Page 151

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 27
3-6-7 Switching OFF the Unit
See  4-2-3 Power Shut Down on page 4-7  for a detailed description.
1.) Press the  ON/STANDBY
 button on the top left of the Control Panel to display the Exit dialog 
window.
2.) Select  SHUTDOWN
.
The shutdown process takes a few seconds and is completed when the Operator Panel 
illumination is turned off.
To completely switch off the unit before disconnec ting the Mains...

Page 152

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 28 Section 3-7 - .  Configuration
Section 3-7. 
Configuration
3-7-1 Vivid 7 Configuration 
3-7-1-1 Select System Settings Screen 1.) Press CONFIG
 (F2) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
NOTE: For software version v5.x and earlier the  default password for ADM (or: adm) was left blank. 
Starting with software vers ion v6.x, the password is: ulsadm
2.) Select  System and then select...

Page 153

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 29
3-7-1-3 Date and Time Adjustments Figure 3-19   Date and Time Adjustments
Table 3-12    Date and Time adjustments
STEPTASKEXPECTED RESULT(S)
1 Open the 
System (Configuration) Window, see  4-2-4 Log On to 
the System as ‘ADM’ on page 4-12 .
Select  System , if needed. The 
System Settings  window is displayed.
2 Adjust the  date, see ( a) in  Figure 3-19 . New date  is displayed
3  Adjust the  time,...

Page 154

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 30 Section 3-7 - .  Configuration
3-7-1-4 Language Selection
 
Figure 3-20   Language Selection
Table 3-13    Language Adjustments
STEPTASKEXPECTED RESULT(S)
1. Open the Configuration Window, see  4-2-4 Log On to the 
System as ‘ADM’ on page 4-12 . 
Select  System , if needed. The 
System Settings  window is displayed.
2. Use the  Language  drop down dialog, see  Figure 3-20, to select 
your preferred language for the on-screen...

Page 155

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 31
3-7-1-5 Online Manual Language Selection
 
Figure 3-21   Online Manual Language Selection
Table 3-14    Online Manual Language Selection
STEPTASKEXPECTED RESULT(S)
1. Open the Configuration Window, see  4-2-4 Log On to the 
System as ‘ADM’ on page 4-12 . 
Select  SYSTEM
, if needed. The 
System Settings  window is displayed.
2. Use the  Manual Language  drop down dialog, see  Figure 3-21, 
to select...

Page 156

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 32 Section 3-7 - .  Configuration
3-7-1-6 Units of Measure
 
Figure 3-22   Select Units of Measure
Table 3-15    Select Units of Measure
STEPTASKEXPECTED RESULT(S)
1. Open the Configuration Window, see  4-2-4 Log On to the 
System as ‘ADM’ on page 4-12 .
Select  SYSTEM
, if needed. The 
System Settings  window is displayed.
2. Use the drop down dialog to select Metric or US Units. The selected units (Metric or US) will
 be used...

Page 157

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 33
3-7-2 Service Screen Setup
3-7-2-1 Overview The Service Screen gives you access to: 
• Select Video Format to be used by DVD or VCR
• Select DVD or VCR type
• Adjust the Operat or Panel LCD’s Contrast and Backlight  Intensity 
• Alphanumeric Keyboard Setups 
• Printer Setup
3-7-2-2 Open Service Screen 1.) Press CONFIG
 (F2) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 ....

Page 158

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 34 Section 3-7 - .  Configuration
3-7-2-3 Select Video Format, PAL or NTSC This selection must correspond to the Video Stan dard (PAL or NTSC) used by the Video Recorder. 
• From the Video Settings drop-down menu, select the correct video format (NTSC or PAL).
3-7-2-4 Select Video Recorder (DVD or VCR) Type •Use the VCR drop down menu to select the type. See  Approved Internal Peripherals on page 3-
36  for approved video...

Page 159

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 35
3-7-2-6 Add Printer NOTE: This function may be unavailabl e for some software versions and it will not always function due 
to that usually, a special Installation Wizard is  to be used. Please follow instructions in the 
respective printer installation manual for correct printer installation. See  3-7-3 Optional 
Peripherals/Peripheral Connection on page 3-36  for more information.
1.) Select  ADD...

Page 160

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 36 Section 3-7 - .  Configuration
3-7-3 Optional Peripherals/Peripheral Connection
3-7-3-1 Approved Internal Peripherals This list covers the internal peripherals approved for Vivid 7:
• Printer, Black & White, Video, SONY.
• Printer, Black & White, Digital, SONY (need s software version v3.0 or above and BEP2). 
• Printer, Monochrome, Dig ital, SONY UP-D897SYN
• Printer, Monochrome, Digital, SO NY UP-D895MD (needs software...