GE Vivid 7 Service Manual
Here you can view all the pages of manual GE Vivid 7 Service Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 571
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 59 7-7-3-3 Monitor Test Double-click Monitor Test from the PC Diagnostic interactive tests menu. The monitor test dialog box presents you with three options: • Combination Test - Checks the monitor for alignment, co lor depth, and monitor resolution. It will put four cross-hairs in each corner, RGB (Red Green Blue) spectrum arrays and horizontal and vertical alignment bars on the...
Page 572
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 60 Section 7-7 - Common Diagnostics 7-7-3-3 Monitor Test (cont’d) •Solid Color Test - The solid color test is for laptop and LCD (Liquid Crystal Display) panel displays. This test displays solid color screens to allow yo u to make sure all color pixels are operating correctly. • VESA (Video Electronics Standard s Association) Test Patterns - The VESA test patterns checks luminance, geometry, and focus. The VESA tests...
Page 573
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 61 7-7-3-3 Monitor Test (cont’d) Geometry Figure 7-65 M onitor, VESA Test Pattern, Geometry
Page 574
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 62 Section 7-7 - Common Diagnostics 7-7-3-3 Monitor Test (cont’d) Focus Always shutdown the system and reboot after a diagnostics session. Figure 7-66 Monitor, VESA Test Pattern, Focus
Page 575
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 63 Section 7-8 Acquisition Diagnostics 7-8-1 Acquisition Di agnostics Introduction • If the scanner is running, but an error that may be a hardware failure, has been registered, then continue to read. 7-8-1-1 Start System Test To start the System Test, power up the scanner, see 4-2-2 - Power ON/ Boot UP on page 4-3 for detailed instructions. After scanner initializat ion is...
Page 576
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 64 Section 7-8 - Acquisition Diagnostics 7-8-2 Calibration Screen After following the steps in 7-8-1-1 Start System Test on page 7-63 , the default screen will be the Calibration screen. The lower part of the Calibration screen is different on units with Front-End Processor version 1 (FEP1) and Front-End Processor version 2 (FEP2). 7-8-2-1 Functions Available from the Calibration Screen These functions are available from...
Page 577
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 65 7-8-2-1 Functions Available from the Calibration Screen (cont’d) To start a test, select the respective button. The “LEDs” in the Calibration Screen will change color when a test has been done; • GREEN indicates that the test PASSED. • RED indicates that the test FAILED. Figure 7-69 Calibration Screen on units with FEP2 Channel Control (used for factory tests) DC Offs. Cal...
Page 578
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 66 Section 7-8 - Acquisition Diagnostics 7-8-3 System Test 7-8-3-1 Select the Test Screen From the Calibration screen, select TEST or select one of the buttons (not CAL) in the lower part of the screen. See Figure 7-70 - Select Test Window (FEP1) and Figure 7-71 - Select Test Window (FEP2). The System Test screen (Figure 7-72 /Figure 7-73 ) gives you access to the tests needed for troubleshooting the Front-End Card...
Page 579
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 7 - Diagnostics/Troubleshooting 7 - 67 7-8-3-2 The System Test Screen - FEP1 Figure 7-72 System Test Screen (FEP1) Table 7-4 Choices on the System Test Screen on units with FEP1 ITEMDESCRIPTIONCOMMENTS 1 MODULE NAME ETC. 2 SYSTEM:TEST ALL See 7-8-3-5 Test Loops, System on page 7-69 3 TEST LOOPS See 7-8-3-5 Test Loops, System on page 7-69 4BREAKSee 7-8-3-5 Test Loops, System on page 7-69 5“BLOCK DIAGRAM” WITH SINGLE...
Page 580
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 7 - 68 Section 7-8 - Acquisition Diagnostics 7-8-3-3 The System Test Screen - FEP2 Figure 7-73 System Test Screen (FEP2) Table 7-5 Choices on the System Test Screen on units with FEP2 ITEMDESCRIPTIONCOMMENTS 1 MODULE NAME ETC. 2 SYSTEM:TEST ALL See 7-8-3-5 Test Loops, System on page 7-69 3 TEST LOOPS See 7-8-3-5 Test Loops, System on page 7-69 4BREAKSee 7-8-3-5 Test Loops, System on page 7-69 5“BLOCK DIAGRAM” WITH SINGLE...