Page 591

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 3
8-1-3 Returning/Shipping  Probes and Repair Parts
Equipment being returned must be clean and free of blood and other infectious substances.
GEHC policy states that body fluids must be proper ly removed from any part or equipment prior to 
shipment. GEHC employees, as well as customers,  are responsible for ensuring that parts/equipment 
have been properly decontaminated prior to ship ment....

Page 592

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 4 Section 8-2 - Side Covers (with Bumpers) Replacement Procedure
Section 8-2
Side Covers (with Bumpers)  Replacement Procedure
8-2-1 Manpower
One person, 5 minutes
8-2-2 Tools
Large coin and a large screwdriver
8-2-3 Preparations
Power off / Shutdown the system, see  4-2-3 Power Shut Down on page 4-7.
8-2-4 Side Covers Removal Procedure
1.) Use a large coin to turn the two recessed fastene rs at the bottom of the side cover a...

Page 593

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 5
8-2-5 Side Covers Installation Procedure
1.) Guide the top edge of the side cover so the hooks rest in the top cover retainer slots.
2.) Push back side in first position so  it is locked in Lower Rear Cover.
3.) Push the front side in position  so it is locked in Front Cover.
4.) With a large coin, turn the two bottom fasten ers to the vertical position.Now, using a large 
screwdriver, push...

Page 594

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 6 Section 8-3 - Upper Rear Cover (Filter Cover) Replacement Procedure
Section 8-3
Upper Rear Cover (Filter Co ver) Replacement Procedure
8-3-1 Manpower
One person, 5 minutes
8-3-2 Tools
No special tools needed
8-3-3 Upper Rear Cover Removal Procedure
Take hold at the lower side of Filter  Cover and pull it loose from the chas sis. Hook it off at the upper side 
and remove it.
8-3-4 Upper Rear Cover  Installation Procedure
Align...

Page 595

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 7
Section 8-4
Front Cover Repla cement Procedure
8-4-1 Manpower
One person, 30 minutes
8-4-2 Tools
Phillips screwdriver
8-4-3 Preparations
1.) Shut down the system, see  4-2-3 Power Shut Down on page 4-7 .
2.) Disconnect all Probes and I/ O cabling. See User manual. 
3.) Remove Side Covers, see  Side Covers Removal Procedure on page 8-4 .
8-4-4 Front Cover Removal Procedure
1.) Unscrew five...

Page 596

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 8 Section 8-5 - Air Filter Replacement Procedure
Section 8-5
Air Filter Replacement Procedure
8-5-1 Manpower
One person, 30 minutes 
8-5-2 Tools
No special tools are needed
 
8-5-3 Filter Removal Procedure
1.) Remove Filter Cover, see  8-3-3 Upper Rear Cover Removal Procedure on page 8-6 .
2.) Pull out plastic plugs and remove the filter.
8-5-4 Filter Installation Procedure
Two different Plastic plugs have been used, see...

Page 597

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 9
8-5-5 Verification - Functional Checks
• Please perform the tests specified in  Section 8-20 Verification - Functional Checks on page 8-105 . 

Page 598

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 10 Section 8-6 - Lower Rear Cover Replacements Procedure
Section 8-6
Lower Rear Cover Replacements Procedure
8-6-1 Manpower
One person, 30 minutes
8-6-2 Tools
Phillips screwdriver size 2
8-6-3 Preparations
Shut down the system, see  4-2-3 Power Shut Down on page 4-7 .
8-6-4 Lower Rear Cover Removal Procedure
1.) Remove Filter Cover shown in  Upper Rear Cover Removal Procedure on page 8-6 .
2.) Remove both Side Covers, see  Side...

Page 599

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 11
8-6-6 Verification - Functional Checks
• Please perform the tests specified in  Section 8-20 Verification - Functional Checks on page 8-105 . 

Page 600

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 12 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
Section 8-7
Software Loading Proced ure - Software Version v7. x, v6.x, v5.x and v4.x
8-7-1 Introduction
 
8-7-2 Preparations Before  Software Re-installation
• Export Patient Images from the internal hard disk to removable removable media. 
• Make backup of Customer Presets, Patient Databas e (Patient Archive, Report Archive, System...