GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 181
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 57 3-9-13 Configurable Remote Path User It is possible to set the User and Password for all remote paths as secondary creditials. These settings are only used if the log in with the default User and Password fails. If the log in with the default User and Password succeed, these settings are not used. Follow these steps to configure the Remote Path User: 1.) Press Config (F2 ) and log on as ADM,...
Page 182
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 58 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14 Set Up Connection to a DICOM Server in a Network In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are first saved on the local image buffer on the scanner. At the end of the examination the images are sent to the DICOM server via a DICOM spooler and to the local database, depending on...
Page 183
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 59 3-9-14-2 DICOM Server IP Address Setting on the Scanner (cont’d) In Figure 3-54 , the selected flow, Worklist/Local Archive - DICOM Server/Int. HD , is shown. Setup of other dataflows are similar to this example. 6.) Select Worklist (so it is highlighted) and then select Properties to display the Properties dialog. If Direct Store is enabled, the image will go to the DICOM server...
Page 184
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 60 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-2 DICOM Server IP Address Setting on the Scanner (cont’d) 7.) Select the IP-Address down-arrow to choose the Worklist Server from the pull-down menu. NOTE: It is not possible to change the setting in the IP-Address field by editing it. See description below, starting with step 8, to change the setting. 8.) Follow the steps below to...
Page 185
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 61 3-9-14-2 DICOM Server IP Address Setting on the Scanner (cont’d) c.) Edit the name and/or the IP address of the server, see Figure 3-58. d.) Select OK to save the new settings and close the dialog. e.) Select OK to close the IPs dialog to re turn to the Properties dialog. f.) Enter the DICOM server AE Title. This entr y is case sensitive and must match exactly. g.) Enter the DICOM server...
Page 186
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 62 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-3 Verify the Network Connection to a Device Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection 1.) Select the “Smiley” button to Ping the server. NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM ping), so it does not check AE title or port number. 2.) If the network...
Page 187
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 63 3-9-14-4 Verify the Connection to a Device (cont’d) NOTE: If the Check button fails directly, the AE Title is probably wrong. If the Check button fails after a long time (corres ponding to the timeout), the IP address or Port Number is probably wrong. In the illustration below, the mess age indicates that the pinging of the server worked fine (line 1), but the application on the server didn’t...
Page 188
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 64 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-5 DICOM Storage Setup 1.) Select Dicom storage . 2.) Select Properties to display the DICOM Storage properties screen. Figure 3-63 Select Dicom storage Figure 3-64 DICOM Storage Properties DICOM STORAGE SOFTWARE VERSION v6.x and v7.x SOFTWARE VERSION v5.x
Page 189
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 65 3-9-14-6 DICOM Storage Properties Adjustments IP-address 1.) Select the down-arrow at the IP-Address pull-down menu to select the DICOM Storage server. 2.) Select from the pull-down menu if you need to modify the server’s parameters or add a new server. This will bring up the IPs dialog. If no changes are needed, continue with step 7. 3.) To modify any of the settings, select the server you...
Page 190
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 66 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-14-6 DICOM Storage Properties Adjustments (cont’d) 7.) Next, verify the other settings and parameters as described below. Name Raw data Storage - Name of service. AE Title DICOMSTORAGESCP - The DICOM server’s AE Title. Port No Verify / set the DICOM server’s Port Number. Image Settings • Allow Raw Data - default ON (checked) • Raw...