8-8-7 Record SW Opti on Keys - Software v3.x
1.) Press CONFIG (F2) and log on as  adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12.
2.) From the  Admin tab, select  SYSTEM ADMIN .
3.) Record SW Options Keys  from SW Options Key window.
 
Figure 8-37   System Admin
WARNINGWARNINGMake sure that the SW Option Keys (passwords) have been recorded correctly. If the 
key is incorrect, you...

Page 632

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 44 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-8 Record TCP/IP Settings  - Software Version v3.x
1.) Select Connectivity .
2.) Select the  Tcpip tab.
3.) Record all settings in  Table 8-11 Record settings from Tcpip screen (v3.x) on page 8-44 .
 
 
Figure 8-38   TCP/IP Example Settings
Table 8-11    Record settings from Tcpip screen (v3.x)
PARAMETERVALUE
Computer Name VIVID7-00 _ _ _ _ AE Title...

Page 633

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 45
8-8-9 Record Service Settings  - Software Version v3.x
1.) Select Service.
.
2.) Record Video Setting Format and VCR Type in  Table 8-12 on page 8-45.
 
Figure 8-39   Service Settings
Table 8-12    Video Setting and VCR type
VIDEO SETTING FORMATVCR TYPE
___ PAL   ___NTSC 

Page 634

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 46 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-10 Information on the About  tab - Software Version v3.x
1.) Select the About tab.
2.) Record software and hardvare versions in  Table 8-13 on page 8-46.
 
Figure 8-40   Information available on the About tab
Table 8-13    Record Software and Hardware versions (v3.x)
DESCRIPTIONRECORD VERSION
APPLICATION SW SYSTEM SW
Processor (BEP) 

Page 635

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 47
8-8-11 Software Loading - Software Version v3.x
8-8-11-1 Overview This software loading procedure covers: 
• Installation of BT’03 software (sof tware version v3.x) on Vivid 7 units.
Refer to  Table 8-13 on page 8-46  where you recorded the unit’s software versions.
The software for Vivid 7 is divided on two CDs, on e CD with the System Software and one with the 
Application Software....

Page 636

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 48 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-11-2              Vivid 7 S ystem SW Loading - Software Version v3.x  (cont’d)
5.) When requested, enter the missing part of the Co mputer Name (the unit’s serial number) as you 
recorded it in  Table 8-6 Record settings from Tcpip screen  - software version v7.x/v6.x/v5.x/v4.x 
on page 8-21 . 
Example:  VIVID7-001890  Always enter two zeros before you...

Page 637

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 49
8-8-11-2              Vivid 7 S ystem SW Loading - Software Version v3.x  (cont’d)
7.) The Software loader program starts and the below Time Bar Info Screen appears.
  
8.) Wait for the SW Installation to be completed. (Typic al installation time: 5 minutes). Status bar on the 
screen indicates process.
9.) You will now be prompted  to remove the first CD and restart th e unit when the SW...

Page 638

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 50 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-11-3 Application SW Installation - Software Version v3.x 1.) Insert the applicable CD labelled  Vivid7 & EchoPAC PC Application SW , see Table 8-9 CDs with 
BT’03 Software, SW. v3.x on page 8-39 .
2.) Keep the ON/Standby switch on the operator panel pressed for a few seconds to power down the  scanner. (This is called Forced Shut Down).
3.) Wait until the...

Page 639

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 51
8-8-12 Verifications -  Software Version v3.x
NOTE: Software Option Strings are software versio n dependant. If a new software version has been 
installed, please follow installa tion instructions for that softwar e. Usually, new software option 
strings are needed.
8-8-12-1 Verify SW Option Key Se ttings - Software Version v3.x
1.) Press  CONFIG
 (F2) and log on as  adm, see 4-2-4 Log On...

Page 640

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 52 Section 8-8 - Software Loading Procedure - Software Version v3.x
8-8-12-2 Verify Video and VCR Se ttings - Software Version v3.x
1.) Select  Service (lower, right corner)
.
6.) From Video Settings pull down menu, verify  Format as recorded in  Table 8-12 on page 8-45 . 
Correct the selection if needed.
7.) From the VCR pull down menu, verify  VCR type as recorded in  Table 8-12 on page 8-45 . Correct 
the selection if...
Start reading GE Vivid 7 Service Manual

Related Manuals for GE Vivid 7 Service Manual

All GE manuals