Page 281

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 17
4-2-7 Formatting Removable Media -  Software Version v3.x and above
This procedure describes how to format an MO disk or a CD-R disc. 
NOTE: CD-RW discs are NOT  supported by Vivid 7, but CD-R discs may be used.
Before you continue, be sure ab out what naming convention you  will use for labelling the Removable 
Media. 
1.) Select  Config > Connectivity > Tools
2.) Select the removable media...

Page 282

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 18 Section 4-2 - General Procedures
4-2-8 Formatting Removable Me dia - Software Version v2.x
NOTE: When formatting the MO disks, give them logi cal names and make sure to physically label the 
disks. 
- A labelling convention should be followed so  that each MO disk gets a unique label. 
- A maximum numbers of eleven (11) label characters are allowed in the label name. 
- Use different labels on the MO disk’s A side and B...

Page 283

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 19
4-2-10 Archiving and Loading Preset s - Software Version v2.x/v3.x
NOTE: Always save presets before  any software reload. This action is necessary to ensure that the 
presets loaded after the software reload are as up–to–date as possible.
4-2-10-1 Archiving Presets to an MO Disk
NOTICEPresets shouldn’t be saved on the same MO disk  as images, because the Archive Menu lists the 
images but not...

Page 284

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 20 Section 4-2 - General Procedures
4-2-10-2 Loading Presets from an MO Disk NOTE: It is not possible to move presets between units with different software versions. 
NOTE: You can successfully copy presets from one V7  to another, provided they run the same software 
version.
Table 4-4    Loading Presets 
STEPTASKFUNCTION
1. Insert the MO disk with the archived 
presets into the MOD.
2.
Go to Config (F2) > Admin > Restore The...

Page 285

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 21
4-2-11 Space Management - Softwar e version v7.x/v6.x/v5.x/v4.x
4-2-11-1 Configure the Disk Management Function
NOTE: Configuration of the Disk management system can only be done by user with administration  rights.
1.)
Press Config (F2).
2.) Log on as  adm (administrator).
3.) Select the category  Admin.
4.) In the  Admin category, select the  Disk Management tab.
4-2-11-2 Disk Management...

Page 286

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 22 Section 4-2 - General Procedures
4-2-11-3 Data Management Settings 1.) Select a number of days, weeks or months or a specific date next to  Manage files older than. 
Only files older than the specif ied setting will be copied or moved.
If  none  is selected, all files w ill be copied or moved.
2.) Next to  Operation check:
• Copy
: the images and reports from the examinations older than the specified setting 
defined in step...

Page 287

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 23
4-2-12 Backup - Software Ve rsion v7.x/v6.x/v5.x/v4.x
1.) Press Archive. The Operator login  window is displayed. 
2.) Log on as  Adm. The  Search/Create patient  window is displayed. 
3.) In the  Search/Create patient  window, select the dataflow  Local Archive - Int. HD. 
4.) Press  Config (F2) . 
5.) Select  Admin. 
6.) Select the  Backup tab. 
7.) In the  Backup sheet select as needed: 
-...

Page 288

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 24 Section 4-2 - General Procedures
4-2-12 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d)
10.) Select Start backup . The following situations may occur: 
- The system is checking that the removable me dia is inserted. If not, a dialogue window is 
displayed prompting the user to insert a media.
- Insert the media and select  OK. 
- The system is checking if the media needs to be  formatted. If yes, the media is...

Page 289

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 4 - Functional Checks 4 - 25
4-2-12 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d)
- The system is checking if there is already  a backup or a Disk management copy on the media. 
If the following error message is displayed, the di sk is ejected and the user is asked to use a 
new media that does not contain any backup or Disk management data. 
Insert a new media and select  OK.
NOTE: To reuse a Backup media when...

Page 290

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
4 - 26 Section 4-2 - General Procedures
4-2-13 Moving and Tr ansporting the Vivid 7
CAUTIONVIVID 7 weighs 190 kg (419 lbs) or more, depending on installed peripherals, when ready for use. 
Care must be used when moving it or replacing its parts. Failure to follow the precautions listed 
below could result in injury, uncontrolled motion and costly damage.
ALWAYS:
- Be sure the pathway is clear.
- Use slow, careful motions.
- Use two...