Home > GE > Ultrasound System > GE Vivid 7 Service Manual

GE Vivid 7 Service Manual

Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.

Page 361

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 15
5-3-10-2 Location in the Unit
5-3-10-3 Input DC Voltages
 
NOTICEPlease be careful about the positions when replacing  the TX card and the RX card. The position of the 
two cards are interchanged between FEP1 and FEP2.
Figure 5-11   TX128-5 Board: Location in Unit
Table 5-4    Input, DC Voltages
INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
+6 Va DC voltage, distributed via...

Page 362

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 16 Section 5-3 - a  Front-End Processor (FEP)
5-3-10-4 Control Signals
5-3-10-5 Outputs
5-3-10-6 Fuses, Jumpers and Dip-switches None Table 5-5    Input, Control Signals 
INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
FE_D0-15 Front-End Data bus, Digital TTL J1 RFI > Backplane
FE_PA0-9 Front-End Page Address bus, Digital TTL J1 RFI > Backplane
FE_SA0-5 Front-End Device Select Address bus, Digital TTL J1 RFI > Backplane...

Page 363

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 17
5-3-10-7 LEDs on the TX128 board
 
Figure 5-12   TX128 Board: LEDs
Table 5-7    LEDs on the TX128 Board
LED #LED COLORDESCRIPTIONNORMAL STATE
LD1  - Red Board Failure Normally OFF
LD2  - Green Board OK Normally ON
LD3  - Yellow Front-End Bus Protocol Error Normally OFF
LD4 - YellowScanNormally OFF
LD5 - YellowSpa reNormally OFF
LD5LD4LD3
LD2 LD1 

Page 364

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 18 Section 5-3 - a  Front-End Processor (FEP)
5-3-10-8 Test points on the TX128 board
 
NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used 
by the transmitter circuits. (The exact [calcu lated] value is 10/101). The measured values 
depends on selected scanning mode  and the Acoustic Power adjustment.
Figure 5-13   TX128 Board: Test points
Table 5-8    Test points on the TX128-5...

Page 365

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 19
5-3-11 Transmitter Board, TX12 8 (version used before BT’05)
5-3-11-1 General Description
The Transmitter board (TX128) contains 128 individually controlled transmit channels and provides 
transmit pulses via the XDBUS board to the Relay boa rd and then to the transducer array (the probes).
Frequency-, delay- and pulswidth parameters for the TX128 board are received on the FE_BUS...

Page 366

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 20 Section 5-3 - a  Front-End Processor (FEP)
5-3-11-2 Location in the Unit
5-3-11-3 Input DC Voltages
 
NOTICEPlease be careful about the positions when replacing  the TX card and the RX card. The position of the 
two cards are interchanged between FEP1 and FEP2.
Figure 5-15   TX128 Board: Location in Unit
Table 5-9    Input, DC Voltages
INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
+6 Va DC voltage, distributed via...

Page 367

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 21
5-3-11-4 Control Signals
5-3-11-5 Outputs
5-3-11-6 Fuses, Jumpers and Dip-switches None Table 5-10    Input, Control Signals 
INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM
FE_D0-15 Front-End Data bus, Digital TTL J1 Front-End Controller/RFI > 
Backplane
FE_PA0-9 Front-End Page Address bus, Digital TTL J1 Front-End Controller/RFI > 
Backplane
FE_SA0-5 Front-End Device Select...

Page 368

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 22 Section 5-3 - a  Front-End Processor (FEP)
5-3-11-7 LEDs on the TX128 board
 
Figure 5-16   TX128 Board: LEDs
Table 5-12    LEDs on the TX128 Board
LED #LED COLORDESCRIPTIONNORMAL STATE
LD1  - Red Board Failure Normally OFF
LD2  - Green Board OK Normally ON
LD3  - Yellow Front-End Bus Protocol Error Normally OFF
LD4 - YellowScanNormally OFF
LD5 - YellowSpa reNormally OFF
LD5LD4LD3
LD2 LD1 

Page 369

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 23
5-3-11-8 Test points on the TX128 board NOTE: If the test points below have not been mounted, and you want to verify the TX voltages, you  must use the TXP test, available via the Di agnostics Test Screen. Please refer to 7-8-3 System 
Test on page 7-66 .
 
NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used 
by the transmitter...

Page 370

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 24 Section 5-3 - a  Front-End Processor (FEP)
5-3-12 Transducer Bus Boards, XD BUS
5-3-12-1 General Description The two XD BUS Boards plug into the rear edge  connectors on the Relay Board, the TX128 Board(s) 
and on the RX-128 Board. 
FEP2 ONLY:  The XD Bus boards used on units with two TX  boards, with 4D (RT3D), are not backward 
compatible and cannot be used on units with one TX card (without 4D (RT3D)). And the XD Bus...
Start reading GE Vivid 7 Service Manual

Related Manuals for GE Vivid 7 Service Manual