Page 411
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 65 5-3-21-15 Output Signals 5-3-21-16 Power Distribution From DC Power via backplane to RFI. 5-3-21-17 Fuses None. Table 5-54 Output Signals OUTPUTDESCRIPTIONCONNECTION TO TXTRIG_L STARTS THE TRANSMITTER SEQUENCE via Backplane to TX128 BOARDS SYNC_L SIGNAL USED TO SYNCHRONIZE THE BEAM FORMERS via Backplane to BF64 BOARDS TSIG TEST SIGNAL via Backplane to RX128 BOARD AT G C V...
Page 412
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 66 Section 5-3 - a Front-End Processor (FEP) 5-3-21-18 Connectors Figure 5-40 Connectors on the RFI Board Table 5-55 RFI - Connectors CONNECTOR #DESCRIPTIONCOMMENTS P1 ATGC OUT P2 TXTRIG_L OUT P3 TEST TRIG OUT 4TEST 5TEST 6TEST P7PCI CABLE CONNECTOR CONNECTED TO PC2IP P8 MANUFACTORING TEST 9 N/A N/A 10TEST 500TEST 501TEST 502TEST P/N: FC200054- MCD REV: S/N: RFI B C D E F 1 25AJ6 C3364...
Page 413
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 67 5-3-21-19 Jumpers 5-3-21-20 Dip-switches None. Figure 5-41 Jumpers on the RFI Board Table 5-56 RFI - Jumpers JUMPER NAME DESCRIPTIONCOMMENTS SW4WD ENABLE BYPASSED, NOT USED. SW5DISABLE PCI EEPROM P/N: FC200054- MCD REV:S/N: RFI B C D E F 1 25AJ6 C3364 8U71C432C433C429C337TP45R326R412R413R415R417TP46 48 24U55 C365C85C84C426C175C152C15114...
Page 414
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 68 Section 5-3 - a Front-End Processor (FEP) 5-3-21-21 Switch on RFI Figure 5-42 Reset Switch on the RFI Board Table 5-57 RFI - Switches NAMEDESCRIPTIONCOMMENTS SW3RESET SWITCH ONLY FOR USE BY ENGINEERING. (NOT INSTALLED ON ALL BOARDS) P/N: FC200054- MCD REV: S/N: RFI 38 2P9 2829C362RN11C24C126C127C146C150C297C31620 10U62C314L10L9R381R110 R109R106TP32 TP27D22 D12D11D109 TP39TP34 R111R501R502R500 R529R531R530 TP35D501...
Page 415
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 69 5-3-21-22 Test Points Table 5-58 RFI - Test Points TEST POINT # DESCRIPTIONCOMMENTS 1 EMIFB CLOCK 133 MHz 2 Doppler Probe Sense Voltage 3 Tempsense Output 130.8 mV/ oC 4MAX158 INTNOT USED 5MAX158 RDYNOT USED 6-5 V 7CLK40 40 MHz 8 N/A N/A 9 2.5 V voltage Locally generated from +3.3 V 10DSP CLOCK50 MHz 1115 V MONITOR voltage 12EMIFA CLOCK133 MHz 13 DSP_BSOE3_L signal 14 1.4 V...
Page 416
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 70 Section 5-3 - a Front-End Processor (FEP) 33 2.5 V voltage 343.3 V voltage 355 V voltage 36- 5 V voltage 37 N/A 38 N/A 39 -2.0 V voltage 40FAN VENT 1 voltage 41FAN VENT 2 voltage 4210V_Ref_voltage 43 - 5V_ref voltage 44 5V_ref voltage 45 AUDIO RIGHT CHANNEL OUT +/- 1 V 46AUDIO LEFT CHANNEL OUT+/- 1 V 47N/A 48N/A 49N/A 50 CLK20 20 MHz 500...507 GND Table 5-58 RFI - Test Points (cont’d) TEST POINT # DESCRIPTIONCOMMENTS
Page 417
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 71 5-3-21-23 LEDs on the RFI Board Figure 5-43 LEDs on the RFI Board Table 5-59 RFI - LEDs LED #LED COLORDESCRIPTIONNORMAL STATE LD505 - Red Demod Error LD504 - Green Lit in scanning mode Slow tick in idle LD503 - Yellow Transmit data to BEP During Processing LD502 - RedNot Used LD501 - YellowNot Used LD500 - GreenNot Used LD12 - Green 5.0 V status ON LD22 - Green -2.0 V...
Page 418
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 72 Section 5-3 - a Front-End Processor (FEP) 5-3-22 Motherboard (Backplane) 5-3-22-1 General Description The Motherboard is the Front End Card Rack’s Ba ckplane. It distributes DC Voltages from the DC Power to all the boards in the Card Rack and interconnects (distributes signals and data buses between) the different boards. Two versions have been used: • The first version is used for FEP1 • The second version is used for...
Page 419
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 73 Section 5-4 Back-End Processor (BEP) 5-4-1 BEP Overview Contents in this section: • 5-4-2 Introduction on page 5-74 • 5-4-3 Signal Flow and Processing on page 5-74 • 5-4-4 Location of the Back-End Processor (BEP) on page 5-75 • 5-4-5 Back-End Processor (BEP) Block Diagrams on page 5-76 • 5-4-6 Back-End Processor (BEP) Descriptions on page 5-88 • 5-4-7 Internal Storage Devices on...
Page 420
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 74 Section 5-4 - Back-End Processor (BEP) 5-4-2 Introduction The Back-End Processor is a computer designed spec ially for the use in the Vivid 7 ultrasound scanners made by GE. The following BEP versions ha ve been used for Vivid 7: • BEP1 was introduced at the in troduction of the Vivid 7 ultrasound scanners in 2001. • BEP2 was introduced September 2002 as a replacement for BEP1, both for new ultrasound scanners from...