GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 271
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 7 4-2-3 Power Shut Down When you switch off the unit, the system pe rforms an automatic shutdown sequence. Depending on software version, the following Shut Down modes are available: • Standby Mode (Only available on unit s with software version v2.x and v3.x.) • Full Shut Down The SYSTEM - EXIT menu, used when switching of f the unit, gives you the choices described below. Figure 4-7 System -...
Page 272
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 8 Section 4-2 - General Procedures 4-2-3 Power Shut Down (cont’d) Standby NOTE: The Standby mode is only supported by units with software version v2.x and v3.x. Use this button to select the Standby mode, allo wing a shorter reboot time. Most of the system is powered down, but a certain portion of the unit remains energized. The standby mode allows a shorter reboot time when the system is used on a daily basis or moved...
Page 273
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 9 4-2-3 Power Shut Down (cont’d) Logoff Use this button to log off the current user. The system remains on and ready for a new user to log on. (If the Logoff button is dimmed, it indicates that no user is logged on to the unit at the moment.) Shutdown Use this button to Shut Down th e system. The entire system is shut down. It is recommended to perform a full shutdown at least once a week....
Page 274
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 10 Section 4-2 - General Procedures 4-2-3-1 Standby Mode NOTE: The Standby mode is supported by units with software version v2.x and v3.x ONLY. Select Standby Mode This procedure describes the needed steps to power down the unit to Standby mode: NOTE: Don’t press the ON/STANDBY button too long. If you press the button too long (more than 3-4 seconds), a full shutdown is started. 1.) Press the ON/STANDBY key on the Operator...
Page 275
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 11 4-2-3-1 Standby Mode (cont’d) NOTE: Connect the power cable to the mains power out let again as soon as possible and at least before you switch on the Mains Circuit Breaker and press the ON/STANDBY key. NOTE: When turning ON a system from standby mode, it takes a few seconds before it responds. Do not push the ON/STANDBY button again during this period. A seco nd push will initiate a...
Page 276
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 12 Section 4-2 - General Procedures 4-2-4 Log On to the System as ‘ADM’ Table 4-2 Log on to the system as ‘ADM’ StepTaskExpected Result(s) 1. Press CONFIG on the alphanumeric keyboard. If not already logged on, the Operator Login window is displayed. 2. a) Select the down-arrow to the right of the Operator field to view the Operator pull- down menu with a list of users- b) Select ADM from the pull-down menu. From the...
Page 277
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 13 4.Select System to select the System Setup window. These windows (above and to the ri ght) may be slightly different, and in other colors, depending on the software version.The System Settings window is displayed . Table 4-2 Log on to the system as ‘ADM’ (cont’d) StepTaskExpected Result(s) System
Page 278
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 14 Section 4-2 - General Procedures 4-2-5 Using Removable Media 4-2-5-1 Introduction Vivid 7 is equipped with a CD/DVD drive (earlier vers ions have a CD drive) and an optional Magneto Optical Disk Drive (MO Drive). Both drives are availabl e from the front of the unit, below the Patient I/O panel. 4-2-5-2 Install MO Disk in MO Drive 1.) Preparations before installing an MO disk in the MO Drive: a.) The unit must be turned...
Page 279
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 15 4-2-5-3 Eject and Remove MO Disk from MO Drive 1.) Eject the MO Disk, using one of th e methods described in the following steps: a.) In some cases, the MO Disk is ejected automatically. b.) If not automatically ejected, use manual ejection method: -Select Eject button on the screen, or -Press ALT + E on the alphanumeric keyboard to display the Eject Device menu. NOTICEDo not eject the MO...
Page 280
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 16 Section 4-2 - General Procedures 4-2-5-3 Eject and Remove MO Disk from MO Drive (cont’d) 2.) Select 5 1/4 MO Disk in the Eject Device menu to eject the MO Disk. 3.) Remove the ejected MO Disk from the MO Drive and store it in it’s cover box in a safe place. 4-2-5-4 Install CD-R Disc in CD Drive NOTE: CD-RW discs are NOT supported by Vivid 7, but CD-R discs may be used. 1.) Press the Eject button on the right side of...