Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 651

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 63
8-9-11 Record TCP/IP Setti ngs - Software Version v2.x
1.) Select Connectivity .
2.) Select the  Tcpip tab.
3.) Record all settings in  Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 .
 
 
Figure 8-57   TCP/IP Example Settings
Table 8-16    Record settings from Tcpip screen (v2.x)
PARAMETERVALUE
Computer Name VIVID7-00 _ _ _ _ AE Title VIVID7-00 _ _ _ _
IP Settings
IP...

Page 652

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 64 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-12 Record Default Dataflow - Software Version v2.x
1.) In the  Connectivity  window, select  Dataflow.
2.) Verify that Default  Dataflow is selected.
3.) Record Default Dataflow Name in table  Table 8-17 on page 8-64.
  Figure 8-58   Select Dataflow
Table 8-17    Record Default Dataflow Name
DEFAULT DATAFLOW NAME 

Page 653

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 65
8-9-13 Record Standard Printer IP and  Port Settings - Software Version v2.x
1.) In Connectivity  window, select  Services.
2.) In the  Services list, Scroll down until  Standard Print appears.
3.) Select  Standard Print .
4.) Select  Open.
5.) Select  Printer to open the Printer scroll down menu.
6.) Select  Properties from the Printer scroll down menu.Figure 8-59   Choosing Standard...

Page 654

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 66 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-13      Record Standard Printer IP and  Port Settings - Software Version v2.x (cont’d)
7.) Select  Ports.
8.) Select  Standard TCP\IP Port .
9.) Select  Configure Port .Figure 8-61   Configure Port
Figure 8-62   Configure Port (Port Settings dialog)
STANDARD TCP\IP PORT
CONFIGURE PORT
PORTS 

Page 655

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 67
8-9-13      Record Standard Printer IP and  Port Settings - Software Version v2.x (cont’d)
10.) Select and record  Printer Name or IP Address  in Table 8-18 on page 8-67 .
11.) Select  Raw.
12.) Select and record  Port Number in Table 8-18 on page 8-67 .
13.) Select  Cancel.
14.) Close rest of the Windows by selecting  Ok.
Table 8-18    Printer Name or IP Address and Port Number
PRINTER...

Page 656

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 68 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-13-1 Record Service Settings - Software Version v2.x 1.) Select Service.
.
2.) Record Video Setting Format and VCR Type in  Table 8-19 Video Settings and VCR Type on 
page 8-68 .
 
Figure 8-64   Service Settings
Table 8-19    Video Settings and VCR Type
Video Setting FormatVCR Type 

Page 657

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 69
8-9-13-2 Information on the About tab - Software Version v2.x 1.) Select the  About tab.
2.) Record Application software version, System  software version and Processor model (BEP) in Table 
8-20 Record Software and hardware versions on page 8-69 .
 
Figure 8-65   Information available on the About tab
Table 8-20    Record Software and hardware versions
DESCRIPTIONRECORD VERSION
APPLICATION...

Page 658

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 70 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-14 Software Loading - Software Version v2.x
8-9-14-1 Overview This software loading procedure covers: 
• Installation of BT’02 Software (application software version v2.x) on Vivid 7 units.
 
Refer to Table 8-20 on page 8-69  where you recorded the unit’s software versions.
The software for Vivid 7 is divided on two CDs, one CD with the System Software and...

Page 659

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 71
8-9-14-2 Vivid 7 System Software Loading - Software Version v2.x 1.) Insert the applicable  Vivid 7 System Software CD  into the CD drive, see Table 8-20 Record 
Software and hardware versions on page 8-69 .
2.) Power down the System, see  4-2-3 Power Shut Down on page 4-7 .
3.) Wait until the LCD’s on the Operator Panel are s witched off and the ON/Standby switch has turned 
amber.
4.)...

Page 660

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 72 Section 8-9 - Software Loading Procedure - Software Version v2.x
8-9-14-2              Vivid 7 Syst em Software Loading - Software Version v2.x (cont’d)
5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you 
recorded it in  Table 8-16 Record settings from Tcpip screen (v2.x) on page 8-63 . 
Example:  VIVID7-001890  Always enter two zeros before you enter the serial  number. This...