Page 311
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 47 4-3-8-7 Check Sample Volume and Tissue Priority Table 4-16 Color Flow Mode: Check Variance, Color Maps and Cineloop StepTask Expected Result(s) 9. Adjust Variance Controls the amount of variance data added to a color display. Variance enables computer-aided detection of non- luminary flow (e.g. jets or regurgitation). Variance is available in live and cine replay. 10. Select Color Maps...
Page 312
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 48 Section 4-3 - Functional Checks 4-3-8-8 Check Frequency, Radial Av eraging, Lateral Averaging, Power 4-3-8-9 Adjust ROI Table 4-17 Color-Mode Soft Menu Controls Check StepTaskExpected Result(s) 1. Adjust Frequency Enables the adjustment of the transmission frequency to control the sensitivity or the level of penetration. The selected frequency is displayed in the status window. Adjusting Frequency may affect...
Page 313
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 49 4-3-9 Doppler Mode Checks 4-3-9-1 Introduction Doppler is used to measure velocity (most often in blood). Doppler mode can be done with a special pencil probe or with an ordinary probe. By using an ordinary probe, you can first bring up a 2D picture for navigation purpose and then add Doppler. 4-3-9-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the...
Page 314
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 50 Section 4-3 - Functional Checks 4-3-9-2 Preparations (cont’d) Figure 4-32 Doppler Mode Controls Gain * Mode Dependant Keys Horizontal Sweep * Velocity Range Baseline * LVR Invert * LPRF (in PW mode only) Color Maps * Soft menu Sample Vol. (PW) Compress * Reject * Frequency Frame Rate Angle Corr. * Power CW Freeze PW Zoom Display zoom HR zoom Controls marked with * (asterisk) are also available in freeze and...
Page 315
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 51 4-3-9-3 Check Horizontal Sweep, Velocity Range, Baseline and Low Velocity Reject Table 4-19 Doppler Mode Controls Checks StepTaskExpected Result(s) 1. Adjust HORIZONTAL SWEEP Adjusts the horizontal refresh rate of the Doppler area of the display. Horizontal sweep is available in live and cine replay. 2. Adjust VELOCITY RANGE Enables the vertical scale of the Doppler spectrum and the...
Page 316
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 52 Section 4-3 - Functional Checks 4-3-9-4 Check Invert, LPRF (in PW only) and Color Maps Table 4-20 Doppler Mode Controls Checks StepTaskExpected Result(s) 5. Press INVERT Enables the Doppler spectrum to be flipped 180 degrees, so that negative velocities are displayed above the baseline and positiv e velocities below the baseline. Invert is available in live and cine replay. 6. Press LPRF In PW, adjusts the...
Page 317
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 53 4-3-9-5 Check Sample Volume, Co mpress, Reject and Frequency Table 4-21 PW/CW Doppler Mode: Check Sample Volume, Compress, Reject and Frequency StepTask Expected Result(s) 1. Adjust SAMPLE VOLUME (SV) In PW mode, set the longitudinal size of the region to be sampled for measurement. Adjusting Sample volume may affect the PRF (Nyquist limit) settings. SV does not apply to CW mode,...
Page 318
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 54 Section 4-3 - Functional Checks 4-3-9-6 Check Frame Rate, Angle Correction and Power Table 4-22 PW/CW Doppler Mode: Che ck Frame Rate, Angle Correction and Power StepTask Expected Result(s) 1. Adjust FRAME RATE Adjusts frame rate. The relative setting of the frame rate is displayed in the status window (under 2D). 2. Adjust ANGLE CORRECTION Enables correction of the Doppler velocity scale by defining the angle...
Page 319
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 55 4-3-10 4D Volume Rendering Test Do a 4D functional check. The 4D and Multi-plane Imaging manual explains how to use the 4D function. The manual is include d on the Vivid 7 Manual CD. Follow these steps to test the basic 4D functionality: 1.) Select the 3V probe. 2.) Create an examination. 3.) Press 4D on the Operator Panel. The 4D Volu me Rendering screen is displayed. 4.) Adjust the Width...
Page 320
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 56 Section 4-3 - Functional Checks 4-3-10 4D Volume Rendering Test (cont’d) 5.) Using the trackball rotate and translate the volume rendering to display the structure of interest as described in the section below. Press Angle to get a pre-defined orientation before translating and/ or rotating the volume rendering. If necessary, press Clear to reset the orient ation to the default position. 6.) Optimize the volume rendering...