Chapter 3 - System Setup 3 - 67
3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
DICOM SR
If SR is enabled, all M&A that is supported by DICO M is sent to the DICOM server (in a separate job).
Allow SR Private Data (introduced with software versions v6.0, v5.2 and v4.2.)
By default, this function is disabled.
Allow SR Private Data  is intended for receivers that can handle “Private Data”, like the CA-1000.
In the...

Page 192

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 68 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
Retry
• Max # - The number of times the  software will try to connect if the connection doesn’t succeed the 
first time(s).
• Interval - The interval in seconds between each attempt to connect.
Smiley
Select the Smiley button to ping th e server. If the network connection  is OK, the...

Page 193

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 69
3-9-14-6      DICOM Storage Properties Adjustments (cont’d)
MPPS
If you want to enable MPPS, check  the check-box and select the button “MPPS”. In the dialog, enter IP 
address, Name, AE Title and Port Number.
Time-out
Adjust Time-out setting
Introduced with software versions v6.0, v5.2 and v4.2:  The retry settings can be used to make jobs 
retry on bad networks. There is no need to set retries for...

Page 194

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 70 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-15 Export Configuration
The destination for Export of patient records to Exce l and MPEG must be configured prior to use. This 
is done from the  Dataflow screen.
3-9-15-1 Setup on the Remote Share Required setup on the remote share (remote server):
1.) Add user and password.Predefined user:  E1c2h3o4C5l6i7e8n9t
Predefined password:  u1l3t5r7a...

Page 195

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 71
3-9-15-2 Display the Dataflow Screen Follow these steps to display the Dataflow Screen:
1.) Press CONFIG (F2)  and log on as administrator.
2.) Select  Connectivity  > Dataflow .
The  Dataflow sheet  is displayed, see  Figure 3-71 The Dataflow sheet on page 3-71 .
3.)
Select the dataflow  Misc. Export in the Name pull-down menu.
Figure 3-71   The Dataflow sheet
Software version v4.x/v5.x Software...

Page 196

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 72 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-15-3 Export to Excel Configuration 1.) Select the  Excel storage  device in the Selected devices pane  and press Properties .
The  Excel properties window  is displayed.
2.) Select a removable media or a network volume re mote path as the destination in the Destination 
pull-down menu.
NOTE: Remote paths for network volumes must be entered once...

Page 197

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 73
3-9-15-4 Export to MPEGVue Configuration 1.) Select the  MPEGVue device in the  Selected devices  pane and press  Properties .
The  MPEGVue properties window  is displayed.
2.) Select a removable media or a network volu me remote path as the destination in the Destination 
Pull-down menu .
NOTE: Remote paths of network volumes must be enter ed once in the Remote path field before they 
can be...

Page 198

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 74 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-15-5 eVue Configuration 1.)
Press CONFIG (F2) and log on as administrator.
2.) Select the  Connectivity category and  Dataflow subgroup. The Dataflow sheet is displayed.
3.) Select the dataflow  Local Archive - Int. HD/eVue  in the Name pull-down menu.
4.) Select the  eVue device in the  Selected devices  pane and press Properties. The eVue...

Page 199

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 75
3-9-16 Create a New Dataflow
3-9-16-1 Overview It is possible to make new dataflows by combining  the predefined settings. The table below describes 
the legal combination of inputs and outputs in a dataflow.
3-9-16-2 Create a New Dataflow with Worklist, Database and DICOM Storage (Introduced for software v4.0)
To create a new dataflow with Work list, Database and DICOM Storage,  do the following in...

Page 200

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 76 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-16-2      Create a New Dataflow with Worklist, Database and DICOM Storage (cont’d) 3.) Add the DICOM Storage Output, highlight it, select properties, and name it (i.e.: DR PACS Store).Go to the IP Address field, scroll down and select  . This will allow you to edit or add IP 
Addresses (New Destinations/devices).
If the correct IP Address is...
Start reading GE Vivid 7 Service Manual

Related Manuals for GE Vivid 7 Service Manual

All GE manuals