GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 461
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 115 5-7-2 Monitors 5-7-2-1 CRT Monitor • 17-inch High Resolution Monitor with no interlace scan • Available for all Vivid 7 models 5-7-2-2 LCD Monitor (Option) • 21-inch High Resolution Monitor with no interlace scan • Mounted on the Monitor Arm so monitor can be mo ved left or right without moving the Vivid 7. • Introduced for BT’06 (June 2006) as an option (Only available on order...
Page 462
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 116 Section 5-7 - Top Console with Monitor and Operator Panel 5-7-3 Operator Panel The Operator Panel includes an ON/OFF switch, an alphanum eric keyboard, different controls for manipulating the picture quality, and cont rols for use in Measure & Analyze (M&A). On units with software version 3.x.x (or above) the Ro taries have a “Push” function in addition to the ordinary rotary functions. NOTE: Units originally produced...
Page 463
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 117 Section 5-8 External I/O 5-8-1 General Description, External I/O The External I/O is the interface between the scanner and all external devices. Several signal are fed from the Internal I/O module to the External I/O module. The External I/O module contains data buffers and a galvanic isolation section. Different types of signals are fed to co nnectors accessible on the...
Page 464
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 118 Section 5-8 - External I/O 5-8-1 General Description, External I/O (cont’d) The External I/O has the following external conn ectors available at the rear of the system: • Sound out (Right and Left channels - Stereo) •Trig Out • Footswitch • RS-232 Serial Port (Not Used) • Remote Expose 1 • Remote Expose 2 • Modem (Not Used) • Composite Video Out • B/W Video Out • Svideo Out • SVGA Out •USB • Ethernet 5-8-2 Location in the...
Page 465
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 119 5-8-3 Input Signals Figure 5-75 EIO - Overview Table 5-79 Input Signals InputDescriptionConnection from/via/to Footswitch Connector Footswitch is a mechanical sw itch that connect a signal to ground. Internal level is TTL (0-3.3V) Footswitch > External I/O > Internal I/O > PC2IO (BEP) DC Voltages from BEP +5 VDC Used for Optical Couplers on RS232 signals PC2IO (BEP) >...
Page 466
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 120 Section 5-8 - External I/O 5-8-4 Bi-directional Signals 5-8-5 Output Signals Table 5-80 Bi-directional Signals Signal NameDescriptionSignal Path Ethernet Standard 10/100 Base TX Ethernet Ethernet - External I/O - Internal I/O - BEP (Ethernet Connector) COM1 Serial RS-232 port (Not used in Vivid 7) External Device - Exte rnal I/O - Internal I/O - BEP (COM1) USB Standard USB (Universal Serial Bus), (0-5V) USB Device -...
Page 467
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 121 5-8-6 Reset Switch A reset switch is located behind the little hole on the External I/O. NOTE: Use this Reset Key if told so by pr ocedure in this manual or by OLC, ONLY. 5-8-7 Jumpers and Dip-switches None 5-8-8 LEDs NoneFigure 5-76 External I/O Conne ctor Panel with Reset Switch hole RESET SWITCH INSIDE THIS HOLE
Page 468
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 122 Section 5-8 - External I/O 5-8-9 Video Specifications 5-8-9-1 PAL Video Specifications 5-8-9-2 NTSC Video Specifications 5-8-9-3 SVGA Table 5-82 PAL Video Specifications HORIZONTAL TIMING Horizontal Total (HT) 64 µs 944 pixels Horizontal Start (HS) 10.3051 µs 152 pixels Horizontal Active (HA) 52 µs 768 pixels Pixel Clock Freq.14.7 MHz(PAL) VERTICAL TIMING (INTERLACED) Vertical Start (VS)53 Lines / field Active Lines...
Page 469
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 123 Section 5-9 Peripherals 5-9-1 Internal Peripherals • DVD R/W (BT’04 and newer) available from the front of the BEP and the scanner. NOTE: Only CD-R and DVD-R (write once and read many) is supported. CD-R/W and DVD R/W discs are not supported. • CD R/W drive (BT’01, BT’02 and BT’03) NOTE: Only CD-R (write once and read many) is supported. CD-R/W discs are not supported. • DVD...
Page 470
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 124 Section 5-10 - Modem (Option) Section 5-10 Modem (Option) NOTE: The modem has almost never been in active use. 5-10-1 General Description The modem is a standard MultiTech modem that connects to an analog phone line. It is used when connecting to GE via iLink and for InSite Remote Service. The modem is included on all units produced before February 2003. On units produced after February 2003, the modem is optional....