Page 231
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 107 3-10-10-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the TCP/IP connection to the server. - DICOM Ping is used to verify that the DICOM application is answering 1.) Select QueryRetrieve from the Selected devices list 2.) Select Check. - If the test pass, a...
Page 232
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 108 Section 3-10 - Connectivity Setup - Software v3.x 3-10-10-3 Query/Retrieve Verification (cont’d) Follow the Steps Below to Test the DICOM Query/Retrieve Workflow (cont’d) - If the test fail, a pop-up warning is displayed, see left illustration in Figure 3-115. Select OK to continue. A red “X” to the left of the Selected devices indicates that the test failed. If Retrieve is Failing ... If Retrieve is failing...
Page 233
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 109 Section 3-11 Connectivity Setup - Software v2.x NOTE: FMI 76037 updates softwa re versions v2.3.0 and below to Software version v2.3.1. NOTE: If connected as a stand-alone network, only scanner, EchoPAC PC Workstation and an optional network printer in a separate network, you should use default delivery settings. 3-11-1 Introduction To be able to use the network functions when con nected to a...
Page 234
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 110 Section 3-11 - Connectivity Setup - Software v2.x 3-11-4 Select TCP/IP Set-up Screen - Software v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . Figure 3-116 Measuring and Analyzes - Global Level Connectivity
Page 235
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 111 3-11-4 Select TCP/IP Set-up Screen - Software v2.x (cont’d) 2.) Select CONNECTIVITY (in the lower part of the window). 3.) Select the TCP/IP TAB (it is named Tcpip). (See example in Figure 3-118.) Figure 3-117 Connectivity Opening Screen Figure 3-118 TCP/IP Set-up Screen, Overview Connectivity TCP/IP Tab IP settings area for the scanner Remote Archive Setup area
Page 236
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 112 Section 3-11 - Connectivity Setup - Software v2.x 3-11-5 Set the Scanner’s Network Information - Software Version v2.x In the IP settings area of the screen, enter the; • IP Address for scanner. (Default IP Address from factory: 10.0.0.3). • Subnet Mask for scanner. (Default Subn et Mask from factory: 255.255.255). • IP address for Default Gateway. (Default Subnet Mask from factory: 0.0.0.0). NOTE: For software v2.x, the...
Page 237
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 113 3-11-6 Set the Remote Archive’s Ne twork Information - Software v2.x In the Remote Archive Setup area of the screen (see example in Figure 3-118.), enter the; 1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4). 2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001). . 3-11-7 Save the New Settings - Software v2.x 1.) Press SAVE SETTINGS to save...
Page 238
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 114 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 2.) Select CONNECTIVITY (in the lower part of the window). 3.) Select the VIEWS tab, located in upper part of window, to display the Current Dataflow Views screen, see Figure 3-121 . 4.) Select...
Page 239
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 3 - System Setup 3 - 115 3-11-8 Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d) . The Remote Archive - Remote HD includes an EchoPAC PC with one Service also called Remote Archive - Remote HD . The name of this service indicates that the archive is located on a hard disk on a remote server (EchoPAC PC is used as an server). 6.) Select the EchoPAC PC’s name to highlight it, see Figure...
Page 240
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 3 - 116 Section 3-11 - Connectivity Setup - Software v2.x 3-11-8 Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) • In the example in Figure 3-125, the red “X” next to the computer sign ( ), means that the check failed. The cause may be; - the setup of the network information, either on Vivid 7 or on the EchoPAC PC, or - the network connection is broken somewhere between the Vivid 7 and the EchoPAC...