GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 77 3-9-16-3 Create a New Dataflow with Worklist and DICOM Storage (for CA-1000 sites) This dataflow is intended for Vivid 7 scanners in a DI COM network, where the data should not be stored locally on the scanner. A site with the integrated EchoPAC PC and CA-1000 is an example of this. To create a new dataflow with Worklist an d DICOM Storage, do the following steps: 1.) Add a new dataflow in Config...
Page 202
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 78 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17 Query/Retrieve (Q/R) Setup for Software v7.x/v6.x/v5.x/v4.x 3-9-17-1 Overview The Query/Retrieve function makes it possible to search for and retrieve DICOM data from a DICOM server for further analysis on the Vivid 7. NOTE: You may have to set up Vivid 7 as a destination on the server. 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with...
Page 203
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 79 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) In Figure 3-78 , the selected dataflow, Query/Retrieve, is shown. 5.) Select QueryRetrieve so it is hi ghlighted and then select Propertie s to display the Properties dialog. Figure 3-78 Query/Retrieve workflow Figure 3-79 Select Qu ery/Retrieve’s Properties SELECT QUERY RETRIEVE AND THEN SELECT...
Page 204
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 80 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) 6.) Select the IP-address down-arrow to choose t he DICOM Query/Retrieve server from the pull down menu. In some cases, the server to use is the same as used for DICOM Storage. If the server to use is missing from the list, select from the pull down...
Page 205
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 81 3-9-17-2 Query/Retrieve Setup on the Vivid 7 with Software v7.x/v6.x/v5.x/v4.x (cont’d) Change Search Criterias It is possible to set up special Se arch Criterias for DICOM Query/Retrieve. In most cases you may leave the Search Criterias as is , and skip this adjustment. Follow the steps below to change the Search Criterias parameters: 1.) Select Search Criterias 2.) Select the correct tag...
Page 206
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 82 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-3 Query/Retrieve Verification Follow the Steps Below to do a First Test (TCP-IP Ping) of the Connection 1.) Select Query retrieve and then select Properties. 2.) Select the “Smiley” button to Ping the server. NOTE: The smiley checks if the remote server is accessible (ping). It is not a DICOM Echo (DICOM ping), so it does not check AE...
Page 207
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 83 3-9-17-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the TCP/IP connection to the server. - DICOM Ping is used to verify that the DICOM application is answering. 1.) Select QueryRetrieve from the Selected devices list. 2.) Select Check. - If the test pass, a...
Page 208
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 84 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-17-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d) - If the test fails, a pop-up dialog reporting the resu lts as “0 of 1 OK” is displayed. In the example in Figure 3-86, the warning message pop up almost at onc e, and it states that for the Server Check (Ping), 1...
Page 209
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 85 3-9-18 Mapping of Parameters from Vivid 7 to DICOM 3-9-18-1 Search Screen Figure 3-89 Search / Create Patient Screen Table 3-17 Search / Create Patient Screen ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME 1 LAST NAME, FIRST NAME (0010,0010) PATIENT’S NAME 2 SEX (0010,0040) PATIENT’S SEX 3 ACCESSION (0008,0050) ACCESSSION NUMBER 4PAT I E N T I D(0010,0020)PATIENT ID...
Page 210
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 86 Section 3-9 - Connectivity Setup - Software v7.x (BT’08), v6.x, v5.x and v4.x 3-9-18-2 Patient Information Screen Figure 3-90 Patient Information Screen Table 3-18 Patient Information Screen ITEMVIVID 7 FIELD NAMEDICOM TAG NUMBERDICOM TAG NAME 1 OPERATOR (0008,1070) OPERATOR’S NAME 2 REF. DOC (0008,0090) REFFERING PHYSICIAN’S NAME 3 DIAGN.PHYS. (0008,1050) PERFORMING PHYSICIAN’S NAME 4WARD/DEPT(0008,1040)INSTITUTIONAL...