GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 251
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 127 3-11-10-6 Set-up of DICOM Server in Vivid 7’s Configuration Screens - Software v2.x 1.) From the Name pull-down menu select DICOMSERVER. 2.) Enter the IP address of the DICO M server in the IP Address field. 3.) Select the Pure DICOM Storage line under Services and the bottom part of the screen will be populated with set-up information for the DICOM server. Figure 3-134 Select DICOM Server...
Page 252
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 128 Section 3-11 - Connectivity Setup - Software v2.x 3-11-10-6 Set-up of DICOM Server in Vivid 7’s Configuration Screens - Software v2.x (cont’d) Figure 3-136 Select Pure DICOM Storage Select Pure DICOM Storage. Before selecting After
Page 253
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 129 3-11-10-6 Set-up of DICOM Server in Vivid 7’s Configuration Screens - Software v2.x (cont’d) 4.) Change the Name to a name you pick for the DICOM server (e.g. ProSolv server). It doesn’t really matter what this name is. 5.) Change the AE Title to the name you got from the network administrator. This name must match exactly and is case sensitive. 6.) Change the port number to the nu mber you...
Page 254
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 130 Section 3-11 - Connectivity Setup - Software v2.x 3-11-10-7 Check the Connection to the DICOM Server (Ping) - Software v2.x 1.) After the system is rebooted, press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Enter the Connectivity screen. 3.) Select the VIEWS tab and the DATAFLOW button and you will see all available data flows. 4.) Select the + by the “DICOM Server”...
Page 255
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 131 3-11-10-8 Setting the DICOM Server as Default Dataflow - Software v2.x 1.) Finish the verification in Check the Connection to the DICOM Server (Ping) - Software v2.x on page 3-130 . You need to have green check marks on both Ha rdware Ping and DICOM Ping before you continue. 1.) Select the DATAFLOW tab. 2.) Select “DICOM Server” dataflow on the “Dataflow Name” drop-down menu. . 3.) Check the...
Page 256
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 132 Section 3-11 - Connectivity Setup - Software v2.x 3-11-10-9 Performing a Study 1.) After restart, hit New Exam (you’ll see that DICOM Server is the default dataflow), enter a last name and patient ID and start scanning and image acquisi tion. Images are stored in the local image buffer on the system. 2.) After having finished the study, Select New Exam and select All when asked if you want to save. The images will now...
Page 257
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 133 Section 3-12 Ethernet Switch / Hub An Ethernet Switch is used to connect Vivid 7 to an EchoPAC PC and a network printer in a Peer-to- Peer network. It can also be connected to the hospital’s network for access to a DICOM server. The Ethernet Switch described here is included as an example. Both five port and eight port switches have been delivered with EchoPAC PC. Please r ead and follow the...
Page 258
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 134 Section 3-12 - Ethernet Switch / Hub 3-12-1 Local Network Connection to EchoPAC PC Workstation Figure 3-145 Local Network Connection EchoPAC PC Network PrinterVivid7 Do not use the fifth connector and the switch can be in any position,Use only Standard non-crossed Ethernet cabling. Vivid3
Page 259
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 135 3-12-2 Hospital Network Connectio n to EchoPAC PC Workstation See the EchoPAC PC Service Manual , Direction Number EP091298, for details about EchoPAC PC, and its set-up. Figure 3-146 Hospital Network Connections Using standard non-crossed cables and MDI switch set to position =HUB, connect like this. Wall connection EchoPAC PC Network Printer Vivid 3 or Vivid 7
Page 260
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 136 Section 3-13 - Installation Paperwork Section 3-13 Installation Paperwork NOTE: During and after installation, the documentat ion (i.e. CDs with documentation, User’s Manuals, Installation Manuals etc.) for the peripheral units must be kept as part of the original system documentation. This will ensure th at all relevant safety and user informations are available during the operation and serv ice of the complete...