GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 331
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 67 4-3-16 Probe/Connectors Check NOTE: Probes can be connected at any time, whether the unit is ON or OFF CAUTIONTake the following precautions with the probe cables: - Keep away from the wheels - Do not bend - Do not cross cables between probes Table 4-31 Probe and Connectors Checks StepTaskExpected Result(s) 1. Press Probe on the Operator Panel. A list of the connected probes will pop...
Page 332
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 68 Section 4-3 - Functional Checks 4-3-17 ECG Check 4-3-17-1 Introduction The ECG capability on this unit, is intended as use as a trigger for measurements, but can also be viewed on the screen. 4-3-17-2 Parts needed • ECG Harness, P/N:16L0026 + P/N:16L0028 • ECG Pads, (3 pc) or • ECG simulator 4-3-17-3 Preparations None 4-3-17-4 ECG Check Table 4-32 ECG Check StepTaskExpected Result(s) 1. Connect the ECG harness to...
Page 333
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 69 4-3-18 Cineloop Check 4-3-18-1 Introduction A cineloop is a sequence of images recorded over a certain time frame. When using ECG the time frame can be adjusted to cover one or more heart cycles. When frozen the System automatically displays the cineloop boundary markers on either side of the last detected heart cycle (Figure 4-36). 4-3-18-2 Preparation Use a phantom when doing these tests....
Page 334
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 70 Section 4-3 - Functional Checks 4-3-18-2 Preparation (cont’d) Figure 4-37 Cineloop controls OffOff/OnO n Mode Dependant Keys Left marker Right marker Cycle select Number of cycles First Cycle Last Cycle Sync Cineloop Select all Trackball Scroll (in freeze) Cine speed (in replay) Freeze Image Store Start/stop cineloop
Page 335
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 71 4-3-18-3 Check Left Marker, Right Marker, Cycle Select and Number of Cycles Table 4-33 Cineloop Controls Checks StepTaskExpected Result(s) 1. Adjust LEFT MARKER and RIGHT MARKER Moves the Left and Right Markers to expand or trim the cineloop boundaries. 2. Adjust CYCLE SELECT Moves the cineloop boundaries from heart cycle to heart cycle. 3. Adjust NUMBER OF CYCLE Increase or...
Page 336
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 72 Section 4-3 - Functional Checks 4-3-18-4 Check First, Last Cycle, Sync and Select all 4-3-18-5 Adjust Scroll and Cine Speed Table 4-34 Cineloop Controls Checks StepTaskExpected Result(s) 1. Press FIRST and LAST CYCLE Select either the first or the last heart cycle. 2. Press SYNC Phase synchronize multiple loops. 3. Press SELECT ALL Moves the cineloop boundaries to include all heart cycles. Table 4-35...
Page 337
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 73 4-3-19 Back-End Processor Checks • If all the previous tests have been passed succe ssfully, the Back-End Processor is most likely OK. • If the system seems to be operat ing erratically, please refer to Chapter 7 - Diagnostics/ Troubleshooting . Figure 4-38 The Test pane
Page 338
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 74 Section 4-3 - Functional Checks 4-3-20 Peripheral Checks 4-3-20-1 Video Printers Checks There may be both a black & white video printer and a color video printer on-board the system. The Video Printers are controlled from the PRINT and ALT. keys on the scanner’s Operator Panel. The factory default is PRINT for the B/W Video Printer and ALT. for the Color Video Printer. 4-3-20-2 DVR Checks Please see chapter 5 in...
Page 339
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 75 4-3-20-3 VCR Checks 4-3-20-4 VCR Counter Check When adding new examinations to a Video cassette, you can set the VCR Counter without rewinding the Video Cassette to the start. The procedure below explains how to check this function. Prerequisite: The VCR Counter number when the Video Cassette was Ejected last time. Figure 4-40 VCR Controls on the Operator Panel Table 4-37 VCR Counter...
Page 340
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 76 Section 4-3 - Functional Checks 4-3-20-5 VCR Record Check 4-3-20-6 Check Rewind, Fast Forward and Replay Video Prerequisite: A Video Tape in the VCR with at least one recorded examination. Table 4-38 VCR Checks StepTask Expected Result(s) 1.When scanning, press REC/PAUSE on the Operator Panel. The VCR starts recording. A red dot is displayed in the screen’s VCR status area on the Title bar. It indicates that...