GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 63 8-9-11 Record TCP/IP Setti ngs - Software Version v2.x 1.) Select Connectivity . 2.) Select the Tcpip tab. 3.) Record all settings in Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . Figure 8-57 TCP/IP Example Settings Table 8-16 Record settings from Tcpip screen (v2.x) PARAMETERVALUE Computer Name VIVID7-00 _ _ _ _ AE Title VIVID7-00 _ _ _ _ IP Settings IP...
Page 652
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 64 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-12 Record Default Dataflow - Software Version v2.x 1.) In the Connectivity window, select Dataflow. 2.) Verify that Default Dataflow is selected. 3.) Record Default Dataflow Name in table Table 8-17 on page 8-64. Figure 8-58 Select Dataflow Table 8-17 Record Default Dataflow Name DEFAULT DATAFLOW NAME
Page 653
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 65 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x 1.) In Connectivity window, select Services. 2.) In the Services list, Scroll down until Standard Print appears. 3.) Select Standard Print . 4.) Select Open. 5.) Select Printer to open the Printer scroll down menu. 6.) Select Properties from the Printer scroll down menu.Figure 8-59 Choosing Standard...
Page 654
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 66 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 7.) Select Ports. 8.) Select Standard TCP\IP Port . 9.) Select Configure Port .Figure 8-61 Configure Port Figure 8-62 Configure Port (Port Settings dialog) STANDARD TCP\IP PORT CONFIGURE PORT PORTS
Page 655
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 67 8-9-13 Record Standard Printer IP and Port Settings - Software Version v2.x (cont’d) 10.) Select and record Printer Name or IP Address in Table 8-18 on page 8-67 . 11.) Select Raw. 12.) Select and record Port Number in Table 8-18 on page 8-67 . 13.) Select Cancel. 14.) Close rest of the Windows by selecting Ok. Table 8-18 Printer Name or IP Address and Port Number PRINTER...
Page 656
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 68 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-13-1 Record Service Settings - Software Version v2.x 1.) Select Service. . 2.) Record Video Setting Format and VCR Type in Table 8-19 Video Settings and VCR Type on page 8-68 . Figure 8-64 Service Settings Table 8-19 Video Settings and VCR Type Video Setting FormatVCR Type
Page 657
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 69 8-9-13-2 Information on the About tab - Software Version v2.x 1.) Select the About tab. 2.) Record Application software version, System software version and Processor model (BEP) in Table 8-20 Record Software and hardware versions on page 8-69 . Figure 8-65 Information available on the About tab Table 8-20 Record Software and hardware versions DESCRIPTIONRECORD VERSION APPLICATION...
Page 658
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 70 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14 Software Loading - Software Version v2.x 8-9-14-1 Overview This software loading procedure covers: • Installation of BT’02 Software (application software version v2.x) on Vivid 7 units. Refer to Table 8-20 on page 8-69 where you recorded the unit’s software versions. The software for Vivid 7 is divided on two CDs, one CD with the System Software and...
Page 659
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 71 8-9-14-2 Vivid 7 System Software Loading - Software Version v2.x 1.) Insert the applicable Vivid 7 System Software CD into the CD drive, see Table 8-20 Record Software and hardware versions on page 8-69 . 2.) Power down the System, see 4-2-3 Power Shut Down on page 4-7 . 3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch has turned amber. 4.)...
Page 660
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 72 Section 8-9 - Software Loading Procedure - Software Version v2.x 8-9-14-2 Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d) 5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you recorded it in Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . Example: VIVID7-001890 Always enter two zeros before you enter the serial number. This...