Page 631
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 43 8-8-7 Record SW Opti on Keys - Software v3.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) From the Admin tab, select SYSTEM ADMIN . 3.) Record SW Options Keys from SW Options Key window. Figure 8-37 System Admin WARNINGWARNINGMake sure that the SW Option Keys (passwords) have been recorded correctly. If the key is incorrect, you...
Page 632
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 44 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-8 Record TCP/IP Settings - Software Version v3.x 1.) Select Connectivity . 2.) Select the Tcpip tab. 3.) Record all settings in Table 8-11 Record settings from Tcpip screen (v3.x) on page 8-44 . Figure 8-38 TCP/IP Example Settings Table 8-11 Record settings from Tcpip screen (v3.x) PARAMETERVALUE Computer Name VIVID7-00 _ _ _ _ AE Title...
Page 633
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 45 8-8-9 Record Service Settings - Software Version v3.x 1.) Select Service. . 2.) Record Video Setting Format and VCR Type in Table 8-12 on page 8-45. Figure 8-39 Service Settings Table 8-12 Video Setting and VCR type VIDEO SETTING FORMATVCR TYPE ___ PAL ___NTSC
Page 634
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 46 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-10 Information on the About tab - Software Version v3.x 1.) Select the About tab. 2.) Record software and hardvare versions in Table 8-13 on page 8-46. Figure 8-40 Information available on the About tab Table 8-13 Record Software and Hardware versions (v3.x) DESCRIPTIONRECORD VERSION APPLICATION SW SYSTEM SW Processor (BEP)
Page 635
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 47 8-8-11 Software Loading - Software Version v3.x 8-8-11-1 Overview This software loading procedure covers: • Installation of BT’03 software (sof tware version v3.x) on Vivid 7 units. Refer to Table 8-13 on page 8-46 where you recorded the unit’s software versions. The software for Vivid 7 is divided on two CDs, on e CD with the System Software and one with the Application Software....
Page 636
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 48 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-11-2 Vivid 7 S ystem SW Loading - Software Version v3.x (cont’d) 5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you recorded it in Table 8-6 Record settings from Tcpip screen - software version v7.x/v6.x/v5.x/v4.x on page 8-21 . Example: VIVID7-001890 Always enter two zeros before you...
Page 637
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 49 8-8-11-2 Vivid 7 S ystem SW Loading - Software Version v3.x (cont’d) 7.) The Software loader program starts and the below Time Bar Info Screen appears. 8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the screen indicates process. 9.) You will now be prompted to remove the first CD and restart th e unit when the SW...
Page 638
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 50 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-11-3 Application SW Installation - Software Version v3.x 1.) Insert the applicable CD labelled Vivid7 & EchoPAC PC Application SW , see Table 8-9 CDs with BT’03 Software, SW. v3.x on page 8-39 . 2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the scanner. (This is called Forced Shut Down). 3.) Wait until the...
Page 639
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 51 8-8-12 Verifications - Software Version v3.x NOTE: Software Option Strings are software versio n dependant. If a new software version has been installed, please follow installa tion instructions for that softwar e. Usually, new software option strings are needed. 8-8-12-1 Verify SW Option Key Se ttings - Software Version v3.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On...
Page 640
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 52 Section 8-8 - Software Loading Procedure - Software Version v3.x 8-8-12-2 Verify Video and VCR Se ttings - Software Version v3.x 1.) Select Service (lower, right corner) . 6.) From Video Settings pull down menu, verify Format as recorded in Table 8-12 on page 8-45 . Correct the selection if needed. 7.) From the VCR pull down menu, verify VCR type as recorded in Table 8-12 on page 8-45 . Correct the selection if...