GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 381
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 35 5-3-15 Beam Former Board(s), BF-64 (cont’d) 5-3-15-7 Dip-switches and Jumpers None. 5-3-15-8 LEDs on BF64, Part Number FB200900 Figure 5-25 BF64, P/N: FB200900 - LEDs Table 5-27 LEDs on the BF-64 Boards, P/N: FB200900 LED #LED COLORDESCRIPTIONNORMAL STATE LD1 - Green “OK”. Board tested OK by System. Normally ON LD2 - Red “ERROR”. Board not initialized or an error...
Page 382
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 36 Section 5-3 - a Front-End Processor (FEP) 5-3-15-9 LEDs on BF64, Part Number: FC200100 Figure 5-26 LEDs on BF64, P/N FC200100 Table 5-28 LEDs on BF-64 Boards, P/N FC200100 LED #LED COLORDESCRIPTIONNORMAL STATE LD2 - Red “ERROR”. Board not initialized or an error has occurred Normally OFF LD1 - Green “OK”. Board tested OK by System. Normally ON LD3 - Yellow Not Used Normally OFF LD4 - YellowNot UsedNormally...
Page 383
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 37 5-3-16 Front-End Contro ller Board, FEC NOTE: The Front-End Controller Board is only used in Front-End Rack version 1. In units with Front-End Rack vers ion 2, the RFI board replaces the functionality of this board, see: 5-3-21 Radio Frequency Interface Board, RFI on page 5-61 . 5-3-16-1 General Description The Front-End Controller Board controls the transmitter and receiver...
Page 384
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 38 Section 5-3 - a Front-End Processor (FEP) 5-3-16-3 Location in the Unit Figure 5-27 FEC-2 Board: Location in Unit (Front-End Rack version 1, only) FEC-2 BoardFront of Scanner
Page 385
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 39 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-4 Inputs 5-3-16-4-1 DC Voltages 5-3-16-4-2 Signal Inputs Table 5-29 DC Input Voltages INPUTDESCRIPTIONCONNECTED TO + 3.3 Vd DC output voltage, distributed via Motherboard (Backplane). The “d” indicates that this volt age is used for digital circuits.DC Power Supply via Backplane + 5 Vd DC output voltage,...
Page 386
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 40 Section 5-3 - a Front-End Processor (FEP) 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-5 Output 5-3-16-5-1 Output Signals 5-3-16-6 LEDs Description, Main Digital Signal Processor (DSP) Table 5-31 Output Signals OUTPUTDESCRIPTIONCONNECTION TO TXTRIG_L Starts the transmitter sequence TX128 boards SYNC_L Signal used to synchronize the Beam Formers BF64 boards TSIG Test Signal RX-128 Board AT G C V P / AT G C...
Page 387
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 41 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-7 LEDs Description, Slave DSP 5-3-16-8 LEDsTable 5-33 FEC-2, Upper Batch of LEDs, Slave DSP LED #LED COLORDESCRIPTIONNORMAL STATE LD27 - Yellow Rack temperature surveillance Blinking LD28 - Yellow TX Power surveillance Blinking LD29 - Green Up and running ON Blinking LD30 - RefFailure or reprogrammingOFF Figure...
Page 388
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 42 Section 5-3 - a Front-End Processor (FEP) 5-3-16 Front-End C ontroller Board, FEC (cont’d) 5-3-16-9 Test points for DC Voltages Figure 5-29 FEC-2, Test points for DC Voltages TP8TP10 TP9 TP7 TP20 TP17TP21 +5V +15V -15V -5Va +3.3V +6Va +10Va
Page 389
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 43 5-3-16-10 Test Points for DC Voltages See Figure 5-29 FEC-2, Test points for DC Voltages on page 5-42 . 5-3-16-11 Fuses and Dip-switches None Table 5-34 Test points for DC Voltages TEST POINTDC VOLTAGEACCEPTABLE RANGE TP 8 +5 V +4.75 V to +5.25 V TP 10 +15 V +/- 10% TP 9 -15 V +/- 10% TP 7-5 Va+/- 10% TP 20+3.3 Va+/- 10% TP 17+6 Va+/- 10% TP 21 +10 Va +/- 10%
Page 390
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 44 Section 5-3 - a Front-End Processor (FEP) 5-3-16-12 Jumpers Figure 5-30 FEC-2, Jumpers and Connectors SW7 SW4 SW1 SW2 SW10 12 910 12 1112 SW3 12 3940 SW5 12 1112 Connector used in factory test Connector used in factory test Connector used in factory test Connector used in factory test