Page 351
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 5 5-2-7 Block Diagram - Units with Fr ont-End Card Cage, version 2 (FEP2) NOTE: In this block diagram, the BEP2.2 is illustrated. Figure 5-3 Vivid 7 Simple Block Diagram - Units with Front-End Card Cage, version 2 (FEP2) FRONT END PROCESSOR PROBES XDBUS DOPPLER ATX POWER SUPPLY UPS AC POWER OP I/O PANEL17 Monitor USER INTERFACE SONY COLOR PRINTERUP-2900MDPC2IO PCVIC PC...
Page 352
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 6 Section 5-3 - a Front-End Processor (FEP) Section 5-3a Front-End Processor (FEP) 5-3-1 General Description The electronics in the Front-End Card Cage is also called the Front-End Processor (FEP). Two versions of the Front-End Card Cage are in use; • Version 1 (FEP1) was used from introduction of Vivid 7 until 2nd half of 2003 • Version 2 (FEP2) was introduced 2nd half of 2003 5-3-2 FEP1 FEP1 has eight different types of...
Page 353
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 7 5-3-3 FEP2 • In units with FEP2, a new card, the Radio Frequency Interface (RFI) card replaces four of the cards used in FEP1. The cards used for FEP2 units are listed in Table 5-3 Front-End Processor Cards used in FEP2 on page 5-7 . Table 5-3 Front-End Processor Cards used in FEP2 SHORT NAME COMPLETE NAMECOMMENT RLY RELAY BOARD TX TRANSMITTER BOARD 2X TO BE USED PER...
Page 354
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 8 Section 5-3 - a Front-End Processor (FEP) 5-3-4 FEP’s Location in the Unit 5-3-5 Input Signals • RX signals from probes • Physio (ECG/Phono) 5-3-6 Bidirectional Signals PC2IP bus to BEP 5-3-7 Output Signals • TX signals to probe • Doppler and Color Flow Data to Back-End Processor for digital signal processing • Digital Video Data to Back-End Processor 5-3-8 Fuses, Jumpers, Dip-switches and LEDs See descriptions for each...
Page 355
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 9 5-3-9 Transmitter and Receiver Subsystem 5-3-9-1 Transmitter Signal Path - Units With One TX Board • The Front-End Controller board (FEP1) or the Radio Frequency Interface board (FEP2) loads scan parameters via the FE_BUS into local RAM on the TX board and on the BF boards. The scan parameters includes probe dependant st eering and focusing delay for a certain scan pattern....
Page 356
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 10 Section 5-3 - a Front-End Processor (FEP) 5-3-9-2 Transmitter Signal Path - Units With Two TX Board • The Radio Frequency Interface board (RFI) loads scan parameters via the FE_BUS into local RAM on the TX boards and on the BF boards. The scan parameters includes probe dependant st eering and focusing delay for a certain scan pattern. Thus, when the Radio Frequency Interfac e board goes through a scan sequence, it loads...
Page 357
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 11 5-3-9 Transmitter and R eceiver Subsystem (cont’d) 5-3-9-3 Phased, Linear and Convex Array Probes • The Phased, Linear and Convex array probes consis t of several identical transducer elements (e.g. 64, 128, 192). • Three probes can be connected to the system at the same time. The probe connectors are physically located on the Relay Board, where on e of them is selected and...
Page 358
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 12 Section 5-3 - a Front-End Processor (FEP) 5-3-9-5 Signal Control • The Front-End Controller (FEC) board (FEP1) or the Radio Frequency Interface (RFI) board (FEP2) controls the transmitter and receiver boards. FE C/RFI loads all parameters to the TX and Beam Former RAMs, it reads the probe identification, selects probe connector on Relay board and controls the high voltage multiple xer in linear and convex probes. In...
Page 359
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 13 5-3-9-6 Signal Path Out from the Beam Former Boards • The output from the BF#2’s Beamadders is fed to the RF and Tissue Processor board (RFT) on units with Front-End version 1, and to the Radio Frequency Interface (RFI) board on units with Front-End version 2. Figure 5-9 Receiver Signal Path RFT - Radio Frequency & Tissue SDP - Spectral Doppler IMP - Image Port...
Page 360
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 14 Section 5-3 - a Front-End Processor (FEP) 5-3-10 Transmitter Board, TX128- 5 (version introduced for BT’05) 5-3-10-1 General Description The Transmitter board (TX128) contains 128 individually controlled transmit channels and provides transmit pulses via the XDBUS board to the Relay boa rd and then to the transducer array (the probes). Frequency-, delay- and pulswidth parameters for the TX128 board are received on the...