GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 291
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 27 4-2-13 Moving and Transpor ting the Vivid 7 (cont’d) • Place the probes in their carrying case. • Lock Operator Panel (Control Console) in place. The Operator Panel (Control Console) must be locked into the center (x/y) position as indicated on the label on the Front Handle, before transporting the system. • Eject any Magneto Optical disk, CD or DVD from their drive. • Secure the unit in an...
Page 292
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 28 Section 4-3 - Functional Checks Section 4-3 Functional Checks 4-3-1 Overview The intention with this section is to keep all functional checks together in one place. 4-3-2 Contents in this Section 4-3-3 Preparation Turn on power to Vivid 7, see 4-2-2 - Power ON/ Boot UP on page 4-3. Table 4-5 Functional Checks overview SUBSECTIONDESCRIPTIONPAGE NUMBER 4-3-1 Overview 4-28 4-3-2 Contents in this Section 4-28 4-3-3...
Page 293
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 29 4-3-4 Basic Controls 4-3-4-1 Introduction Most of the mode dependant controls can be foun d as Mode Dependant Keys (Mode Dependant Rotary Knobs and Mode Dependant Push Buttons) and Trackball. The functions available on the Mode Dependant Keys and Trackball vary according to the scan mode and position that is currently active. 4-3-4-2 Mode Dependant Rotary Knobs Vivid 7 has four mode...
Page 294
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 30 Section 4-3 - Functional Checks 4-3-4-5 Soft Menu Control (4-way Rocker) The 4-way Rocker works together with the Soft menu on the screen. • The vertical arrows (up/down) are used to do selections from the Soft Menu. • The horizontal arrows (left/right) are used to do adjustments on the selected function. 4-3-4-6 Trackball Area Different functions can be assigned to the trackball depending of the current active mode. The...
Page 295
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 31 4-3-5 Performance Tests 4-3-5-1 Recommended Test Phantoms GE Healthcare recommends the RMI 430GS phantom. It is the most current phantom recommended to our field service personnel and provides the necessary targets and extended life necessary for consistent system testing. Figure 4-26 Performance tests 2 432 1 4 6 8 10 12 14 16 RMI 403GS Lateral DistanceMeasurement Functional Resolution...
Page 296
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 32 Section 4-3 - Functional Checks 4-3-6 2D Mode (B mode) Checks 4-3-6-1 Introduction The 2D Mode is the system’s default mode. 4-3-6-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the scanner’s left-most active probe connector. -See Probe Connection on page 3-21 for info about connecting the probes. - See section Section 9-26 Probes on page 9-80 for available probes. 2.) Turn ON the...
Page 297
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 33 4-3-6-2 Preparations (cont’d) 4.) Press 2D on the Operator Panel to access 2D Mode. 5.) These Image Controls are used to optimize the 2D picture: -Use Gain and TGC controls to optimize the ov erall image together with the Power control. -Use Depth to adjust the range to be imaged. -Use Focus to center the focal point(s) around the region of interest. -Use Frequency (move to higher...
Page 298
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 34 Section 4-3 - Functional Checks 4-3-6-3 Check Width, Focus, Framerate, Frequency These 2D Mode Functions are located on Mode Dependant Rotaries. The results of these adjustments must be verified on the 2D sector on the screen. Table 4-6 2D Mode: Check Width, Focus, Framerate and Frequency StepTaskExpected Result 1.Turn the rotary labeled WIDTH left and right. Controls the size or angular width of the 2D image...
Page 299
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 35 4-3-6-4 Check Up/Down, Left/Right, B Color Maps and Cineloop These 2D Mode Functions are located on Mode Dependant Keys (Push Buttons) The results of these adjustments must be verified on the 2D sector on the screen. Table 4-7 2D Mode: Check Up/Down, Left/Right, B Color Maps and Cineloop StepTaskExpected Result 1. Press once on the Mode Dependant button labeled UP/DOWN Enables the 2D...
Page 300
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 36 Section 4-3 - Functional Checks 4-3-6-5 Check Gain, TGC and Depth Table 4-8 2D Mode: Check Gain, TGC and Depth in 2D Mode StepTaskExpected Result 1. Adjust TISSUE GAIN Increases the overall gain equally for all depth when rotated clockwise 2. Adjust Time Gain Compensation (TGC) Compensates for depth-related attenuation in an image. The upper sliders affect the near field. The lower sliders (nearest the operator)...