Page 501
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 1 Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this Chapter This section describes how to adjust the scanner. 6-1-2 Contents in this Chapter Section 6-2 Date and Time Adjustments Please refer to: 3-7-1-3 Date and Time Adjustments on page 3-29 . Section 6-3 Daylight Saving Time (DST) Adjustments From the factory, the automatic DST function has been disabled. Manually...
Page 502
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 2 - Section 6-5 LCD Monitor Adjustments 6-5-1 Overview This procedure describes how to adjust the LCD moni tor for use with Vivid 7, software version v6.x and newer. It also describes how to set up the Vivid 7 to the factory default for LCD use. The LCD adjustment procedure contains two main tasks: 1.) Adjust LCD monitor 2.) Select gamma curve 6-5-2 Cautions and Warnings N/A 6-5-3 Access to Adjustments The adjustments of...
Page 503
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 3 Section 6-6 CRT Monitor Adjustments 6-6-1 Cautions and Warnings N/A 6-6-2 Access to Adjustments The adjustments of the monitor is done via c ontrols (buttons) on front of the monitor. 6-6-3 Degaussing the CRT Monitor Degaussing refers to the process of removing magnetic-field effects from the monitor. Operation of the monitor within a magnetic field may adversely effect co lor purity....
Page 504
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 4 Section 6-6 - CRT Monitor Adjustments 6-6-4 Adjusting the CRT’s Contrast and Brightness The display monitor’s contrast and brightness contro ls may need periodic adjustment due to changes in ambient light. They are adjusted from the three buttons on the front part of the display monitor. 6-6-4-1 Adjust Contrast The factory default Contrast settings: • BT03/04/05/06 units: Contrast = 65% • BT02 units: Contrast = 100%...
Page 505
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 5 6-6-4-2 Adjust Brightness The factory default Brightness settings: • BT03/04/05/06 units: Brightness = 32% • BT02 units: Brightness = 40% Follow the steps below to adjust the monitor’s brightness setting: 1.) Press the Left button on the display monitor twice (see Figure 6-1 Adjust Contrast and Brightness on page 6-4 ). 2.) Press the Right button to increase brightness. 3.) Press the Center...
Page 506
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 6 Section 6-7 - Front-End Alignm ent Procedure (Beamformer Calibration) Section 6-7 Front-End Alignment Procedure (Beamformer Calibration) 6-7-1 Introduction Each A/D converter on the BF boards has an offset error (i.e. gives an output code different from 00 Hex with an input voltage of 0V). This offset error may cause artifacts in some modes of operation (2D and TVI) due to inaccuracy in the demodulation process. In order...
Page 507
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 7 6-7-3 Front-End Alignment Procedure 1.) Disconnect all connected probes. 2.) Turn on the Vivid 7. 3.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 4.) Select System 5.) Select DC Offs. Cal to start the DC Offset calibration of the Beam Formers (Front-End Alignment). 6.) Wait for the calibration to take place. Front-End Alignment Description...
Page 508
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 8 Section 6-7 - Front-End Alignm ent Procedure (Beamformer Calibration) 6-7-3 Front-End Alignment Procedure (cont’d) NOTE: The LEDs displayed in the below individual b oard sections are relevant when the system is in the default setting (2D-mode). 8.) Restart Vivid 7. NOTE: Always reboot the scanner after Front-End Alig nment is finished and verify proper operation. NOTE: Do Not select “Cal Reset” af ter performing...
Page 509
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 6 - Service Adjustments 6 - 9 Section 6-8 Direction Lock and Brake Adjustment 6-8-1 Front Caster Brakes Adjustment Procedure 6-8-1-1 Manpower 2 persons 6-8-1-2 Tools • Hexagon key 3/32” inch • Phillips screw driver size 2 6-8-1-3 Preparations 1.) Set Rear Brakes to pr event system from rolling. 2.) Remove Front Bumper , see Front Bumper Removal Procedure on page 8-100 . 6-8-1-4 Front Caster Brakes Adjustment 1.) Use a...
Page 510
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 6 - 10 Section 6-8 - Direction Lock and Brake Adjustment 6-8-2 Direction Lock Adjustment Procedure 6-8-2-1 Manpower 1 person, 30 minutes 6-8-2-2 Tools Open end wrench, 13 mm 6-8-2-3 Preparations 1.) Block Rear Casters. 2.) Push down Direction Lock Pedal. 3.) Turn Casters until they locks in forward pointing position. 6-8-2-4 Direction Lock Adjustment 1.) Loosen the retaining nuts on the brake rods on both Front Casters to allow...