GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 87
3-9-18-3 Examination List Screen Figure 3-91   Examination List Screen
Table 3-19    Examination List Screen
ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
1 REFERRAL REASON (0010,2180) ADDITIONAL PATIENT’S HISTORY
2 DIAGNOSIS (0008,103E) SERIES DESCRIPTION
1
2 

Page 212

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 88 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-18-4 Worklist Search Screen Figure 3-92   Worklist Search Screen
Table 3-20    Worklist Search Screen
ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME
2 SEX (0010,0040) PATIENT’S SEX
3 ACCESSION (0008,0050) ACCESSION NUMBER
4BIRTHDATE(0010,0030)PAT I E N T ’ S  B I RT H D AT E
5PATIENT...

Page 213

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 89
Section 3-10
Connectivity Setup - Software v3.x
NOTE: If connected to a stand-alone network (Peer-to-Peer network with a Vivid 7 scanner, an  EchoPAC PC Workstation and an optional network printer), you should use default delivery 
settings. 
3-10-1 Introduction
To be able to use the network functions when con nected to a hospital network, the scanner must have 
a proper network address. 
• Before...

Page 214

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 90 Section 3-10 - Connectivity Setup - Software v3.x
3-10-4 Select TCP/IP Set-up Scr een - Software Version v3.x
1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
2.) If not already selected, select  CONNECTIVITY
 from the bottom row of “buttons” on the screen.
3.) Select the TCP/IP TAB (it is named Tcpip). 
.
Figure 3-93   TCP/IP Set-up screen for Vivid 7, Overview (Example)...

Page 215

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 91
3-10-5 Set the Scanner’s  Network Information - Software Version v3.x
In the IP settings area of the screen, enter the;
1.) IP Address for scanner. (Default IP  Address from factory: 10.0.0.3).
2.) Subnet Mask for scanner. (Default Subn et Masks from factory: 255.255.255.0).
3.) IP address for Default Gateway. (Default  Gateways from factory IP address: 0.0.0.0).
4.) In addition, the AE Title for the...

Page 216

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 92 Section 3-10 - Connectivity Setup - Software v3.x
3-10-6 Set the Remote Archive’s Networ k Information - Software Version v3.x
In the Remote Archive Setup area of the screen (see example in  Figure 3-95), enter the;
1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001).
.
3-10-7 Save the New Settings
1.) Press SAVE...

Page 217

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 93
3-10-8 Connect to a DICOM Server in  a Network - Software Version v3.x
In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are 
first saved on the local image buffer on the scanner. At the end of the examination the images are sent 
to the DICOM server via a DICOM spooler and to  the local database, depending on dataflow.
This scenario requires that the...

Page 218

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 94 Section 3-10 - Connectivity Setup - Software v3.x
3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  In Figure 3-98 , the selected flow, Worklist/Local Arch ive - DICOM Server/Int. HD, is shown. 
 
6.) Select Worklist (so it is highlighted) and then se lect Properties to display the Properties dialog. 
 
Figure 3-98   Worklist/Local Archive - DICOM Server/Int. HD flow
Figure 3-99...

Page 219

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 95
3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  7.) Select the IP-Address down-arrow to choose the Worklist Server from the pull-down menu.
NOTE: It is not possible to change the setting in  the IP-Address field by editing it. See description 
below, starting with step  8, to change the setting.
8.) Follow the steps below to change the IP-Address...

Page 220

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 96 Section 3-10 - Connectivity Setup - Software v3.x
3-10-8-2      DICOM Server IP Address Setting on the Scanner - Software Version v3.x (cont’d)  c.) Edit the name and/or the IP address of the server, see  Figure 3-103. 
d.) Select OK to save the new  settings and close the dialog.
e.) Select OK to close the IPs dialog  to return to the Properties dialog.
f.) Enter the DICOM server AE Title. This entr y is case sensitive and...