Page 341
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 77 4-3-20-7 Check Cineloop from Video Prerequisite A Video Tape in the VCR with at least one recorded examination. Procedure 4-3-20-8 Turn OFF Power to Vivid 7 See 4-2-3 Power Shut Down on page 4-7 . Table 4-40 VCR Cineloop Checks StepTask Expected Result(s) 1. When in video play back (of an examination), press FREEZE . The last part of the video is stored in the computer’s memory as a cine...
Page 342
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 78 Section 4-3 - Functional Checks 4-3-21 Mechanical Functions Checks 4-3-21-1 Alphanumeric Keyboard and Display Platform (Console) Checks Figure 4-41 Console with handles Table 4-41 Display Platform Maneuverability Checks STEPTASK NOTE(S) 1. Pull handle up under the platform towards you to release horizontal lock It is possible to maneuver the Display Platform horizontally at present level of height 2. Pull handle...
Page 343
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 79 4-3-21-2 LCD Screen Movement Figure 4-42 LCD Display Platform Maneuverability Checks Table 4-42 LCD Display Platform Maneuverability Checks STEPTASKNOTE(S) 1 MOVE UP/DOWN Use both hands, one at the top of the LCD and one at the bottom, to adjust the LCD up and down, 2 TILT FORWARD/BACK Use both hands, one at the top of the LCD and one at the bottom, to rotate the screen forward and back....
Page 344
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 80 Section 4-3 - Functional Checks 4-3-21-3 Brakes and Direction Lock Checks Follow this procedure to verify that Brak es and Direction Locks function as intended. Table 4-43 Brakes and Direction Lock Checks StepTaskExpected Result(s) 1. Press on pedal no. 2 To engage the brakes in full lock 2. Press on pedal no.1 To release the brake 3. Press on pedal no.1 To engage swivel lock. You may need to move the system around a...
Page 345
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 81 Table 4-44 Brakes and Direction Lock Checks STEPTASKFUNCTIONCOMMENTS 1. Press on pedal no. 2 To engage the brakes in full lockTest Brake effect, if not proper, see section 6-8-1 on page 6-9 for adjustment 2. Press on pedal no.1 To release the brake 3. Press on pedal no.1 To engage swivel lock. You may need to move the system around a little to get the Casters to lock into the...
Page 346
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 82 Section 4-4 - Application Turnover Check List Section 4-4 Application Turnover Check List Complete these checks before returning the scanner to customer for use: 4-4-1 Software Configuration Checks Section 4-5 Power Supply 4-5-1 Power Supply Test Procedure 4-5-1-1 Introduction There is no need to do any special tests on the Power Supplies if there don’t seems to be a problem that may be related to the Power Supplies....
Page 347
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 1 Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Vivid 7’s system concepts, component arran gement, and subsystem functions. It also describes the Power Distribution System and the Common Service Interface. 5-1-2 Contents in Chapter 5 Table 5-1 Contents in Chapter 5 SECTIONDESCRIPTION PAGE NUMBER 5-1 Overview...
Page 348
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 2 Section 5-2 - General Information Section 5-2 General Information 5-2-1 Introduction The Vivid 7 ultrasound scanner can be used with both phased array and linear array ultrasound probes and Doppler (Pedof) probes. The system can be used for: • 2D Gray Scale and 2D Color Flow Imaging • M-Mode Gray Scale Imaging • Color M-Mode • Doppler • 4D (Realtime 3D, introduced 2004) • Different combinations of the above modes Vivid 7 is...
Page 349
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 3 5-2-2 Signal Flow The TX board(s) in the Front-End Processor, generat es the strong bursts transmitted by the probes as ultrasound into the body. The Transmit bursts are rout ed from the TX board via the XD bus to the Relay board where the ultrasound probes are connected. Weak ultrasound echoes from blood cells and body st ructure are received by the probes and routed via the...
Page 350
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 4 Section 5-2 - General Information 5-2-6 Block Diagram - Units with Fr ont-End Card Cage, version 1 (FEP1) Figure 5-2 Vivid 7 Simple Block Diagram - Un its with Front-End Card Cage version 1 (FEP1) FRONT END PROCESSOR PROBES XDBUS DOPPLER ATX POWER SUPPLYUPS AC POWER OP I/O PA N EL17 Monitor USER INTERFACE SONY COLOR PRINTERUP-2900MDPC2IO PCVIC PC MOTHERBOARD POWER ON/OFF (RESET*) Ethernet PC2IP AGP VIDEO EXTERNAL I/ O...