DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 107
3-10-10-3      Query/Retrieve Verification (cont’d)
Follow the Steps Below to Test th e DICOM Query/Retrieve Workflow 
NOTE: This check uses both Ping and DICOM Ping. - Ping is used to verify the  TCP/IP connection to the server.
- DICOM Ping is used to verify that  the DICOM application is answering
1.) Select QueryRetrieve from the Selected devices list
2.) Select Check.  - If the test pass, a...

Page 232

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 108 Section 3-10 - Connectivity Setup - Software v3.x
3-10-10-3      Query/Retrieve Verification (cont’d)       Follow the Steps Below to Test  the DICOM Query/Retrieve Workflow (cont’d)
- If the test fail, a pop-up warning is  displayed, see left illustration in Figure 3-115. 
Select OK to continue. A red “X” to the left of the Selected devices indicates that the test failed.
If Retrieve is Failing ...
If Retrieve is failing...

Page 233

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 109
Section 3-11
Connectivity Setup - Software v2.x
NOTE: FMI 76037 updates softwa re versions v2.3.0 and below to Software version v2.3.1.
NOTE: If connected as a stand-alone network, only  scanner, EchoPAC PC Workstation and an optional 
network printer in a separate network, you should use default delivery settings. 
3-11-1 Introduction
To be able to use the network functions when con nected to a...

Page 234

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 110 Section 3-11 - Connectivity Setup - Software v2.x
3-11-4 Select TCP/IP Set-up  Screen - Software v2.x
1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 .
Figure 3-116   Measuring and Analyzes - Global Level
Connectivity 

Page 235

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 111
3-11-4      Select  TCP/IP Set-up Screen -  Software v2.x (cont’d)
2.) Select CONNECTIVITY (in the lower part of the window).
3.) Select the TCP/IP TAB (it is named Tcpip). (See example in  Figure 3-118.)
Figure 3-117   Connectivity Opening Screen
Figure 3-118   TCP/IP Set-up Screen, Overview
Connectivity
TCP/IP Tab
IP settings area for 
the scanner
Remote Archive 
Setup area 

Page 236

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 112 Section 3-11 - Connectivity Setup - Software v2.x
3-11-5 Set the Scanner’s  Network Information - Software Version v2.x
In the IP settings area of the screen, enter the;
• IP Address for scanner. (Default IP  Address from factory: 10.0.0.3).
• Subnet Mask for scanner. (Default Subn et Mask from factory: 255.255.255).
• IP address for Default Gateway. (Default  Subnet Mask from factory: 0.0.0.0).
NOTE: For software v2.x, the...

Page 237

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 113
3-11-6 Set the Remote Archive’s Ne twork Information - Software v2.x
In the Remote Archive Setup area of the screen (see example in  Figure 3-118.), enter the;
1.) Remote Archive IP address. (Default  IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Ar chive Name from factory: ECHOPAC7-000001).
.
3-11-7 Save the New Settings - Software v2.x
1.) Press SAVE SETTINGS to save...

Page 238

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 114 Section 3-11 - Connectivity Setup - Software v2.x
3-11-8 Verification of a Connection Set  Up to EchoPAC PC - Software v2.x
1.) Press CONFIG (F2) and log on as  adm, see  4-2-4 Log On to the System as ‘ADM’ on page 4-12 . 
2.) Select  CONNECTIVITY
 (in the lower part of the window).
3.) Select the  VIEWS
 tab, located in upper part of window,  to display the Current Dataflow Views 
screen, see  Figure 3-121 .
 
4.) Select...

Page 239

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 3 - System Setup 3 - 115
3-11-8      Verification of a Connection Se t Up to EchoPAC PC - Software v2.x (cont’d)
.
The Remote Archive - Remote HD includes an EchoPAC PC with one Service also called  Remote 
Archive - Remote HD . The name of this service indicates that the archive is located on a hard disk on 
a remote server (EchoPAC PC is used as an server). 
6.) Select the EchoPAC PC’s name to highlight it, see  Figure...

Page 240

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
3 - 116 Section 3-11 - Connectivity Setup - Software v2.x
3-11-8      Verification of a Connection Set Up to EchoPAC PC - Software v2.x (cont’d) • In the example in  Figure 3-125, the red “X” next to the computer sign ( ), means that the check 
failed. The cause may be;
- the setup of the network information, either on Vivid 7 or on the EchoPAC PC, or
- the network connection is broken somewhere between the Vivid 7 and the EchoPAC...