Chapter 8 - Replacement Procedures 8 - 13
8-7-3 Customer Provided Prerequisite
• Formatted and labelled removable media for Images storage
• Formatted and labelled removable media for Patient Archive and User Defined Settings.
8-7-4 Tools Provided With Un it or After an Upgrade
Table 8-1    Software used on BT’08 (Software Version v7.x)
ITEMPART NAME
PART 
NUMBER
DESCRIPTIONQTY
1 VIVID 7 WITH BEP4.3, SYSTEM SW CD FD200078System...

Page 602

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 14 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-5 Preparations
1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see  4-2-2 Power ON/ Boot UP 
on page 4-3 .
2.) Store images to removable media , see  8-7-7 Backup - Software Version v7.x/v6.x/v5.x/v4.x on 
page 8-17 .
4APPLICATION SOFTWAREFC200979APPLICATION SOFTWARE v4.2.x 
(THIS CD ALSO INCLUDE SOFTWARE v5.2.x and...

Page 603

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 15
8-7-6 Move Images - Software  Version v7.x/v6.x/v5.x/v4.x
For general information, please refer to the  Disk management description in the Vivid 7 User Manual.
8-7-6-1 Configure the Disk Management Function NOTE: Configuration of the Disk management system can only be done by user with administration  rights.
1.)
Press Config (F2).
2.) Log on as  ADM (administrator). 
- Password, software...

Page 604

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 16 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-6-3 Data Management Settings 1.) Select a number of days, weeks or months or a specific date next to  Manage files older than. 
Only files older than the specif ied setting will be copied or moved.
If  none  is selected, all files w ill be copied or moved.
2.) Next to  Operation check:
• Copy
: the images and reports from the...

Page 605

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 17
8-7-7 Backup - Software Ve rsion v7.x/v6.x/v5.x/v4.x
1.) Press Archive. The Operator login  window is displayed. 
2.) Log on as  Adm. The  Search/Create patient  window is displayed. 
3.) In the  Search/Create patient  window, select the dataflow  Local Archive - Int. HD. 
4.) Press  Config (F2) . 
5.) Select  Admin. 
6.) Select the  Backup tab. 
7.) In the  Backup sheet select as needed:...

Page 606

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 18 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-7 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d)
10.) Select Start backup . The following situations may occur: 
- The system is checking that the removable me dia is inserted. If not, a dialogue window is 
displayed prompting the user to insert a media.
- Insert the media and select  OK. 
- The system is checking if the media...

Page 607

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 19
8-7-7 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d)
- Insert a new media and select  OK.
NOTE: To reuse a Backup media when performing a new archive backup, the media has to be re- formatted first.
11.) During backup, progress windows are displayed showing the current operation being performed. 
12.) At the end of the process, the media is ejected and the  Backup completed window...

Page 608

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 20 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-8 Record SW Option Keys - Softw are Version v7.x/v6.x/v5.x/v4.x
1.) Press CONFIG (F2) and log on as  adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12.
2.) Select  Admin from the lower part of the screen.
3.) Select the  System Admin  tab.
4.) Record the SW Options Key(s) - alphanumer ic string - from the SW Options Key window....

Page 609

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 8 - Replacement Procedures 8 - 21
8-7-9 Record TCP/IP Sett ings - Software Version v7.x/v6.x/v5.x/v4.x
1.) Select Connectivity .
2.) Select the  TCPIP
 tab.
3.) Record all settings in  Table 8-6 Record settings from Tcpip screen - software version v7.x/v6.x/v5.x/
v4.x on page 8-21 .
  
Figure 8-18   TCP/IP Example Settings
Table 8-6    Record settings from Tcpip sc reen - software version v7.x/v6.x/v5.x/v4.x
PARAMETERVALUE...

Page 610

GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
8 - 22 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x
8-7-10 Record Service Settings - So ftware Version v7.x/v6.x/v5.x/v4.x
1.) Select Service.
.
2.) Record Video Setting Format and VCR Type in  Table 8-7 Video Setting and VCR type - Software 
Version v7.x/v6.x/v5.x/v4.x on page 8-22 .
 
Figure 8-19   Service Settings
Table 8-7    Video Setting and VCR type  - Software Version...