GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 581
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 69 7-8-3-4 System Test Screen Description 7-8-3-5 Test Loops, System Figure 7-74 System Test Screen, Upper Part Table 7-6 Test Loops, System - Test All, System - Break (FEP1) FIELDKEYDESCRIPTION TEST LOOPS -IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES. ENTER A NUMBER TO INDICATE HOW MANY TI MES YOU WANT THE TEST TO REPEAT ITSELF. DEFAULT VALUE = 1.. SYSTEM TEST...
Page 582
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 70 Section 7-8 - Acquisition Diagnostics 7-8-3-5 Test Loops, System (cont’d) 7-8-3-6 Mode (Calibration/Test Screen Selection) Table 7-7 Test Loops, System - Test All, System - Break (FEP2) FIELDKEYDESCRIPTION TEST LOOPS -IN SOME CASES YOU MAY WANT TO DO THE SAME TEST SEVERAL TIMES. ENTER A NUMBER TO INDICATE HOW MANY TI MES YOU WANT THE TEST TO REPEAT ITSELF. DEFAULT VALUE = 1.. SYSTEM TEST ALL USE TEST ALL TO...
Page 583
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 71 7-8-3-7 “Block Schematics” with Single Tests Select one of the buttons in the “block schematics” in the upper part of the screen (or one of the duplicate buttons in the lower row of the screen) to select a particular test. Some of the tests are described after the figure below. Figure 7-75 Diagnostics Test Screen, Upper Part BLOCK SCHEMATICS, FEP2BLOCK SCHEMATICS, FEP1
Page 584
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 72 Section 7-8 - Acquisition Diagnostics 7-8-3-8 Receiver Board (RX) Test 7-8-3-9 Relay Board Test NOTE: A Test Attenuator, Part Number AA200215 is needed for this test. Table 7-9 Receiver Board (RX) Test Test (Key)Sub-testsDescription RX Receiver Board (RX) Te s t A l l Activates the following tests, in s equence, see description for each sub test; - RECV0 - RECV1 RECV.0 Max Gain test for RX using FEC signal generator....
Page 585
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 73 7-8-3-10 Image Port Test When selecting the ImagePort Test, the scanner issues a test command to the local software running on the I960RP processor on the Image Port board. The test command will start the following tests; • test the interrupt based command/replay mechanism • reset the ImagePort board • test the PipeLink data fifo • test the video data fifos • test the PipeLink...
Page 586
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 74 Section 7-8 - Acquisition Diagnostics 7-8-4 General Recommendations to Troubleshoot Noise Issues Ultrasound machines are susceptible to Electromagne tic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid 7 complies with limits as stated on the EMC label. Ho wever there is no guarantee that interference will not occur in a particular...
Page 587
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 75 7-8-5 EMI Prevention/abatement Table 7-12 EMI Prevention/abatement EMI RULEDETAILS BE AWARE OF RF SOURCES - Keep the unit at least 5 meters (15 feet) away from other emi sources. - Special shielding may be requi red to eliminate interference problems caused by high frequency, - High powered radio or video broadcast signals. GROUND THE UNIT - Poor grounding is the most...
Page 588
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 76 Section 7-8 - Acquisition Diagnostics This page was intentionally left blank.
Page 589
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 1 Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter holds replacement procedures for the following modules/subsystems. NOTE: Use the ESD-kit during replacing boards. WARNINGWARNINGDO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE NECESSARY ESD PRECAUTIONS: ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD...
Page 590
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 2 - 8-1-2 Definitions of Le ft, Right, Front and Back Figure 8-1 Definition of Left, Right, Front and Back of Vivid 7 RIGHT SIDE REAR / BACK LEFT SIDE FRONT