GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 83 8-11-4 AC Power Removal Procedure 1.) Disconnect Speed Control cable, Fan cable and Input Voltage cable. 2.) Remove four (4) screws holding AC Power. 3.) Turn AC Power Module to get better access and u nplug the rest of the cables. Make sure each connector and socket are properly marked, a nd easy to identify during installation. Figure 8-80 AC Power Module Fan Cable Speed Cable...
Page 672
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 84 Section 8-11 - AC Power Replacement Procedure 8-11-5 AC Power Installation Procedure 1.) Connect the cables. You may need to leave the Speed Control cable and the Fan cable unconnected until you have fastened the unit. Input Voltage connector and socket are coded to prevent mistakes. In 115V system the connector matches one end of the 8-holed socket. In 230V system it matches the other end of the socket. Figure 8-81 is...
Page 673
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 85 8-11-6 Verification - Functional Checks • Please perform the tests specified in Section 8-20 Verification - Functional Checks on page 8-105 .
Page 674
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 86 Section 8-12 - DC Power Replacement Procedure Section 8-12 DC Power Replacement Procedure 8-12-1 Manpower One person, 30 minutes 8-12-2 Tools 4 mm blade screwdriver. 8-12-3 Preparations 1.) Shut down the system, see 4-2-3 Power Shut Down on page 4-7 . 2.) Remove Right Side Cover, see Side Covers Removal Procedure on page 8-4 . 3.) Remove the Front End’s Cover. a.) Turn seven (7) quarter-turn screws to loosen Front-End...
Page 675
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 87 8-12-4 DC Power Removal Procedure 1.) Unscrew three (3) screws as shown in Figure 8-82. 2.) Unplug PCI cable from the socket on IMP2 board. 3.) Pull the DC Power a little to get access to power cables. 4.) Remove retaining clasp and unplug Power Cable and HV Cable. 5.) Remove DC Power. Figure 8-82 DC Power and Cabling. Screws HV power cable DC power cable PCI cable IMP2 Board DC Power
Page 676
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 88 Section 8-12 - DC Power Replacement Procedure 8-12-5 DC Power Installation Procedure 1.) Slide DC Power module into its guides. 2.) Plug in HV cable and Power cable and secure Power cable with the retaining clasp. 3.) Push DC Power module gently in position. 4.) Fasten three (3) screws. 5.) Connect PCI cable to IMP2 Board or RFI Board. 6.) Install Front-End Cover. 7.) Install Right Side Cover, see Side Covers Installation...
Page 677
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 89 Section 8-13 TX Power Replacement Procedure 8-13-1 Manpower One person, 30 minutes 8-13-2 Tools • Phillips screwdriver 2 • 4 mm blade screwdriver 8-13-3 Preparations 1.) Shut down the system, see 4-2-3 Power Shut Down on page 4-7 . 2.) Remove Right Side Cover, see Side Covers Removal Procedure on page 8-4 . 3.) Remove the Front-End Door. a.) Turn seven (7) quarter-turn screws to loosen...
Page 678
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 90 Section 8-13 - TX Power Replacement Procedure 8-13-4 TX Power Removal Procedure 1.) Remove retaining clasp from DC Power connector. 2.) Unplug DC Power and TX Power. 3.) Unscrew two (2) screws in front of TX Power. . 4.) Push cables aside and pull out TX Power from its socket. 8-13-5 TX Power Installation Procedure 1.) Push cables aside and install TX Power. 2.) Fasten two (2) screws. 3.) Connect DC Power Cable and secure...
Page 679
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL Chapter 8 - Replacement Procedures 8 - 91 Section 8-14 Rear Casters Repl acement Procedure 8-14-1 Manpower 2 persons, 30 minutes (one Caster) 8-14-2 Tools • Phillips screwdriver size 2 • 5 mm hexagon key • Bevel Edged Board and Wooden Wedge (service parts, see Table 9-77 Service Kits, Overview on page 9-99 .) 8-14-3 Preparations 1.) Shut down the system, see 4-2-3 Power Shut Down on page 4-7 . 2.) Remove the Side Covers, see...
Page 680
GE HEALTHCARE DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL 8 - 92 Section 8-14 - Rear Casters Replacement Procedure 8-14-4 Rear Casters Removal Procedure 1.) Turn Front Casters as shown in Figure 8-87. 2.) Activate Direction Lock. 3.) Put the bevel-edged board on the floor. 4.) Pull the system backwards up th e board incline. This will lift the Rear Wheel on the opposite side of the System from the floor. WARNINGWARNINGPrior to elevating scanner, verify that the floating Operator...