DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 77
3-9-16-3 Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites)  This dataflow is intended for Vivid 7 scanners in a DI COM network, where the data should not be stored 
locally on the scanner. A site with the integrated  EchoPAC PC and CA-1000 is an example of this.
To create a new dataflow with Worklist an d DICOM Storage, do the following steps:
1.) Add a new dataflow in Config...

Page 202

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 78 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17 Query/Retrieve (Q/R) Setup  for Software v7.x/v6.x/v5.x/v4.x
3-9-17-1 Overview
The Query/Retrieve function  makes it possible to search for and  retrieve DICOM data from a DICOM 
server for further analysis on the Vivid 7. 
NOTE: You may have to set up Vivid 7 as a destination on the server.
3-9-17-2 Query/Retrieve Setup on the Vivid 7 with...

Page 203

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 79
3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
In  Figure 3-78 , the selected dataflow, Query/Retrieve, is shown.
5.) Select QueryRetrieve so it is hi ghlighted and then select Propertie s to display the Properties dialog.
Figure 3-78   Query/Retrieve workflow
Figure 3-79   Select Qu ery/Retrieve’s Properties
SELECT 
QUERY RETRIEVE 
AND THEN SELECT...

Page 204

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 80 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
6.) Select the IP-address down-arrow to choose t he DICOM Query/Retrieve server from the pull down 
menu. In some cases, the server to use  is the same as used for DICOM Storage.
If the server to use is missing  from the list, select  from the pull down...

Page 205

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 81
3-9-17-2      Query/Retrieve Setup on the  Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d)
Change Search Criterias
It is possible to set up special Se arch Criterias for DICOM Query/Retrieve. In most cases you may leave 
the Search Criterias as is , and skip this adjustment.
Follow the steps below to change the Search Criterias parameters:
1.) Select Search Criterias
2.) Select the correct tag...

Page 206

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 82 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-3 Query/Retrieve Verification
Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection
1.) Select  Query retrieve  and then select  Properties.
2.) Select the “Smiley” button to Ping the server.
NOTE: The smiley checks if the remote server is  accessible (ping). It is not a DICOM Echo (DICOM 
ping), so it does not check AE...

Page 207

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 83
3-9-17-3      Query/Retrieve Verification (cont’d)
Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow 
NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the  TCP/IP connection to the server.
- DICOM Ping is used to verify that  the DICOM application is answering.
1.) Select QueryRetrieve from the Selected devices list.
2.) Select Check.  - If the test pass, a...

Page 208

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 84 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-17-3      Query/Retrieve Verification (cont’d)       Follow the Steps Below to Test  the DICOM Query/Retrieve Workflow (cont’d)
- If the test fails, a pop-up dialog reporting the resu lts as “0 of 1 OK” is displayed. In the example 
in  Figure 3-86,  the warning message pop up almost at onc e, and it states that for the Server 
Check (Ping), 1...

Page 209

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 85
3-9-18 Mapping of Parameters  from Vivid 7 to DICOM
3-9-18-1 Search Screen
Figure 3-89   Search / Create Patient Screen
Table 3-17    Search / Create Patient Screen
ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME
2 SEX (0010,0040) PATIENT’S SEX
3 ACCESSION (0008,0050) ACCESSSION NUMBER
4PAT I E N T  I D(0010,0020)PATIENT ID...

Page 210

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 86 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x
3-9-18-2 Patient Information Screen Figure 3-90   Patient Information Screen
Table 3-18    Patient Information Screen
ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME
1 OPERATOR (0008,1070) OPERATOR’S NAME
2 REF. DOC (0008,0090) REFFERING PHYSICIAN’S NAME
3 DIAGN.PHYS. (0008,1050) PERFORMING PHYSICIAN’S NAME
4WARD/DEPT(0008,1040)INSTITUTIONAL...