GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 511
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 11 6-8-2-4 Direction Lock Adjustment (cont’d) 5.) Turn adjustment nuts un til the balljoints matches. 6.) Press the ba lljoint together. 7.) Install the lock pins, see Figure 6-5. 8.) Fasten the retaining nuts. 9.) Test Direction Lock. Figure 6-6 Balljoint with Lock Pin Adjustment nut Retaining nut Lock Pin
Page 512
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 12 Section 6-8 - Direction Lock and Brake Adjustment 6-8-3 Rear Brakes Adjustment 6-8-3-1 Manpower 2 persons, 1 hour 6-8-3-2 Tools • Hexagon key 3/32 inch • Phillips screw driver size 2 6-8-3-3 Preparations 1.) Remove Filter Cover , see Upper Rear Cover Removal Procedure on page 8-6 . 2.) Remove Lower Rear Cover , see Lower Rear Cover Removal Procedure on page 8-10 . 3.) Remove the Caster , see Rear Casters Removal Procedure...
Page 513
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 1 Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of this Chapter This section describes how to setup and run the tool s and software that help maintain image quality and system operation. Very basic host, system and boa rd level diagnostics are run whenever power is applied. Some Service Tools may be run at the application level. NOTE: Diagnostics must be run...
Page 514
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 2 Section 7-2 - Service Tools Overview Section 7-2 Service Tools Overview 7-2-1 Service Software Tools Two major Service Software tools are available: • Common Service Diagnostic Interface (GSUI) This software tool makes it possible for you to do tests on various functions on the Back-End Processor. It also includes an acquisition test that tests most of the PCBs in the Front-End Rack. If this test fails, detailed...
Page 515
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 3 Section 7-3 Frequently Asked Questions - FAQ 7-3-1 Error Message(s) During Start Up Table 7-2 Error Message(s) During Start Up ERROR MESSAGEDESCRIPTIONWHAT TO DO This message, displayed when starting Vivid 7, indicates that one or more of the installed options soon will expire. Typical, this will be the case when a demo option is about to expire. If you want to keep the...
Page 516
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 4 Section 7-4 - Service Safety Considerations Section 7-4 Service Safety Considerations DANGERDANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. WARNINGWARNINGIF THE COVERS ARE REMOVED FROM AN OPERATING Vivid 7, SOME METAL SURFACES MAY BE WARM ENOUGH TO POSE A POTENTIAL HEAT HAZARD IF TOUCHED, EVEN WHILE IN SHUT DOWN MODE....
Page 517
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 5 Section 7-5 Gathering Troubleshooting Data 7-5-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition through remote diagnos tics (InSite) or to be sent back to the manufacturer for analysis. There are different options to acquire this data that would give different results. 7-5-2 Collect Vital System Information...
Page 518
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 6 Section 7-5 - Gathering Troubleshooting Data 7-5-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously . This will collect a screen capture of the monitor, system presets and several lo g files in a date and time stamped zip file. NOTE: Power Supply and Temperature logs are not currently being updated by the Vivid 7. This Alt-D function is available at all times. When...
Page 519
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 7 Section 7-6 Diagnostics Tools 7-6-1 Diagnostic Procedure Summary 7-6-1-1 Purpose of this Sub-Section These software based diagnostics tools are available on Vivid 7: •Common Service Diagnostic Interface (GSUI) and • Vivid 7’s Front-End System Test To be able to use these Service Tools, the Back-End Processor must be up and running to some extent. If the Back-End Processor doesn’t...
Page 520
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 8 Section 7-6 - Diagnostics Tools 7-6-2 Common Service Diagnostic Interface (GSUI) NOTE: It is not necessary to connect a PC to Vivid 7 to use the GSUI. 7-6-2-1 Log-in to GSUI Double-click the button with the wren ch icon ( ) in the status bar at the bottom of the video screen. This button links the user or the Field En gineer (FE) to the Service Login Screen. Log on as External Service with password: gogems (all...