Page 191
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 67 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) DICOM SR If SR is enabled, all M&A that is supported by DICO M is sent to the DICOM server (in a separate job). Allow SR Private Data (introduced with software versions v6.0, v5.2 and v4.2.) By default, this function is disabled. Allow SR Private Data is intended for receivers that can handle “Private Data”, like the CA-1000. In the...
Page 192
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 68 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) Retry • Max # - The number of times the software will try to connect if the connection doesn’t succeed the first time(s). • Interval - The interval in seconds between each attempt to connect. Smiley Select the Smiley button to ping th e server. If the network connection is OK, the...
Page 193
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 69 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) MPPS If you want to enable MPPS, check the check-box and select the button “MPPS”. In the dialog, enter IP address, Name, AE Title and Port Number. Time-out Adjust Time-out setting Introduced with software versions v6.0, v5.2 and v4.2: The retry settings can be used to make jobs retry on bad networks. There is no need to set retries for...
Page 194
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 70 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15 Export Configuration The destination for Export of patient records to Exce l and MPEG must be configured prior to use. This is done from the Dataflow screen. 3-9-15-1 Setup on the Remote Share Required setup on the remote share (remote server): 1.) Add user and password.Predefined user: E1c2h3o4C5l6i7e8n9t Predefined password: u1l3t5r7a...
Page 195
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 71 3-9-15-2 Display the Dataflow Screen Follow these steps to display the Dataflow Screen: 1.) Press CONFIG (F2) and log on as administrator. 2.) Select Connectivity > Dataflow . The Dataflow sheet is displayed, see Figure 3-71 The Dataflow sheet on page 3-71 . 3.) Select the dataflow Misc. Export in the Name pull-down menu. Figure 3-71 The Dataflow sheet Software version v4.x/v5.x Software...
Page 196
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 72 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15-3 Export to Excel Configuration 1.) Select the Excel storage device in the Selected devices pane and press Properties . The Excel properties window is displayed. 2.) Select a removable media or a network volume re mote path as the destination in the Destination pull-down menu. NOTE: Remote paths for network volumes must be entered once...
Page 197
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 73 3-9-15-4 Export to MPEGVue Configuration 1.) Select the MPEGVue device in the Selected devices pane and press Properties . The MPEGVue properties window is displayed. 2.) Select a removable media or a network volu me remote path as the destination in the Destination Pull-down menu . NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they can be...
Page 198
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 74 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-15-5 eVue Configuration 1.) Press CONFIG (F2) and log on as administrator. 2.) Select the Connectivity category and Dataflow subgroup. The Dataflow sheet is displayed. 3.) Select the dataflow Local Archive - Int. HD/eVue in the Name pull-down menu. 4.) Select the eVue device in the Selected devices pane and press Properties. The eVue...
Page 199
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 75 3-9-16 Create a New Dataflow 3-9-16-1 Overview It is possible to make new dataflows by combining the predefined settings. The table below describes the legal combination of inputs and outputs in a dataflow. 3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (Introduced for software v4.0) To create a new dataflow with Work list, Database and DICOM Storage, do the following in...
Page 200
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 76 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (cont’d) 3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).Go to the IP Address field, scroll down and select . This will allow you to edit or add IP Addresses (New Destinations/devices). If the correct IP Address is...