Page 691
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 103 Section 8-19 Brake Pedal and Direct ion Lock Assembly Replacement Procedures 8-19-1 Manpower One person, 1 hour 8-19-2 Tools • Phillips screwdriver size 2 • Hexagon key 5 mm • Torx screwdriver T-20 • Wrench 11 and 13 mm 8-19-3 Preparations 1.) Remove Front Bumper, see Front Bumper Removal Procedure on page 8-100 . 2.) Remove footrests on both pedals by unscrewing 4 Torx screws on each....
Page 692
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 104 Section 8-19 - Brake Pedal and Direction Lock Assembly Replacement Procedures 8-19-5 Brake Pedal and Directi on Lock Installation Procedure 1.) Put the Brake Pedal and Direction Lock Assembly in position and fasten the 4 hexagon screws loosely in the sliding nuts. 2.) Fasten the ball joint nuts before tightening the 4 hexagon screws. 3.) Install Front Bumper, see Front Bumper Installation Procedure on page 8-100 . 4.)...
Page 693
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 105 Section 8-20 Verification - Functional Checks 8-20-1 Functional Checks and Tests The following table lists the required Function al Tests, based upon the FRU being replaced. Table 8-22 Functional Test Matrix sheet 1 of 3 SectionFRU Replacement Diagnostics Leakage Current Functional Test Basic Functional Checks 3-6-6-1 Connect Mains Power to the Unit on page 3-23 4-2-2 Power ON/ Boot...
Page 694
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 106 Section 8-20 - Verifi cation - Functional Checks 8-6 Lower Rear Cover Replacements Procedure on page 8-10 4-2-2 Power ON/ Boot UP on page 4-3 4-2-3 Power Shut Down on page 4-7 3-6-5-1 Connect a Probe on page 3-21 4-3-5 Performance Tests on page 4-31 4-3-6 2D Mode (B mode) Checks on page 4-32 4-3-21-3 Brakes and Direction Lock Checks on page 4-80 8-7 Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x...
Page 695
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 107 8-14 Rear Casters Replacement Procedure on page 8-91 4-3-21-3 Brakes and Direction Lock Checks on page 4-80 Perform Basic Functional Checks on page 8-105 8-15 Front Casters Replacement Procedure on page 8-95 4-3-21-3 Brakes and Direction Lock Checks on page 4-80 Perform Basic Functional Checks on page 8-105 8-16 Front Bumper Replacement Procedure on page 8-100 4-3-21-3 Brakes and...
Page 696
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 108 Section 8-20 - Verifi cation - Functional Checks This page was intentionally left blank.
Page 697
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 9 - Renewal Parts 9 - 1 Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts for Vivid 7. 9-1-2 Contents in Chapter 9 Table 9-1 Contents in Chapter 9 SECTIONDESCRIPTION PAGE NUMBER 9-1 Overview 9-1 9-2 Definitions of Left, Right, Front and Back 9-2 9-3 List of Abbreviations 9-3 9-4Parts List Groups9-4 9-5Vivid 7 Models and Hardware/Software...
Page 698
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 9 - 2 - Section 9-2 Definitions of Left, Right, Front and Back 9-23Cables - Vivid 79-69 9-24ECG Cables9-78 9-25 Physio TX Parts 9-79 9-26 Probes 9-80 9-27 Probe Service Parts 9-90 9-28Other Probe Part(s)9-90 9-29Options9-91 9-30Kits9-99 9-31 Accessory Boxes, Vivid 7 / Vivid 7 PRO 9-105 9-32 Product Manuals 9-108 9-33 Packing Parts for Reshipment - Vivid 7 with LCD 9-120 9-34Packing Parts for Reshipment - Vivid 7 with CRT9-121...
Page 699
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 9 - Renewal Parts 9 - 3 Section 9-3 List of Abbreviations Table 9-2 List of Abbreviations ABBREVIATIONSDESCRIPTION 3D THREE DIMENSIONAL (SEE RT3D and 4D) 4D 4 Dimensional is the same as RT3D. 3D view in real time. The time is the 4th dimension. ACP AC CONTROLLER (AC POWER) ACTAC TRANSFORMER ASSY ASSEMBLY BEPBACK-END PROCESSOR CTRL CONTROL EXT. EXTERNAL F. O . B . FLOCK OF BIRDS—FOR 3D OPTION NOTE: “FLOCK OF BIRDS”...
Page 700
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 9 - 4 Section 9-4 - Parts List Groups Section 9-4 Parts List Groups Table 9-3 Parts List Groups sheet 1 of 3 TABLE NO.DESCRIPTION PAGE NUMBER Table 9-4 Vivid 7 Models and Hardware/Software Compatibility 9-7 Table 9-5 Software used on BT’08 (Software Version v7.x) 9-10 Table 9-6 Software used on BT’06 (Software Version v6.x) 9-11 Table 9-7Software used on BT’05 (Software Version v5.x)9-12 Table 9-8Software used on BT’04...