GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 97 3-10-8-3 Verify the Connection to a Device - Software Version v3.x 1.) Select (highlight) the device you want to verify the connection to. 2.) Select CHECK to start the verification process of the connection to the device. The verification process may use some amount of time (several seconds). When done, a sign in front of the device i ndicates if the test passed (green check-mark) or failed...
Page 222
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 98 Section 3-10 - Connectivity Setup - Software v3.x 3-10-9 HL7 Communication Setup - Software v3.x 3-10-9-1 Introduction The procedures below tell you how to set up the Vivid 7 so it can connect to the Hospital Information System (HIS), via the Vivid HL7 Gateway. By connec ting to the HIS, demographic information i.e Patient ID, Name, Gender etc. can be pulled from the HI S, reducing “double-work” and typing errors....
Page 223
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 99 3-10-9-3 Configuration of Modality Worklist Services - Software v3.x (cont’d) Configure the Search Criterias NOTE: This configuration is required. 1.) Select the Search Criterias button. 2.) Select the criteria Modality and select the Remove button. 3.) Press the OK button. Configure Search Criterias: Constrain Query by AE Title NOTE: This configuration is optional. It should only be...
Page 224
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 100 Section 3-10 - Connectivity Setup - Software v3.x 3-10-9-3 Configuration of Modality Worklist Services - Software v3.x (cont’d) Test Connectivity with the Vivid HL7 Gateway The purpose of this test is to verify that the Vivid 7 has connectivity with the Vivid HL7 Gateway. 1.) From either the Current or Dataflow view, expand the Dataflow incorporating the Gateway Modality Worklist service. 2.) Expand the Gateway...
Page 225
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 101 3-10-9-3 Configuration of Modality Worklist Services - Software v3.x (cont’d) Test Connectivity with the Vivid HL7 Gateway 1.) Start a new study and acquire an image, or use an existing study. 2.) In the Measurements mode, make some of the measurements. 3.) In the Report screen, cr eate and store a report. 4.) In the Patient screen, select the study that you created the report in from the...
Page 226
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 102 Section 3-10 - Connectivity Setup - Software v3.x 3-10-10 Query/Retrieve (Q/R) Setup for Software v3.x 3-10-10-1 Overview Query/Retrieve was introduced with the V3.0 software. The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM server for further analysis on the Vivid 7. NOTE: You may have to set up Vivid 7 as a destination on the server. 3-10-10-2 Query/Retrieve Setup on...
Page 227
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 103 3-10-10-2 Query/Retrieve Setup on the Vivid 7 (cont’d) In Figure 3-108 , the selected dataflow, Query/Retrieve, is shown. 5.) Select QueryRetrieve so it is hi ghlighted and then select Propertie s to display the Properties dialog. Figure 3-108 Query/Retrieve workflow Figure 3-109 Select Query/Retrieve’s Properties SELECT QUERY RETRIEVE AND THEN SELECT PROPERTIES QUERY RETRIEVE PROPERTIES DIALOG
Page 228
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 104 Section 3-10 - Connectivity Setup - Software v3.x 3-10-10-2 Query/Retrieve Setup on the Vivid 7 (cont’d) 6.) Select the IP-address down-arrow to choose t he DICOM Query/Retrieve server from the pull down menu. In some cases, the server to use is the same as used for DICOM Storage. If the server to use is missing from the list, select from the pull down menu and edit the setup for one of the predefined servers. 7.)...
Page 229
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 105 3-10-10-2 Query/Retrieve Setup on the Vivid 7 (cont’d) Change Search Criterias If is possible to set up special Se arch Criterias for DICOM Query/Retrieve. In most cases you may leave the Search Criterias as is , and skip this adjustment. Follow the steps below to change the Search Criterias parameters: 1.) Select Search Criterias 2.) Select the correct tag from the Select Tag pull-down menu. 3.)...
Page 230
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 106 Section 3-10 - Connectivity Setup - Software v3.x 3-10-10-3 Query/Retrieve Verification Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection 1.) Select the “Smiley” button to Ping the server. NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM ping), so it does not check AE title or port number. 2.) If the network connection to the server is OK, it will be...