Page 641

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 53
8-8-12-3 Verify TCP/IP Setti ngs - Software Version v3.x
1.) From the top of the con nectivity tab, select the Tcpip tab.
2.) Verify that all settings on the TC PIP tab matches those recorded in Table 8-11 Record settings from 
Tcpip screen (v3.x) on page 8-44 .
3.) Select  Save settings .
4.) Select  Ok to continue.
8-8-12-4 DC Offset Calibration 1.) Disconnect all probes.
2.) From the...

Page 642

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 54 Section 8-9 - Software Loading Procedure - Software Version v2.x
Section 8-9
Software Loading Procedur e - Software Version v2.x
8-9-1 Introduction
This procedure describes how to perform a complete  installation of the software on the Vivid 7 hard disk. 
8-9-2 Preparations To Be Done Be fore Software Re-installation
• Export Patient Images from the internal  hard disk to removable MO disks. 
• Make backup of Customer...

Page 643

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 55
8-9-3 Customer Provided Prerequisite
• Formatted and labelled MO  Disks for Images storage
• Formatted and labelled MO Disk for Patient Archive and User Defined Settings.
8-9-4 Tools Provided Wi th Unit or After a Software Upgrade
8-9-4-1 Units with BT’02 Software
NOTE: Units with software version v2.3 .0 or below and all version v1.x should be upgraded to software 
version v2.3.1 as...

Page 644

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 56 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-6 Prepare MO Disks for Image  Storage - Software Version v2.x
NOTE: When formatting the MO disks, give them logi cal names and make sure to physically label the 
disks. 
- A labelling convention should be followed so  that each MO disk gets a unique label. 
- A maximum numbers of eleven (11) label characters are allowed in the label name. 
- Use different...

Page 645

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 57
8-9-7 Move Images - Software Version v2.x
The  Move Images  function enables the movement of archived images from one volume/media to 
another and the deletion of the moved images from the original volume. The pointers to the database 
are restored in the new volume/media.
NOTE: This function copies image data on ly, no database information is copied.
Move Images from Local Archive to a...

Page 646

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 58 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-7Move Images - Software Version v2.x (cont’d) 4.) Select Move Images  from the More pull-down menu ( Figure 8-51) to display the MOVE IMAGES 
window ( Figure 8-52 ).
5.) Adjust  From date  and To  d a t e  to select the images fr om a defined time period.
6.) Check  Keep days together  if images from examinations performed on the same day are to be kept 
on...

Page 647

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 59
8-9-8 Prepare MO Disk for Patien t Archive - Software Version v2.x
NOTE: A separate MO disk is required for backup of  Patient Database, System Configuration and User 
Defined Configurations (customer presets).
This MO disk must be a separate disk with no images stored.
1.) Label the MO Disk “Backup”.
2.) If the MO Disks is already formatted and labelled, move to  section 8-9-9 on page...

Page 648

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 60 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-8 Prepare MO Disk fo r Patient Archive - Software Version v2.x (cont’d)
12.) Verify the disk formattin g status has been updated.
13.) Press  ALT-E.
14.) Select the  5 1/4 MO Disk, to eject Disk.
15.) Optional:  Insert the MO disk, side B and repeat step  7 to  14.
8-9-9 Backup of Patient Database &  User Presets - Software Version v2.x
This procedure...

Page 649

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 61
8-9-9 Backup of Patient Data base & User Presets - Software Version v2.x (cont’d)
5.) Select Backup now . A Warning will appear on th e screen, as shown in Figure 8-55 on page 8-61 .
.
6.) Select OK.
NOTE: No process indicator will appear on screen. Veri fy that the “Finished, OK” Status appears in the 
Result column, and that the date appearing in th e Last Successful Backup column is the...

Page 650

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 62 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-10 Record SW Option K eys - Software Version v2.x
1.) Press CONFIG (F2) and log on as  adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12.
2.) From the  Admin tab, select  SYSTEM ADMIN .
3.) Record SW Options Keys  from SW Options Key window.
 
Figure 8-56   System Admin
WARNINGWARNINGMake sure that the SW Option Keys (passwords) have been recorded...