Page 471

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 125
5-10-3 Modem (Option) Inputs/Outputs
• DC Voltage from Modem  Power (AC/DC Converter)
• RS232 from Back-End Processor vi a Internal I/O and External I/O
• Analog Phone line via External I/O
5-10-4 Fuses
None
5-10-5 Jumpers and Dip-switches
None Figure 5-78   Modem Option Block Diagram
External I/O
ModemInternal I/O
Connect to hospital’s 
analog phone line RS-232C
Analog phone...

Page 472

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 126 Section 5-10 - Modem (Option)
5-10-6 LEDs on Modem
The Modem has ten LEDs on its front.
 
Figure 5-79   LEDs on Modem’s Front Panel
Table 5-85    LEDs on Modem
LED COLORDESCRIPTIONNORMAL STATE
 - Red TD (Transmit Data) Flashes during use. 
 - Red RD (Receive Data) Flashes during use.
 - Red CD (Carrier Detect) ON when detecting a carrier from another modem and during 
communication.
OFF indicates no or broken connection.
 -...

Page 473

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 127
Section 5-11
Power Distribution
5-11-1 Overall AC Power Distribution
Figure 5-80   Overall Power Distribution
Select AC 
Voltage for 
Peripherals on 
AC Distribution 
Box (PWB): 
115 VAC or 230 
VAC.
Filter Mains Circuit 
Breaker
Inrush 
Limiter
115 or 230 
VAC 
Transformer
20 VAC230 VAC 115 VAC
Fan RegulatorFan
DC Power
TX Power
DC 
Voltages to 
Cards
 Front-End Processor
DC...

Page 474

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 128 Section 5-11 - Power Distribution
5-11-1 Overall AC Powe r Distribution (cont’d)
Figure 5-81   AC Cables
P5
AC
DISTRIBUTIO N BO X
XFMR BO X
P 6
P100
P101 P
7
P8
P9
P10
P11
P art o fM ain C able
To Monitor
To Colo r P rinter
To VC R
To BW Printe r
To FE P
Card Rack To BEP
INTERNAL I/O
F an
A 13 A16A 15 A12
A
14
AC CONTRO L
BO X
AC
Adapter Mode m
Optiona l G loba l
Mode m
Assembly
AC Outlet 

Page 475

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 129
5-11-2 AC Controller
5-11-2-1 General Description of the AC Controller The AC Controller’s main task is to supply the va rious internal subsystems with AC power and to 
galvanically isolate the scanner from  the on-site Mains Power System. 
An inrush current limiter reduce the peek input current  when the unit is switched on. An EMI filter helps 
to reduce EMI to acceptable...

Page 476

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 130 Section 5-11 - Power Distribution
5-11-2-2 The AC Transformer The AC Transformer is the galvanic barrier between the rest of the scanner and the on-site AC Mains.
Input voltage to the transformer can be either 2 30 VAC or 115 VAC. The transformer has two 115 VAC 
primary windings which are series- or parallel coupled depending on input voltage (see  Figure 5-83). 
Each of the primary windings has  one thermal protection...

Page 477

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 131
5-11-2-3 Distribution of Voltages via AC Controller The 20 VAC is routed via the fan regulator to the system fan.
Both 115 VAC and 230 VAC are routed via fuses.
• 115 VAC is routed to th e Monitor outlet (MONITOR)
• An AC Voltage selector is used to select the  correct voltage for the Peripherals. The selected 
voltage (230 VAC or 115 VAC) is routed  to the Peripherals AC outlet...

Page 478

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 132 Section 5-11 - Power Distribution
5-11-2-4 Location in the Unit - AC Controller and AC Transformer Figure 5-85   AC Controller
Figure 5-86   AC Transformer
AC 
TRANSFORMER 

Page 479

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 5 - Components and Functions (Theory) 5 - 133
5-11-3 DC Power
5-11-3-1 General Description • The DC Power module is a switch mode power suppl y, enclosed in a metal frame that slides into 
the Card Cage. 
• The module connects to the Motherboard (Backplane) through two F48 pin male connectors (P3  and P4) and supplies DC voltage via the Backplane, to all the boards in the Front End Card Cage 
(FEP), to the I/O and to  the...

Page 480

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 134 Section 5-11 - Power Distribution
5-11-3-3 Inputs AC Voltage. Plug P1, 230 VAC
Figure 5-88   AC Voltage input to DC Power module 
Table 5-86    Input, Plug P1, 230 VAC
INPUTDESCRIPTIONCONNECTION FROM:
230 VAC AC Input via a three pin male (IEC) mains inlet Cable from the AC Power Distribution 
Box
Table 5-87    Sense and Control Signals, Plug P3 and P4
INPUT/OUTPUTDESCRIPTIONCONNECTION FROM/TO:
+ 3.3 Sense Remote sense +3.3...