GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 57 4-3-11 Tissue Velocity Imaging (TVI) Checks 4-3-11-1 Introduction TVI calculates and color codes the velocities in tissue. The tissue velocity information is acquired by sampling of tissue Doppler veloci ty values at discrete points. 4-3-11-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the scanner’s left-most probe connector. -See Probe Connection ,...
Page 322
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 58 Section 4-3 - Functional Checks 4-3-11-2 Preparations (cont’d) Figure 4-34 Doppler Mode Controls Mode Dependant Keys 2D Width Scale Transparency Threshold Invert * Tissue Tracking Simultaneous * Cineloop (Freeze only) Color Maps Soft menu Tissue Priority * Frame Rate Compress * Reject * Frequency Lateral Average Radial Average Baseline * Power TVI Freeze 2D Alt. Alternatives Anat. M-Mode Curved Anat. M-Mode Zoom...
Page 323
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 59 4-3-11-3 Check 2D Width, Scale, Transparency and Threshold Table 4-23 TVI Mode Controls Checks StepTaskExpected Result(s) 1. Adjust 2D WIDTH Controls the size or angular width of the 2D sector. 2. Adjust SCALE Controls the repetition rate of the Doppler pulses transmitted to acquire the data for color mapping. The scale value influences the Nyquist limit (the ability to detect...
Page 324
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 60 Section 4-3 - Functional Checks 4-3-11-4 Check Invert, Tissue Tracking, Simultaneous and Color Maps Table 4-24 TVI Mode Controls Checks St e pTa s kExpected Result(s) 5. Press INVERT Enables the color scheme assigned to positive and negative tissue velocities to be inverted. Invert is available in live and cine replay. 6. Press TISSUE TRACKING Starts Tissue Tracking mode. 7. Choose COLOR MAPS Displays a drop...
Page 325
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 61 4-3-11-5 Check Tissue Priority, Frame Rate, Compress, Reject, Frequency, Lateral & Radial Average, Baseline and Power Table 4-25 TVI Mode: Soft Menu Controls Check StepTask Expected Result(s) 8. Adjust TISSUE PRIORITY Emphasizes either the color of the TVI or the greyscale tissue detail of the 2D image. Tissue priority is avail able in both Live and Freeze. 9. Adjust FRAME RATE...
Page 326
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 62 Section 4-3 - Functional Checks 13.Adjust LATERAL AVERAGE Smooths the image by averaging collected data along the same horizontal line. An increase of the lateral averaging will reduce noise, but this will also reduce the lateral resolution. 14. Adjust RADIAL AVERAGE Smooths the image by averaging collected data along the same radial line. An increase of the radial averaging will reduce noise, but this will also...
Page 327
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 63 4-3-12 Contrast Checks Refer to Chapter 5 in Vivid 7 User Manual. 4-3-13 Stress Echo Refer to Chapter 4 in Vivid 7 User Manual.
Page 328
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 64 Section 4-3 - Functional Checks 4-3-14 Measurements and Multi Image Checks 4-3-14-1 Introduction to Measurements & Analysis (M&A) • A complete M&A software package for Cardiac is included in the Vivid 7. • The M&A software runs on the Back-End Processor. • In this test, we just check one measurement, Calipers. 4-3-14-2 Preparations, M&A Figure 4-35 Measurement Keys Table 4-26 Measurement Key Checks StepTask Expected...
Page 329
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 65 4-3-14-3 Check Caliper Measurement, Distance 4-3-14-4 Check Post-measurement assignment labels Table 4-27 Measurement Key Checks StepTask Expected Result(s) 4. Press MEASURE on the Operator Panel. Select Caliper in the Measurement menu. or Press CALIPER on the Operator Panel. Press the CALIPER mode dependant button The Measurement menu with a list of studies is displayed in the...
Page 330
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 66 Section 4-3 - Functional Checks 4-3-15 Multi Image Checks 4-3-15-1 Introduction to Multi Image This unit has a special function for displaying tw o or four examination pictures at the same time 4-3-15-2 Preparations Use a phantom when doing these tests. 1.) Connect one of the probes, to the scanner’s left-most probe connector. -See Probe Connection , page 3-21 for info about connecting the probes -See Probes, page 9-80...