GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 17 FC000970 VIVID 7 DIMENSION VOLUME ULTRASOUND (BT’06) 230 VAC 230 VAC ±10% 5 A 50-60 Hz FC000980 VIVID 7 DIMENSION VOLUME ULTRASOUND (BT’06) 120 VAC 110-120 VAC ±10% 10 A 50-60 Hz FC000760 VIVID 7 DIMENSION, MULTIDIMENSIONAL RENDERING AND VOLUME 230 VAC ±10% 5 A 50-60 Hz FC000770VIVID 7 DIMENSION, MULTIDIMENSIONAL RENDERING AND VOLUME110-120 VAC±10%10 A50-60 Hz FC000780VIVID 7 DIMENSION230...
Page 142
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 18 Section 3-5 - Preparing for Setup FC000190Vivid 7 PRO (BT ’02)100-120 VAC±10%10 A50-60 Hz FC000180 Vivid 7 PRO (BT ’02) 230 VAC ±10% 5 A 50-60 Hz FC000060 Vivid 7 (BT ’01) 100-120 VAC ±10% 10 A 50-60 Hz FC000030 Vivid 7 (BT ’01) 230 VAC ±10% 5 A 50-60 Hz Table 3-10 Electrical Specifications for Vivid 7 (cont’d) sheet 3 of 3 GE VINGMED PART NUMBER DESCRIPTIONVOLTAGETOLERANCESCURRENTFREQUENCY
Page 143
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 19 3-6-3 Connections on the External IO 3-6-3-1 Connect Footswitch Connect the Footswitch to the External I/O at the rear side of Vivid 7. When not in use, store it in the tray below. 3-6-3-2 Connect Ethernet Connect Ethernet to the External I/O at the rear side of Vivid 7. 3-6-3-3 Connect USB Flash Card NOTE: Support for USB Flash Card was in troduced in software version v4.0. NOTE: USB Flash Cards...
Page 144
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 20 Section 3-5 - Preparing for Setup 3-6-4 Connections on the Patient I/O Panel 3-6-4-1 Connect ECG Connect ECG to Patient I/O Modu le in the front of Vivid 7. 3-6-4-2 Connect Phono Connect Phono to Patient I/O Mo dule in the front of Vivid 7. 3-6-4-3 Connect Pulse Pressure Transducer Connect Pulse Pressure Transducer to Pati ent I/O Module in the front of Vivid 7. Figure 3-7 ECG connecti on to Patient I/O Module Figure 3-8...
Page 145
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 21 3-6-5 Probe Connection V7/V7 PRO has four positions to plug in probes. • On systems without the 4D Imaging option, the left connector is a “dummy” connector, only used for parking one probe connector. The three other connectors can be selected and used for scanning. • On systems with the 4D Imaging option, the three leftmost connectors are standard probe ports and the probe port to the right...
Page 146
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 22 Section 3-5 - Preparing for Setup 3-6-5-1 Connect a Probe (cont’d) 5.) Position the Probe Cables in a way which the are not sweeping the floor. 3-6-5-2 Disconnect Probe 1.) Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector. (See Figure 3-11 .) 2.) Remove the connector from the port. 3.) Ensure that the probe head is clean before placing the probe in its storage case, see...
Page 147
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 23 3-6-6 Power ON/Bootup 3-6-6-1 Connect Mains Power to the Unit NOTE: Do not cycle the Circuit Breake r ON-OFF-ON in less than five (5) seconds. When turning OFF the Circuit Breaker, the system should de-energize completely before turning the circuit breaker ON. NOTE: When turning on a system from standby mode, it ta kes a few seconds before it responds. Do not push the On/off button again...
Page 148
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 24 Section 3-5 - Preparing for Setup 3-6-6-1 Connect Mains Power to the Unit (cont’d) 5.) Connect the Power Cable’s female plug to the Power Inlet at the rear of the unit. 6.) Lock the plug in position wi th the Retaining Clamp (ACC Clamp). 7.) Verify that the Mains Power Circuit Breaker is in OFF position, if not, switch it OFF. 8.) Connect the Power Cable’s other end (male plug ) to a hospital grade mains power outlet...
Page 149
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 25 3-6-6-2 Turn Unit ON 1.) Switch ON the Mains Power Circuit Breaker at the rear of the unit. You should hear a “click” from the relays in the AC Power/AC Controller and the unit is ready to boot. NOTE: If the unit has been in the OFF co ndition for an extended period of time, (3 to 5 days or more), the unit may not boot, or may beep when turned on. 2.) Press once on the ON/OFF KEY on the...
Page 150
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 26 Section 3-5 - Preparing for Setup 3-6-6-2 Turn Unit ON (cont’d) f.) A start-up bar indicating the time used fo r software loading, is displayed on the monitor. g.) The software initiates and se ts up the Front-End electronics and the rest of the instrument. h.) The backlight in the keyboard is lit. i.) As soon as the software has been loaded, a 2D screen is displayed on the screen. NOTE: Total time used for start-up is...