GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 361
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 15 5-3-10-2 Location in the Unit 5-3-10-3 Input DC Voltages NOTICEPlease be careful about the positions when replacing the TX card and the RX card. The position of the two cards are interchanged between FEP1 and FEP2. Figure 5-11 TX128-5 Board: Location in Unit Table 5-4 Input, DC Voltages INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM +6 Va DC voltage, distributed via...
Page 362
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 16 Section 5-3 - a Front-End Processor (FEP) 5-3-10-4 Control Signals 5-3-10-5 Outputs 5-3-10-6 Fuses, Jumpers and Dip-switches None Table 5-5 Input, Control Signals INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM FE_D0-15 Front-End Data bus, Digital TTL J1 RFI > Backplane FE_PA0-9 Front-End Page Address bus, Digital TTL J1 RFI > Backplane FE_SA0-5 Front-End Device Select Address bus, Digital TTL J1 RFI > Backplane...
Page 363
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 17 5-3-10-7 LEDs on the TX128 board Figure 5-12 TX128 Board: LEDs Table 5-7 LEDs on the TX128 Board LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Red Board Failure Normally OFF LD2 - Green Board OK Normally ON LD3 - Yellow Front-End Bus Protocol Error Normally OFF LD4 - YellowScanNormally OFF LD5 - YellowSpa reNormally OFF LD5LD4LD3 LD2 LD1
Page 364
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 18 Section 5-3 - a Front-End Processor (FEP) 5-3-10-8 Test points on the TX128 board NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used by the transmitter circuits. (The exact [calcu lated] value is 10/101). The measured values depends on selected scanning mode and the Acoustic Power adjustment. Figure 5-13 TX128 Board: Test points Table 5-8 Test points on the TX128-5...
Page 365
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 19 5-3-11 Transmitter Board, TX12 8 (version used before BT’05) 5-3-11-1 General Description The Transmitter board (TX128) contains 128 individually controlled transmit channels and provides transmit pulses via the XDBUS board to the Relay boa rd and then to the transducer array (the probes). Frequency-, delay- and pulswidth parameters for the TX128 board are received on the FE_BUS...
Page 366
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 20 Section 5-3 - a Front-End Processor (FEP) 5-3-11-2 Location in the Unit 5-3-11-3 Input DC Voltages NOTICEPlease be careful about the positions when replacing the TX card and the RX card. The position of the two cards are interchanged between FEP1 and FEP2. Figure 5-15 TX128 Board: Location in Unit Table 5-9 Input, DC Voltages INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM +6 Va DC voltage, distributed via...
Page 367
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 21 5-3-11-4 Control Signals 5-3-11-5 Outputs 5-3-11-6 Fuses, Jumpers and Dip-switches None Table 5-10 Input, Control Signals INPUTDESCRIPTIONCONNECTOR-PIN#CONNECTION FROM FE_D0-15 Front-End Data bus, Digital TTL J1 Front-End Controller/RFI > Backplane FE_PA0-9 Front-End Page Address bus, Digital TTL J1 Front-End Controller/RFI > Backplane FE_SA0-5 Front-End Device Select...
Page 368
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 22 Section 5-3 - a Front-End Processor (FEP) 5-3-11-7 LEDs on the TX128 board Figure 5-16 TX128 Board: LEDs Table 5-12 LEDs on the TX128 Board LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Red Board Failure Normally OFF LD2 - Green Board OK Normally ON LD3 - Yellow Front-End Bus Protocol Error Normally OFF LD4 - YellowScanNormally OFF LD5 - YellowSpa reNormally OFF LD5LD4LD3 LD2 LD1
Page 369
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 23 5-3-11-8 Test points on the TX128 board NOTE: If the test points below have not been mounted, and you want to verify the TX voltages, you must use the TXP test, available via the Di agnostics Test Screen. Please refer to 7-8-3 System Test on page 7-66 . NOTE: The Voltages on these test points are approxi mately one tenth (1/10) of the TX Voltages used by the transmitter...
Page 370
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 24 Section 5-3 - a Front-End Processor (FEP) 5-3-12 Transducer Bus Boards, XD BUS 5-3-12-1 General Description The two XD BUS Boards plug into the rear edge connectors on the Relay Board, the TX128 Board(s) and on the RX-128 Board. FEP2 ONLY: The XD Bus boards used on units with two TX boards, with 4D (RT3D), are not backward compatible and cannot be used on units with one TX card (without 4D (RT3D)). And the XD Bus...