Page 171
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 47 3-9-5-3 Change AE Title and/or Port Number (Port No.) NOTE: You don’t need to use the Modify butt on when changing AE Title and/or Port No. 1.) To change AE Title and/or Port No ., edit the respective fields. 2.) Select Save settings to store your changes. This will bring up a new Warning screen. 3.) Select Ok to save your changes or Cancel to return without saving any changes. 4.) Reboot Vivid...
Page 172
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 48 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later NOTE: For software version v4.0/v4.1/v5.0/v5.1/v6.0, see: 3-9-7 Set the Scanner’s Network Information - S/W v4.0/v4.1/v5.0/v5.1/v6.0 on page 3-50 . 1.) When in the Tcpip (TCP/IP) screen, select Advanced Settings to display the Network Connections screen. 2.)...
Page 173
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 49 3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later (cont’d) 3.) Select (highlight) Internet Protocol (TCP/IP) and select Properties to display the Internet Protocol (TCP/IP) Properties screen. The Internet Protocol (TCP/IP) Properties screen is used to set the following parameters: If using DHCP Check the Obtain an IP address automatically check box. If not...
Page 174
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 50 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-7 Set the Scanner’s Network Info rmation - S/W v4.0/v4.1/v5.0/v5.1/v6.0 In the IP settings area of the screen, enter the: 1.) IP Address for scanner. (Default IP Ad dress setting from factory: 10.0.0.3). 2.) Subnet Mask for scanner. (Default Subnet Mask setting from factory: 255.255.255.0). 3.) IP address for Default Gateway. (Default Gate...
Page 175
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 51 3-9-8 Advanced Settings NOTE: New for software V4.0 (BT’04). Advanced Settings give you access to the Network and Dial-up Connections screen where you can set up special network connections and ad just settings for network connectivity. On Vivid 7, Advanced Settings is used to enable/disable DHCP. . Figure 3-43 Advanced Settings ADVANCED SETTINGS TO BE USED FOR ENABLING/DISABLING DHCP.
Page 176
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 52 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-9 DHCP Configuration 3-9-9-1 Pre-Requisite • Software version v4.0 or higher. NOTE: The remote connectivity fo r the Global Service Platform will not function if DHCP is active. 3-9-9-2 Turn DHCP ON Follow the instructions below to configure the Vivid 7 to use DHCP . 1.) When in the TCP/IP screen, select ADVANCED SETTINGS to display the...
Page 177
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 53 3-9-9-2 Turn DHCP ON (cont’d) 2.) Select Local Area Connection 3.) Select Properties to display the Local Area Connection Properties 4.) Scroll down to Internet Protocol (TCP/IP) and select it. Figure 3-45 Network Connections Figure 3-46 Local Area Connection Properties
Page 178
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 54 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-9-2 Turn DHCP ON (cont’d) 5.) Select Properties 6.) Select Obtain an IP address automaticallyFigure 3-47 Local Area Connection Status
Page 179
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 55 3-9-10 Set the Remote Arch ive’s Network Information To be able to connect to a remote archive, on a remote computer or server, you must configure Vivid 7 to communicate with it. In the Remote Archive Setup area of the Tcpip screen, see example in Figure 3-48, enter the; 1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4). 2.) Remote Archive Name. (Default Remote Ar chive...
Page 180
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 56 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-12 Remote Path Remote Path is used for Save As, Export from Q-A nalysis, and for exporting error logs with the command. Follow this procedure to Configure Vivid 7 to use the Remote Path. 1.) Press Config (F2 ) and log on as ADM, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) If not already selected, select Connectivity from...