GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 49 7-7-1 Common Diagnostics - Utilities Utilities has two tools, both inte nded for use when connecting fr om a remote connected computer: 1.) Disruptive Mode 2.) System Shutdown Figure 7-52 Common Diagnostics - Utilities
Page 562
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 50 Section 7-7 - Common Diagnostics 7-7-1-1 Disruptive Mode Disruptive Mode utility is used to enable/disable Disruptive Mode. Disruptive Mode must be used if a remote comput er is used to control Vivid 7. Example: InSite. 7-7-1-2 System Shutdown Intended for use from remote computer. See: System Shutdown on page 7-43. Figure 7-53 Common Diagnostics - Utilities - Disruptive Mode Utility
Page 563
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 51 7-7-2 PC (Back-End Processor) Dia gnostics, Non-Interactive Tests The following sub-sections outline the non-interactive BEP Diagnostic tests. 7-7-2-1 CPU Tests •CPU Test (All) - This diagnostic performs a battery of tests on your CPU to ensure it is working properly. The following tests are a complete lis t of the tests performed on the PCs CPU. • CPU: Registers - The CPU...
Page 564
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 52 Section 7-7 - Common Diagnostics 7-7-2-1 CPU Tests (cont’d) •CPU: Transcendental - Tests coprocessor transcendental commands F2XM1, FPATAN, FPTAN, FYL2X, FYL2XP1, FCOS, FSIN and FSINCOS. If errors are detected, the failed commands are listed. • CPU: MMX Arithmetic - Tests MMX commands PADD, PSUB, PMUL and PMADD. If errors are detected, the failed commands are listed. • CPU: MMX Logical - Tests MMX commands PAND,...
Page 565
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 53 7-7-2-2 Hard Drive Tests •Hard Drive Test (All) - The Hard Drive test examines your hard disk drive using a series of tests to ensure that your drive can perform its functions. The following tests are performed on your hard drive to ensure the hard disk controller and the drive mechanism are working correctly. The disk surface itself is also checked. NOTE: The hard drive test can...
Page 566
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 54 Section 7-7 - Common Diagnostics 7-7-2-3 Memory Tests •Memory Test (All) - Eighteen test patterns are used to test memory locations. If errors are detected, the address of the block the error was detected in is shown. During testing, these diagnostics will also look out for pa rity errors and other exceptions. These will be added to the Test Log as errors if encountered. • Memory: Pattern Test - The program uses...
Page 567
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 55 7-7-2-4 CD/DVD-ROM Drive Tests •CD-ROM Test (All) - The CD-ROM test examin es the CD-ROM drive to make sure it is working properly. Use a data CD-ROM disc for these tests. You cannot use an audio CD, such as a regular commercial music CD, with this test. NOTE: The BEP is sometimes slow in recognizing a ne wly inserted CD-ROM disc. It is best to wait a few seconds after inserting a...
Page 568
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 56 Section 7-7 - Common Diagnostics 7-7-2-5 Video Test •Video Test (All) - This diagnostic tests your systems vi deo capabilities. This involves testing the video memory with 18 patterns, testing your graphics acceleration, and text output. You will see these tests being performed on your monitor. You can cancel this test at any time by hitting the Escape (Esc) key. The following tests are performed on your monitor: •...
Page 569
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 57 7-7-3 PC (Back-End Processor) Di agnostics, Interactive Tests 7-7-3-1 Keyboard Test 1.) Double-click Keyboard Test from the PC Diagnostics interactive tests menu. 2.) Select the EXECUTE button. A keyboard layout is displayed on which you can see the keys of a standard computer keyboard. Some of these keys are missing on Vivid 7. 3.) Press each key on your keyboard once and make...
Page 570
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 58 Section 7-7 - Common Diagnostics 7-7-3-2 Audio Test 1.) Double-click Audio Test from the PC Diagnostic interactive tests menu. 2.) To test the WAV sound reproduction select Left Ch annel, Right Channel, or Both Channels to test your speakers. You should hear a guitar chord. 3.) To test MIDI sound reproduction select the button under MIDI. You should hear a few seconds of recorded music. 4.) Select the Beep button. You...