Page 171

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 47
3-9-5-3 Change AE Title and/or Port Number (Port No.) NOTE: You don’t need to use the Modify butt on when changing AE Title and/or Port No.
1.) To change  AE Title and/or Port No ., edit the respective fields.
2.) Select  Save settings  to store your changes. This  will bring up a new Warning screen.
3.) Select  Ok to save your changes or  Cancel to return without saving any changes.
4.) Reboot Vivid...

Page 172

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 48 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later
NOTE: For software version v4.0/v4.1/v5.0/v5.1/v6.0, see:  3-9-7 Set the Scanner’s Network 
Information - S/W v4.0/v4.1/v5.0/v5.1/v6.0 on page 3-50
.
1.) When in the  Tcpip (TCP/IP) screen, select  Advanced Settings to display the Network 
Connections  screen.
2.)...

Page 173

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 49
3-9-6 Set the Scanner’s Ne twork Information - S/W v4.2/v5.2/v6.1/v7.x and later 
(cont’d)
3.) Select (highlight)  Internet Protocol (TCP/IP)  and select Properties to display the  Internet 
Protocol (TCP/IP) Properties  screen.
The  Internet Protocol (TCP/IP) Properties  screen is used to set the following parameters: 
If using DHCP
Check the  Obtain an IP address automatically  check box.
If not...

Page 174

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 50 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-7 Set the Scanner’s Network Info rmation - S/W v4.0/v4.1/v5.0/v5.1/v6.0
In the IP settings  area of the screen, enter the:
1.) IP Address for scanner. (Default IP Ad dress setting from factory: 10.0.0.3).
2.) Subnet Mask for scanner. (Default Subnet  Mask setting from factory: 255.255.255.0).
3.) IP address for Default Gateway. (Default Gate...

Page 175

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 51
3-9-8 Advanced Settings
NOTE: New for software V4.0 (BT’04). Advanced Settings  give you access to the  Network and Dial-up Connections  screen where you can 
set up special network connections and ad just settings for network connectivity.
On Vivid 7, Advanced Settings is used to enable/disable DHCP.
.
Figure 3-43   Advanced Settings
ADVANCED SETTINGS
TO BE USED FOR 
ENABLING/DISABLING 
DHCP. 

Page 176

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 52 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-9 DHCP Configuration
3-9-9-1 Pre-Requisite • Software version v4.0 or higher.
NOTE: The remote connectivity fo r the Global Service Platform will not function if DHCP is active.
3-9-9-2 Turn DHCP ON  Follow the instructions below to  configure the Vivid 7 to use DHCP
.
1.) When in the TCP/IP screen, select  ADVANCED SETTINGS to display the...

Page 177

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 53
3-9-9-2      Turn DHCP ON (cont’d) 2.) Select Local Area Connection
3.) Select  Properties  to display the  Local Area Connection Properties
4.) Scroll down to  Internet Protocol (TCP/IP)  and select it.
Figure 3-45   Network Connections 
Figure 3-46   Local Area Connection Properties 

Page 178

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 54 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-9-2      Turn DHCP ON (cont’d) 5.) Select Properties
6.) Select  Obtain an IP address automaticallyFigure 3-47   Local Area Connection Status 

Page 179

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 55
3-9-10 Set the Remote Arch ive’s Network Information
To be able to connect to a remote archive, on a remote computer or server, you must configure Vivid 7 
to communicate with it.
In the Remote Archive Setup area of the Tcpip screen, see example in  Figure 3-48, enter the;
1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Ar chive...

Page 180

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 56 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-12 Remote Path
Remote Path is used for Save As, Export from Q-A nalysis, and for exporting error logs with the  
command. 
Follow this procedure to Configure Vivid 7 to use the  Remote Path.
1.) Press  Config (F2 ) and log on as  ADM, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
2.) If not already selected, select  Connectivity from...