. 7.) Select Standard TCP\IP Port . 8.) Select Configure Port .Figure 8-75 Properties Figure 8-76 Configure Port Printer Properties Standard TCP\IP Port. Configure Port
Page 667
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 79 8-9-16 Restore Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 9.) Verify Printer Name or IP Address as recorded in Table 8-18 on page 8-67 . 10.) Select Raw. 11.) Verify Port Number as recorded in Table 8-18 on page 8-67 . 12.) Select ok. 13.) Close the rest of the Windows by selecting ok. 8-9-16-1 DC Offset Calibration 1.) Disconnect all probes. 2.) From...
Page 668
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 80 Section 8-10 - BEP Filter Assembly Replacement Procedure Section 8-10 BEP Filter Assembly Re placement Procedure 8-10-1 Manpower One person, 5 minutes 8-10-2 Tools No tool required for installation and removal of the Filter Assembly. 8-10-3 Preparations None 8-10-4 BEP Filter Removel Procedure Follow these steps to remove the Filter Assembly: 1.) Press down the upper rib and pull out the upper part of the assembly before...
Page 669
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 81 8-10-6 Verification - Functional Checks • Please perform the tests specified in Section 8-20 Verification - Functional Checks on page 8-105 .
Page 670
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 82 Section 8-11 - AC Power Replacement Procedure Section 8-11 AC Power Replacement Procedure 8-11-1 Manpower One person, 1 hour. 8-11-2 Tools #2 Phillips screwdriver 8-11-3 Preparations 1.) Shut down the system, see 4-2-3 Power Shut Down on page 4-7 . 2.) Remove Right Side Cover, see Side Covers Removal Procedure on page 8-4 . 3.) Remove Filter Cover, see Upper Rear Cover Removal Procedure on page 8-6 . 4.) Remove Lower Rear...