Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 391

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 45
5-3-17 Digital Signal Processors  Subsystem In Units with FEP1
NOTE: The Digital Signal Processors Subsystem, as  described here, is only used in Front-End Rack 
version 1 (FEP1). 
In units with Front- End Rack version 2 (FEP2), the RFI boa rd plus software on the BEP replaces 
the functionality of these boards, see:  5-3-21 Radio Frequency Interface Board, RFI on 
page 5-61 ....

Page 392

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 46 Section 5-3 - a  Front-End Processor (FEP)
5-3-17-2 Location in the Unit • Color Flow Processing is done by  software on the Back-End Processor.
• Tissue processing and Doppler processing are done by the cards listed in  Table 5-35 on page 5-46.
5-3-17-3 Input Signals • Digitized ultrasound signals from the Beam Formers
• S-Video/C-Video replay from  VCR (via Internal I/O)
5-3-17-4 Bi-directional Signals
5-3-17-4-1 PC2IP Bus...

Page 393

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 47
5-3-18 RF & Tissue Processor Board, RFT
NOTE: The RF & Tissue Processor Board (RFT) is only used in Front-End Rack version 1.  In units with Front-End Rack vers ion 2, the RFI board replaces the functionality of this board, 
see:  5-3-21 Radio Frequency Interface Board, RFI on page 5-61 .
5-3-18-1 General Description The RF & Tissue Processor board receives digitize d ultrasound...

Page 394

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 48 Section 5-3 - a  Front-End Processor (FEP)
5-3-18 RF & Tissue Processor Board, RFT (cont’d)
5-3-18-1-5 Control The RFT board has a local Digital Signal Processor (DSP) with an external EPROM.
The DSP performs multiple control tasks:
1.) It handles communication with the BEP over the VME bus.
2.) It updates filter coefficients and  other parameters in the data path.
3.) When output events from the Fron t-End Controller is...

Page 395

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 49
5-3-18 RF & Tissue Processor Board, RFT (cont’d)
5-3-18-3 Inputs
5-3-18-3-1 Input Signals 
5-3-18-4 Bi-directional Signals
 
5-3-18-5 Outputs
5-3-18-6 DC Voltages
 
Table 5-36    Input Signals 
INPUTDESCRIPTIONCONNECTION FROM
MLA 0 &  MLA 1 Input data stream
VME Bus Data communication BEP
Front-End Bus Information about the type of data and the data sample.  FEC
Table 5-37...

Page 396

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 50 Section 5-3 - a  Front-End Processor (FEP)
5-3-18 RF & Tissue Processor Board, RFT (cont’d)
5-3-18-7 Fuses and Dip-switches None
5-3-18-8 LEDs and Jumpers
 
Figure 5-33   RFT, LEDs and Jumpers
Table 5-40    LEDs on the RFT Board
LED #LED COLORDESCRIPTIONNORMAL STATE
LD8  - Red Board Failure OFF
LD7  - Green Board OK ON
LD6  - Yellow DSP Runs BLINKING
LD5 - YellowData Out on PipeLinkON During Data Transfers to PipeLink
LD4 -...

Page 397

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 51
5-3-19 Spectrum Doppler  Processor Board, SDP
NOTE: The Spectrum Doppler Processor Board is only used in Front-End Rack version 1. 
In units with Front-End Rack ve rsion 2, the RFI board plus software on the BEP replaces the 
functionality of this board, see:  5-3-21 Radio Frequency Interfac e Board, RFI on page 5-61.
5-3-19-1 General Description • This board performs Spectrum...

Page 398

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 52 Section 5-3 - a  Front-End Processor (FEP)
5-3-19 Spectrum Doppler Pr ocessor Board, SDP (cont’d)
5-3-19-7 Location in the Unit
5-3-19-8 Input Signals
 
5-3-19-9 Bi-directional Signals
 
Figure 5-34   SPD Board - Location in Unit
Table 5-41    Input signals 
INPUTDESCRIPTIONCONNECTION FROM:
PipeLink Doppler Signals extracted RFT Board
Serial Trace Data Analog Traces Patient I/O via Internal I/O
Table 5-42    Bi-directional...

Page 399

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 53
5-3-19 Spectrum Doppler Pr ocessor Board, SDP (cont’d)
5-3-19-10 Outputs
 
5-3-19-11 DC Voltages
 
5-3-19-12 Fuses and Dip-switches
None Table 5-43    Output Signals 
INPUTDESCRIPTIONCONNECTION TO:
PipeLink Doppler Signals  extracted IMP Board
Audio Output Analog Doppler (I and Q) signals to Operator Panel 
speakers and VCR Internal I/O
VME Bus Data communication IMP
Table 5-44...

Page 400

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 54 Section 5-3 - a  Front-End Processor (FEP)
5-3-19 Spectrum Doppler Pr ocessor Board, SDP (cont’d)
5-3-19-13 LEDs and Jumpers
Figure 5-35   SDP, LEDs and Jumpers
LD8
LD9
LD10
LD11
LD12
SW3
LD1
LD2
LD3
LD4
LD5
LD6
LD7
P4
12
3940
P6
12
1112SW1
P5