Page 241

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 117
3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
8.) Select the  DATAFLOW
 tab to view the Dataflow Services window.
9.) Select  Remote Archive - Remote HD  on the drop-down menu to see  the contents of that dataflow. 
Figure 3-126   Dataflow Services Window
Figure 3-127   Obtaining Dataflow Contents Overview
Dataflow 
selection 
area
Dataflow 
Services 
area...

Page 242

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 118 Section 3-11 - Connectivity Setup - Software v2.x
3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) Figure 3-128   Remote Archive - Remote HD Dataflow Contents Overview
Remote Archive -
Remote HD 

Page 243

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 119
3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
10.) Select the  SERVICES
 tab (located near the top of the window).
11.) Select the EchoPAC PC Workstation you want to connect to, from the “Name” drop-down menu. In  this example, it is called “EchoPAC7-000000”.  Figure 3-129   Dataflow Services
Figure 3-130   Dataflow Services, Choices and Setup
Name drop-down...

Page 244

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 120 Section 3-11 - Connectivity Setup - Software v2.x
3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) 12.) Verify that the IP-address is correct, in dicating that you have selected the right unit.
If the IP number is wrong, it means that yo u are addressing the wrong device (computer). 
Correct the address and do a new verification before you continue. 
 
Figure 3-131   Verify/change IP...

Page 245

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 121
3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
13.) In the list of “Services”, select “Remote Archive - Remote Storage”. 
14.) In the list of  Services select Remote Archive -  Remote Storage. 
15.) Change the  Corresponding sql.ini name  to reflect the name of the ECHOPAC7-nnnnnn 
Workstation. The  Corresponding sql.ini name  is a combination of the host...

Page 246

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 122 Section 3-11 - Connectivity Setup - Software v2.x
3-11-9 HL7 Communication Se tup for Software v2.x
3-11-9-1 Introduction
The procedures below tell you how to set up the Vivid 7 workstation so it can connect to the Hospital 
Information System (HIS), via th e Vivid HL7 Gateway. By connec ting to the HIS, demographic 
information i.e Patient ID, Name, Gender etc. can  be pulled from the HIS, reducing “double-work” and...

Page 247

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 123
3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
Configure the Search Criterias
NOTE: This configuration is required. 1.) Select the Search Criterias button.
2.) Select the criteria Modality and select the Remove button.
3.) Press the OK button.
Configure Search Criterias:  Constrain Query by AE Title
NOTE: This configuration is optional. It should only be  performed if the site...

Page 248

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 124 Section 3-11 - Connectivity Setup - Software v2.x
3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
Test Connectivity with the Vivid HL7 Gateway
The purpose of this test is to verify that the  Vivid 7 has connectivity with the Vivid HL7 Gateway.
1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality  Worklist service.
2.) Expand the Gateway destination device....

Page 249

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 125
3-11-9-3      Configuration of  Modality Worklist Services (cont’d)
Test Connectivity with the Vivid HL7 Gateway
1.) Start a new study and acquire an  image, or use an existing study.
2.) In the Measurements mode, make some of the measurements.
3.) In the Report screen, cr eate and store a report.
4.) In the Patient screen, select the study that you  created the report in from the Examination List....

Page 250

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 126 Section 3-11 - Connectivity Setup - Software v2.x
3-11-10 Set Up Connection to a DICOM Se rver in a Network - Software v2.x
No special service tools are required to do this set-up.
3-11-10-1 Preparations Obtain the following information  from the network administrator:
• DICOM server’s IP address, subnet mask (and gateway, if required).
• Server’s Port Number.
• Server’s AE Title.
• IP address for the Vivid 7.
3-11-10-2...