Page 601
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 13 8-7-3 Customer Provided Prerequisite • Formatted and labelled removable media for Images storage • Formatted and labelled removable media for Patient Archive and User Defined Settings. 8-7-4 Tools Provided With Un it or After an Upgrade Table 8-1 Software used on BT’08 (Software Version v7.x) ITEMPART NAME PART NUMBER DESCRIPTIONQTY 1 VIVID 7 WITH BEP4.3, SYSTEM SW CD FD200078System...
Page 602
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 14 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-5 Preparations 1.) Switch ON the Mains Power Circuit Breaker and power up the unit, see 4-2-2 Power ON/ Boot UP on page 4-3 . 2.) Store images to removable media , see 8-7-7 Backup - Software Version v7.x/v6.x/v5.x/v4.x on page 8-17 . 4APPLICATION SOFTWAREFC200979APPLICATION SOFTWARE v4.2.x (THIS CD ALSO INCLUDE SOFTWARE v5.2.x and...
Page 603
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 15 8-7-6 Move Images - Software Version v7.x/v6.x/v5.x/v4.x For general information, please refer to the Disk management description in the Vivid 7 User Manual. 8-7-6-1 Configure the Disk Management Function NOTE: Configuration of the Disk management system can only be done by user with administration rights. 1.) Press Config (F2). 2.) Log on as ADM (administrator). - Password, software...
Page 604
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 16 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-6-3 Data Management Settings 1.) Select a number of days, weeks or months or a specific date next to Manage files older than. Only files older than the specif ied setting will be copied or moved. If none is selected, all files w ill be copied or moved. 2.) Next to Operation check: • Copy : the images and reports from the...
Page 605
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 17 8-7-7 Backup - Software Ve rsion v7.x/v6.x/v5.x/v4.x 1.) Press Archive. The Operator login window is displayed. 2.) Log on as Adm. The Search/Create patient window is displayed. 3.) In the Search/Create patient window, select the dataflow Local Archive - Int. HD. 4.) Press Config (F2) . 5.) Select Admin. 6.) Select the Backup tab. 7.) In the Backup sheet select as needed:...
Page 606
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 18 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-7 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d) 10.) Select Start backup . The following situations may occur: - The system is checking that the removable me dia is inserted. If not, a dialogue window is displayed prompting the user to insert a media. - Insert the media and select OK. - The system is checking if the media...
Page 607
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 19 8-7-7 Backup - Software Versio n v7.x/v6.x/v5.x/v4.x (cont’d) - Insert a new media and select OK. NOTE: To reuse a Backup media when performing a new archive backup, the media has to be re- formatted first. 11.) During backup, progress windows are displayed showing the current operation being performed. 12.) At the end of the process, the media is ejected and the Backup completed window...
Page 608
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 20 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-8 Record SW Option Keys - Softw are Version v7.x/v6.x/v5.x/v4.x 1.) Press CONFIG (F2) and log on as adm, see 4-2-4 Log On to the System as ‘ADM’ on page 4-12. 2.) Select Admin from the lower part of the screen. 3.) Select the System Admin tab. 4.) Record the SW Options Key(s) - alphanumer ic string - from the SW Options Key window....
Page 609
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 21 8-7-9 Record TCP/IP Sett ings - Software Version v7.x/v6.x/v5.x/v4.x 1.) Select Connectivity . 2.) Select the TCPIP tab. 3.) Record all settings in Table 8-6 Record settings from Tcpip screen - software version v7.x/v6.x/v5.x/ v4.x on page 8-21 . Figure 8-18 TCP/IP Example Settings Table 8-6 Record settings from Tcpip sc reen - software version v7.x/v6.x/v5.x/v4.x PARAMETERVALUE...
Page 610
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 22 Section 8-7 - Software Loading Procedure - Software Version v7.x, v6.x, v5.x and v4.x 8-7-10 Record Service Settings - So ftware Version v7.x/v6.x/v5.x/v4.x 1.) Select Service. . 2.) Record Video Setting Format and VCR Type in Table 8-7 Video Setting and VCR type - Software Version v7.x/v6.x/v5.x/v4.x on page 8-22 . Figure 8-19 Service Settings Table 8-7 Video Setting and VCR type - Software Version...