GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 121
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 2 - Site Preparations 2 - 11 2-3-2 Required Facility Needs NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a...
Page 122
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 2 - 12 Section 2-3 - Facility Needs 2-3-3 Desirable Features • Door is at least 92 cm (3 ft.) wide • Circuit breaker for dedicated po wer outlet is easily accessible • Sink with hot and cold water • Receptacle for bio–hazardous waste, like used probe sheaths • Emergency oxygen supply • Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet...
Page 123
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 2 - Site Preparations 2 - 13 2-3-5 Networking Setup Requirements 2-3-5-1 Stand Alone Scanner (without Network Connection) None. 2-3-5-2 Scanner Connected to Hospital’s Network Supported networks: DICOM Network (option) 2-3-5-3 Purpose of the DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network. Examples of DI COM...
Page 124
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 2 - 14 Section 2-3 - Facility Needs 2-3-5-5 Pre-quote Worksheet Figure 2-2 Pre-quote Worksheet for DICOM Network Information Vivid 7 Host Name AE Title Local PortIP Address Net Mask. .. . .. ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 . .. . .. . .. . .. . .. . .. . .. Destination IP Addresses GATEWAY IP Addresses Default DICOM APPLICATION INFORMATION NAME Store 2 Store 3 Store 1 MAKE/REVISION IP ADDRESSES PORT AE TITLE Store 5...
Page 125
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 1 Chapter 3 System Setup Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install Vivid 7. Included is a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim. How to prepare the facility and unit of the actual installation, and how to check a nd test the unit, probes, and external peripherals for...
Page 126
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 2 Section 3-2 - Installation Reminders Section 3-2 Installation Reminders 3-2-1 Average Installation Time 3-2-2 Installation Warnings Table 3-2 Average Installation Time Description Average Installation Time Comments Unpacking the scanner 0.5 hour Install Scanner wo/options 4 hours Dependent on the configuration DICOM Network Configuration 2 hours Dependent on the configuration DANGERWHEN USING ANY TEST INSTRUMENT THAT...
Page 127
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 3 3-2-2 Installation Warnings (cont’d) CAUTIONDO NOT OPERATE THIS UNIT UNLESS ALL BOARD COVERS AND FRAME PANELS ARE SECURELY IN PLACE. SYSTEM PERFORMANCE AND COOLING REQUIRE THIS. CAUTIONOPERATOR MANUAL(S) THE USER MANUAL(S) SHOULD BE FULLY READ AND UNDERSTOOD BEFORE OPERATING THE Vivid 7 AND KEPT NEAR THE UNIT FOR QUICK REFERENCE. CAUTIONACOUSTIC OUTPUT HAZARD ALTHOUGH THE ULTRASOUND ENERGY...
Page 128
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 4 Section 3-3 - Receiving and Unpacking the Equipment Section 3-3 Receiving and Unpacking the Equipment 3-3-1 Overview 3-3-1-1 Purpose of this section This section describes how to receive and unpack Vivid 7. Table 3-4 Contents in this section SectionDescriptionPage Number 3-3-1 Overview 3-4 3-3-2 The Post Delivery Checklist 3-5 3-3-3 The Tilt & Shock Indicators 3-6 3-3-4Receiving the Vivid 73-7 3-3-5Unpacking Vivid 73-10...
Page 129
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 5 3-3-2 The Post Delivery Checklist 3-3-2-1 Introduction Before shipment from the factory, the Vivid 7 has been thoroughly tested and visually inspected. Vivid 7 ultrasound scanners are fine tuned electronic inst ruments and should be treated properly during transportation. To learn about any issues that are discovered at th e reception of the package or when unpacking and installing the Vivid 7,...
Page 130
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 6 Section 3-3 - Receiving and Unpacking the Equipment 3-3-3 The Tilt & Shock Indicators 3-3-3-1 Overview Improper handling during transportation may harm the equipment inside the package even if the package itself is undamaged. To make it easier to detection if the handling duri ng transportation has been improper, a set of Tilt & Shock indicators have been attached to the transportation box. 3-3-3-2 Position of the Tilt &...