Page 271

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 7
4-2-3 Power Shut Down
When you switch off the unit, the system pe rforms an automatic shutdown sequence. 
Depending on software version, the following Shut Down modes are available:
• Standby Mode (Only available on unit s with software version v2.x and v3.x.)
• Full Shut Down
The SYSTEM - EXIT menu, used when switching of f the unit, gives you the choices described below.
Figure 4-7   System -...

Page 272

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 8 Section 4-2 - General Procedures
4-2-3 Power Shut Down (cont’d)
Standby 
NOTE: The Standby mode is only supported by  units with software version v2.x and v3.x.
Use this button to select the Standby mode, allo wing a shorter reboot time. Most of the system is 
powered down, but a certain portion of the unit remains energized. The standby mode allows a shorter 
reboot time when the system is used on a daily  basis or moved...

Page 273

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 9
4-2-3 Power Shut Down (cont’d)
Logoff 
Use this button to log off the current user. 
The system remains on and ready for a new user to log on.
(If the Logoff button is dimmed, it indicates that no  user is logged on to the unit at the moment.) 
Shutdown
Use this button to Shut Down th e system. The entire system is shut  down. It is recommended to perform 
a full shutdown at least once a week....

Page 274

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 10 Section 4-2 - General Procedures
4-2-3-1 Standby Mode NOTE: The Standby mode is supported by units with software version v2.x and v3.x ONLY.
Select Standby Mode
This procedure describes the needed steps to power down the unit to Standby mode:
NOTE: Don’t press the 
ON/STANDBY button too long. If you press the button too long (more than 3-4 
seconds), a full shutdown is started.
1.) Press the 
ON/STANDBY key on the Operator...

Page 275

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 11
4-2-3-1      Standby Mode (cont’d)
NOTE: Connect the power cable to the mains power out let again as soon as possible and at least 
before you switch on the Mains  Circuit Breaker and press the ON/STANDBY
 key.
NOTE: When turning ON a system from standby mode, it  takes a few seconds before it responds. Do not push 
the  ON/STANDBY
 button again during this period. A seco nd push will initiate a...

Page 276

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 12 Section 4-2 - General Procedures
4-2-4 Log On to the System as ‘ADM’
Table 4-2    Log on to the system as ‘ADM’ 
StepTaskExpected Result(s)
1. Press CONFIG
 on the alphanumeric keyboard. If not already logged on, the Operator Login window is displayed.
2. a) Select the down-arrow to the right of the Operator field to view  the Operator pull- down menu with a list of users-
b) Select  ADM from the pull-down menu. From the...

Page 277

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 13
4.Select  System  to select the System Setup window.
These windows (above and to the ri ght) may be slightly different, 
and in other colors, depending on the software version.The System Settings window is displayed
.
Table 4-2    Log on to the system as ‘ADM’ (cont’d)
StepTaskExpected Result(s)
System 

Page 278

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 14 Section 4-2 - General Procedures
4-2-5 Using Removable Media
4-2-5-1 Introduction Vivid 7 is equipped with a CD/DVD drive (earlier vers ions have a CD drive) and an optional Magneto 
Optical Disk Drive (MO Drive). Both drives are availabl e from the front of the unit, below the Patient I/O 
panel.
4-2-5-2 Install MO Disk in MO Drive 1.) Preparations before  installing an MO disk in the MO Drive: 
a.) The unit must be turned...

Page 279

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 15
4-2-5-3 Eject and Remove MO Disk from MO Drive 1.) Eject the MO Disk, using one of th e methods described in the following steps:
a.) In some cases, the MO Disk is ejected automatically.
b.) If not automatically ejected, use manual ejection method: 
-Select  Eject button on the screen, or
-Press  ALT 
+ E on the alphanumeric keyboard to  display the Eject Device menu.
NOTICEDo not eject the MO...

Page 280

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 16 Section 4-2 - General Procedures
4-2-5-3      Eject and Remove MO Disk from MO Drive (cont’d) 2.) Select 5 1/4 MO Disk  in the Eject Device menu to eject the MO Disk.
3.) Remove the ejected MO Disk from the MO Drive and  store it in it’s cover box in a safe place.
4-2-5-4 Install CD-R Disc in CD Drive NOTE: CD-RW discs are NOT supported by Vivid 7, but CD-R discs may be used. 1.) Press the Eject button on the right side  of...
Start reading GE Vivid 7 Service Manual