GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 481
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 135 5-11-3-4 Outputs Figure 5-89 AC Voltage from DC Power module (to TX Power) Table 5-88 AC Voltage Output, Plug P2, 230 VAC OUTPUTDESCRIPTIONCONNECTED TO 230 VAC AC output via a three pin female (IEC) mains plug on cable. This power is output enabled by use of a relay. It is controlled by PS_ON* (active low) and PWR_OK (high) TX Power Table 5-89 DC Output Voltages, Plug...
Page 482
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 136 Section 5-11 - Power Distribution 5-11-3-5 Fuses • None. • A current limiter will switch o ff the power if + 3.3 V, +5 V or +15 V outputs are overloaded • Over-voltage Protection is provided for these voltages: - + 3.3 V -+ 5 Vd -+/- 5 Va - +/- 15 Va • Voltage will be switched off if the temperature grows too high. 5-11-3-6 Jumpers and Dip-switches None. 5-11-3-7 LEDs None. (Observe the LEDs on Internal I/O to check DC...
Page 483
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 137 5-11-4 TX Power Supply 5-11-4-1 General Description The TX Power is a switch mode power supply module. It is used to generate two symmetrical pairs of digitally programmable voltages for the transmitters in the Vivid 7 scanner. A fixed Probe MUX voltage with either +/- 95 VDC or +/- 100 VDC, depending on TX Power Supply model, is used by the multiplexers in Linear Array...
Page 484
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 138 Section 5-11 - Power Distribution 5-11-4-2 Location in the Unit Figure 5-91 TX Power Supply: Location in the Unit (below the DC Power Supply) TX Power Supply Front of scanner
Page 485
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 139 5-11-4-3 Inputs 5-11-4-4 Outputs - FC200386 Figure 5-92 AC Voltage to TX Power Table 5-90 Input, Plug P1, 230 VAC INPUTDESCRIPTIONCONNECTION FROM: 230 VAC AC Input via a three pin male (IEC) mains inlet Cable from the DC Power Module Table 5-91 Output Voltages, Plug P2 OUTPUTDESCRIPTIONCONNECTED TO: TSV1 TX Voltage 1 Voltage can be programmed to vary from +/- 2.5 V to...
Page 486
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 140 Section 5-11 - Power Distribution 5-11-4-5 Outputs - FB200574 5-11-4-6 Fuses • Output Power Protection Circuit with these functions:- Protect probe and TX board if a failure make the voltage exceed limits - Protect the power supply if there is a failure in probe or transmitter circuit • A watch-dog disables the voltage (set it to zero Volt) if the; - serial communication with the FEC-2 module stops - nominal Power limit...
Page 487
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 141 Section 5-12 Mechanical Descriptions 5-12-1 General Description, Top Console on Vivid 7 NOTE: Vivid 7 PRO is described in section 5-12-4 on page 5-145. The Top Console includes Monitor, Monitor Shelf, Speakers and Operator Panel. The Console’s movement backwards and forwards and to the sides is made possible by the Frogleg unit that links the Top Console to the rest of the S...
Page 488
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 142 Section 5-12 - Mechanical Descriptions 5-12-2 Transporting Vivid 7 The Top Console must be locked into the center (x /y) position as indicated on the label on the Front Handle, before transp orting the system. Figure 5-93 Label on Front Handle OR OR
Page 489
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 5 - Components and Functions (Theory) 5 - 143 5-12-3 Location in the Unit Figure 5-94 Top Console, upper positions Figure 5-95 Release Handles location (below the Top Console) Operator Panel Extended Frogleg Flexible harness Monitor Monitor Shelf with Speakers Z-release Handle, releases Top Console for up and down movement.XY-release Handle, releases Top Console for forward, backward and sideward movement,
Page 490
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 5 - 144 Section 5-12 - Mechanical Descriptions 5-12-3 Location in the Unit (cont’d) Figure 5-96 Top Console movement, forward and sideward Top Console in low, locked position Top Console in low, extended position Top Console turned left