Page 151
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 27 3-6-7 Switching OFF the Unit See 4-2-3 Power Shut Down on page 4-7 for a detailed description. 1.) Press the ON/STANDBY button on the top left of the Control Panel to display the Exit dialog window. 2.) Select SHUTDOWN . The shutdown process takes a few seconds and is completed when the Operator Panel illumination is turned off. To completely switch off the unit before disconnec ting the Mains...
Page 152
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 28 Section 3-7 - . Configuration Section 3-7. Configuration 3-7-1 Vivid 7 Configuration 3-7-1-1 Select System Settings Screen 1.) Press CONFIG (F2) and log on as ADM, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . NOTE: For software version v5.x and earlier the default password for ADM (or: adm) was left blank. Starting with software vers ion v6.x, the password is: ulsadm 2.) Select System and then select...
Page 153
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 29 3-7-1-3 Date and Time Adjustments Figure 3-19 Date and Time Adjustments Table 3-12 Date and Time adjustments STEPTASKEXPECTED RESULT(S) 1 Open the System (Configuration) Window, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Select System , if needed. The System Settings window is displayed. 2 Adjust the date, see ( a) in Figure 3-19 . New date is displayed 3 Adjust the time,...
Page 154
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 30 Section 3-7 - . Configuration 3-7-1-4 Language Selection Figure 3-20 Language Selection Table 3-13 Language Adjustments STEPTASKEXPECTED RESULT(S) 1. Open the Configuration Window, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Select System , if needed. The System Settings window is displayed. 2. Use the Language drop down dialog, see Figure 3-20, to select your preferred language for the on-screen...
Page 155
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 31 3-7-1-5 Online Manual Language Selection Figure 3-21 Online Manual Language Selection Table 3-14 Online Manual Language Selection STEPTASKEXPECTED RESULT(S) 1. Open the Configuration Window, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Select SYSTEM , if needed. The System Settings window is displayed. 2. Use the Manual Language drop down dialog, see Figure 3-21, to select...
Page 156
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 32 Section 3-7 - . Configuration 3-7-1-6 Units of Measure Figure 3-22 Select Units of Measure Table 3-15 Select Units of Measure STEPTASKEXPECTED RESULT(S) 1. Open the Configuration Window, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Select SYSTEM , if needed. The System Settings window is displayed. 2. Use the drop down dialog to select Metric or US Units. The selected units (Metric or US) will be used...
Page 157
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 33 3-7-2 Service Screen Setup 3-7-2-1 Overview The Service Screen gives you access to: • Select Video Format to be used by DVD or VCR • Select DVD or VCR type • Adjust the Operat or Panel LCD’s Contrast and Backlight Intensity • Alphanumeric Keyboard Setups • Printer Setup 3-7-2-2 Open Service Screen 1.) Press CONFIG (F2) and log on as ADM, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 ....
Page 158
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 34 Section 3-7 - . Configuration 3-7-2-3 Select Video Format, PAL or NTSC This selection must correspond to the Video Stan dard (PAL or NTSC) used by the Video Recorder. • From the Video Settings drop-down menu, select the correct video format (NTSC or PAL). 3-7-2-4 Select Video Recorder (DVD or VCR) Type •Use the VCR drop down menu to select the type. See Approved Internal Peripherals on page 3- 36 for approved video...
Page 159
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 35 3-7-2-6 Add Printer NOTE: This function may be unavailabl e for some software versions and it will not always function due to that usually, a special Installation Wizard is to be used. Please follow instructions in the respective printer installation manual for correct printer installation. See 3-7-3 Optional Peripherals/Peripheral Connection on page 3-36 for more information. 1.) Select ADD...
Page 160
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 36 Section 3-7 - . Configuration 3-7-3 Optional Peripherals/Peripheral Connection 3-7-3-1 Approved Internal Peripherals This list covers the internal peripherals approved for Vivid 7: • Printer, Black & White, Video, SONY. • Printer, Black & White, Digital, SONY (need s software version v3.0 or above and BEP2). • Printer, Monochrome, Dig ital, SONY UP-D897SYN • Printer, Monochrome, Digital, SO NY UP-D895MD (needs software...