GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 17 4-2-7 Formatting Removable Media - Software Version v3.x and above This procedure describes how to format an MO disk or a CD-R disc. NOTE: CD-RW discs are NOT supported by Vivid 7, but CD-R discs may be used. Before you continue, be sure ab out what naming convention you will use for labelling the Removable Media. 1.) Select Config > Connectivity > Tools 2.) Select the removable media...
Page 282
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 18 Section 4-2 - General Procedures 4-2-8 Formatting Removable Me dia - Software Version v2.x NOTE: When formatting the MO disks, give them logi cal names and make sure to physically label the disks. - A labelling convention should be followed so that each MO disk gets a unique label. - A maximum numbers of eleven (11) label characters are allowed in the label name. - Use different labels on the MO disk’s A side and B...
Page 283
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 19 4-2-10 Archiving and Loading Preset s - Software Version v2.x/v3.x NOTE: Always save presets before any software reload. This action is necessary to ensure that the presets loaded after the software reload are as up–to–date as possible. 4-2-10-1 Archiving Presets to an MO Disk NOTICEPresets shouldn’t be saved on the same MO disk as images, because the Archive Menu lists the images but not...
Page 284
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 20 Section 4-2 - General Procedures 4-2-10-2 Loading Presets from an MO Disk NOTE: It is not possible to move presets between units with different software versions. NOTE: You can successfully copy presets from one V7 to another, provided they run the same software version. Table 4-4 Loading Presets STEPTASKFUNCTION 1. Insert the MO disk with the archived presets into the MOD. 2. Go to Config (F2) > Admin > Restore The...
Page 285
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 21 4-2-11 Space Management - Softwar e version v7.x/v6.x/v5.x/v4.x 4-2-11-1 Configure the Disk Management Function NOTE: Configuration of the Disk management system can only be done by user with administration rights. 1.) Press Config (F2). 2.) Log on as adm (administrator). 3.) Select the category Admin. 4.) In the Admin category, select the Disk Management tab. 4-2-11-2 Disk Management...
Page 286
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 22 Section 4-2 - General Procedures 4-2-11-3 Data Management Settings 1.) Select a number of days, weeks or months or a specific date next to Manage files older than. Only files older than the specif ied setting will be copied or moved. If none is selected, all files w ill be copied or moved. 2.) Next to Operation check: • Copy : the images and reports from the examinations older than the specified setting defined in step...
Page 287
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 23 4-2-12 Backup - Software Ve rsion v7.x/v6.x/v5.x/v4.x 1.) Press Archive. The Operator login window is displayed. 2.) Log on as Adm. The Search/Create patient window is displayed. 3.) In the Search/Create patient window, select the dataflow Local Archive - Int. HD. 4.) Press Config (F2) . 5.) Select Admin. 6.) Select the Backup tab. 7.) In the Backup sheet select as needed: -...
Page 288
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 24 Section 4-2 - General Procedures 4-2-12 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d) 10.) Select Start backup . The following situations may occur: - The system is checking that the removable me dia is inserted. If not, a dialogue window is displayed prompting the user to insert a media. - Insert the media and select OK. - The system is checking if the media needs to be formatted. If yes, the media is...
Page 289
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 4 - Functional Checks 4 - 25 4-2-12 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d) - The system is checking if there is already a backup or a Disk management copy on the media. If the following error message is displayed, the di sk is ejected and the user is asked to use a new media that does not contain any backup or Disk management data. Insert a new media and select OK. NOTE: To reuse a Backup media when...
Page 290
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 4 - 26 Section 4-2 - General Procedures 4-2-13 Moving and Tr ansporting the Vivid 7 CAUTIONVIVID 7 weighs 190 kg (419 lbs) or more, depending on installed peripherals, when ready for use. Care must be used when moving it or replacing its parts. Failure to follow the precautions listed below could result in injury, uncontrolled motion and costly damage. ALWAYS: - Be sure the pathway is clear. - Use slow, careful motions. - Use two...