GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 55 5-3-19 Spectrum Doppler Pr ocessor Board, SDP (cont’d) Table 5-45 LEDs on the SDP Board LED #LED COLORDESCRIPTIONNORMAL STATE Upper Batch of LEDs LD1 - Red DSP1 Status OFF. ON if errors/warnings detected. LD2 - Green Board OK ON OFF if error detected. LD3 - Yellow 0DSP1 Local SW RunningFLASHING LD4 - Yellow 1DSP1 Ready for DataON when scanning, else OFF LD5 - Yellow 2DSP1...
Page 402
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 56 Section 5-3 - a Front-End Processor (FEP) 5-3-20 Image Port (IMP) Board NOTE: Image Port is called “IMPORT” in some documents. NOTE: The Image Port Board is only used in FEP1. In units with FEP2, the RFI board replaces the functionality of this board, see: 5-3-21 Radio Frequency Interface Board, RFI on page 5-61 . NOTE: The Image Port Board is available in two versions for Vivid 7. - IMP2, Part Number FB 200991, can be...
Page 403
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 57 5-3-20 Image Port (I MP) Board (cont’d) 5-3-20-2 Location in the Unit Figure 5-36 IMP Board - Location in Unit IMP2 BoardFront of Scanner
Page 404
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 58 Section 5-3 - a Front-End Processor (FEP) 5-3-20 Image Port (I MP) Board (cont’d) 5-3-20-3 Inputs 5-3-20-4 Bi-directional Signals 5-3-20-5 Outputs • Data to the Back-End Processor via PC2IP bus. • Control signals and parameters via VME bus. 5-3-20-6 Power Distribution 5-3-20-7 Fuses None Table 5-46 Video and PipeLink Input Data INPUTDESCRIPTIONCONNECTION FROM: PipeLink Live scan data SDP Video Composite Video or Super...
Page 405
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 59 5-3-20 Image Port (I MP) Board (cont’d) 5-3-20-8 Jumpers and Dip-switches Figure 5-37 IMP, Jumpers and Switch SW4 SW2 12 910 SW12 SW9 SW10 SW8 SW3 SW5 SW13
Page 406
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 60 Section 5-3 - a Front-End Processor (FEP) 5-3-20 Image Port (I MP) Board (cont’d) 5-3-20-9 LEDs All the LEDs should be OFF in Freeze (LD1 dimmed red light). Figure 5-38 IMP2, LEDs Table 5-49 IMP, LEDs LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Red CPU self test failure or reset Normally OFF (dimmed red light) LD2 - Yellow CPU addressing local bus ON during scanning OFF in Freeze LD3 - Yellow Lit when firmware...
Page 407
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 61 5-3-21 Radio Frequency Interface Board, RFI 5-3-21-1 General Description The RFI board was introduced into the manufacturin g in the 2nd half of 2003 together with the Front- End Processor version 2 (FEP2). The functionality of the following boards are either implem ented on the RFI board or on the BEP: • Front-End Controller board • RF & Tissue board • Spectrum Doppler...
Page 408
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 62 Section 5-3 - a Front-End Processor (FEP) 5-3-21-4 Data Buffer After RF filtering, the data is written into different sliding ring buff ers, dedicated to the different types of data. While data is written into the buffers sample by sample in vector s, multiple samples from the same range (depth) can be read out. Both input addressing (start and length) and output addressing (through output events) are controlled by...
Page 409
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 63 5-3-21-9 Probe Management • Control the selection of a probe connector • Sense the Probe type • Sense the Scan Plane Angle on MPTE and PAMPTE probes • Sense Probe Temperature • Turn off the probe if temperature exceed maximum temperature. The temperature references are set to 41.3 oC and 43.0 oC. • Turn off the probe if temperature reading falls below minimum temperature The...
Page 410
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 64 Section 5-3 - a Front-End Processor (FEP) 5-3-21-12 DC Voltages 5-3-21-13 Input Signals 5-3-21-14 Bi-directional Signals Table 5-50 DC Input Voltages INPUTDESCRIPTIONCONNECTED TO + 3.3 Vd DC output voltage, distributed via Motherboard (Backplane). The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane + 5 Vd DC output voltage, distributed via Motherboard (Backplane)....