Page 301
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 37 4-3-6-6 2D Soft Menu Controls 4-3-6-7 Check Compress, Contour, Reject and Tilt Table 4-9 4-way rocker used for Soft Menu control KEYDESCRIPTION A 4-way rocker (4-way multi direction switch) used to; - access mode-specific menus on the screen. - select a menu option. - adjust option-related values. When using the Rocker; - the vertical arrows are used to select the menu options. - the...
Page 302
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 38 Section 4-3 - Functional Checks 4-3-6-8 Check DDP, Diff, Power and Dynamic Range Table 4-11 2D Mode: Check DDP, Diff, Power and Dynamic Range StepTaskExpected Result(s) 1. Use the vertical arrows to select DDP (Data Dependent Processing). Use the horizontal arrows to adjust DDP. Performs temporal processing which reduces random noise without affecting the motion of significant tissue structures. An index number...
Page 303
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 39 4-3-7 M Mode Checks 4-3-7-1 Introduction 4-3-7-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the scanner’s left-most probe connector. -See 3-6-5 Probe Connection on page 3-21 for info about connecting the probes. -See Section 9-26 Probes on page 9-80 for available probes. 2.) Turn ON the scanner. -See see 4-2-2 Power ON/ Boot UP on page 4-3 for...
Page 304
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 40 Section 4-3 - Functional Checks 4-3-7-2 Preparations (cont’d) Figure 4-30 M Mode, Freeze and Cine Replay controls Exit Gain * Mode Dependant Keys Horizontal Sweep* Frequency Focus Contour Up/Down Color Maps Soft menu Compress Reject Dynamic Range Power Freeze Controls marked with * are also available in freeze and cine replay SteCurZoom Alt. PW CW TVI 2D MM Color Tra ck B al lU pd at e M e nu Act. Win Act. Mode Img....
Page 305
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 41 4-3-7-3 Check Horizontal Sweep, Frequency, Focus and Contour Table 4-12 M-Mode: Check Horizontal Sweep, Frequency, Focus and Contour StepTaskExpected Result(s) 1. Adjust HORIZONTAL SWEEP Adjusts the horizontal refresh rate of the M-Mode area of the display. Horizontal sweep does not change the acquisition resolution, so you can change the horizontal sweep in replay without any...
Page 306
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 42 Section 4-3 - Functional Checks 4-3-7-4 Check Compress, Reject, Power and Dynamic Range Table 4-13 M-Mode Soft menu Controls check StepTask Expected Result(s) 1. Adjust Compress Controls the amount of contrast in the image. An index number is displayed in the status window to indicate the relative level of compression. 2. Adjust Reject Adjust reject level. When Reject is increased, low-level echoes are rejected and...
Page 307
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 43 4-3-8 Color Mode Checks 4-3-8-1 Introduction Color Flow screens is 2D or M Mode screens with colors representing blood or tissue movement. Color Flow may be selected both from 2D mode or from M mode or a combination of these. 4-3-8-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the scanner’s left-most probe connector. -See Probe Connection on page 3-21...
Page 308
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 44 Section 4-3 - Functional Checks 4-3-8-4 Color M Mode (cont’d) Figure 4-31 Color Flow Mode Controls Exit Depth Gain * Mode Dependant Keys Horiz. sweep (Color M-Mode only) * Scale Baseline * LVR Invert * Variance * Color maps * Simultaneous (Color 2D only) Angio (Color 2D only) Cineloop (in Freeze, Color 2D mode only) Soft menu Sample Volume Tissue Priority * DDP (Color 2D only) * Frequency Frame rate Radial avg....
Page 309
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 45 4-3-8-5 Check Horizontal Sweep, PRF, Baseline, LVR and Invert Table 4-14 Color Flow Mode Controls Check StepTask Expected Result(s) 1. Adjust Horizontal Sweep (Color M-Mode only) Adjusts the horizontal refresh rate of the M-Mode area of the display. 2. Adjust Baseline Adjusts the color map to emphasize flow either toward or away from the probe. Baseline is available in both Live and...
Page 310
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 46 Section 4-3 - Functional Checks 4-3-8-6 Check Variance, Color Maps and Cineloop Table 4-15 Color Flow Mode: Check Variance, Color Maps and Cineloop StepTask Expected Result(s) 6. Adjust Variance Controls the amount of variance data added to a color display. Variance enables computer-aided detection of non- luminary flow (e.g. jets or regurgitation). Variance is available in live and cine replay. 7. Select Color Maps...