Page 541
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 29 7-6-5-4 Control Frame Contains the user interface elements used for: • Diagnostic control • Operator feedback a.) Execute ButtonThis button has two modes - each with appropriate text: * Execute - to start the diagnostic, and * Abort - to stop a diagnostic b.) Loop Count This is an editable text field th at will only accept numeric values with 4 digits or less. When the button...
Page 542
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 30 Section 7-6 - Diagnostics Tools 7-6-5-4 Control Frame (cont’d) Examine the log in the status frame to find the test(s) that failed. (In the example in Figure 7-33, all the tests failed.) Figure 7-33 User Interface Diagnostic Screen, Test All Failed
Page 543
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 31 7-6-5-4 Control Frame (cont’d) In Figure 7-34 , the test passed. The green background color in the Control Frame indicates that the test passed. 7-6-5-5 Common Diagnostics See: Section 7-7 Common Diagnostics on page 7-48.Figure 7-34 User Interface Diagnostic Screen, Test All Passed
Page 544
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 32 Section 7-6 - Diagnostics Tools 7-6-6 Image Quality Field is not populated. Figure 7-35 Image Quality Page
Page 545
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 33 7-6-7 Calibration This screen gives you access to: • DC Offset calibration • Reset DC-Offset cal (calibration) Figure 7-36 Calibration
Page 546
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 34 Section 7-6 - Diagnostics Tools 7-6-7-1 DC Offset Calibration DC Offset calibration also called “Beamformer Calibra tion” or “Front-End Alignment” is performed to calibrate each ADC channel of the two Beamformer (BF) board inputs. The system calibrates the Beamformer ADCs to zero output when there is no signal on the input. These bias voltages are stored on hard disk in Back-End Processor. You can either access DC...
Page 547
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 35 7-6-8 Configuration Use the Configuration Screen to add or edit Software Options Strings, alphanumeric strings to enable Software Options (the password(s) used for enabling optional functions). Figure 7-38 Configuration Page
Page 548
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 36 Section 7-6 - Diagnostics Tools 7-6-9 Utilities The Utilities page contains several miscellaneous tools: • Windows 2000 Utilities - Disk Usage - IP Configuration - Network Status - Windows 2000 Services - User Accounts - Shared Resources - System Shutdown - Disk Defragmenter • Scanner Utilities - System Shutdown - Set Debug Utility - Reset Database - Clean Userdefs - Resource File Utility • iLink Configuration Utility •...
Page 549
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 37 7-6-9-1 Windows 2000 Utilities Disk Usage The Disk Usage page displays an ov erview of Vivid 7’s disk usage. Figure 7-40 Disk Usage
Page 550
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 38 Section 7-6 - Diagnostics Tools 7-6-9-1 Windows 2000 Utilities (cont’d) IP Configuration The IP Configuration page gives you an overview of Vivid 7’s Internet Protocol (IP) configuration. Figure 7-41 IP Configuration