GE Logiq P5 Service Manual
Here you can view all the pages of manual GE Logiq P5 Service Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 371
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 125 8-3-5-3 Printer setting Back-up (cont’d) 6.) Select removable disk or other media. 7.) Click the “Save” to write the printer configuration. Figure 8-172 Printer set backup Figure 8-173 Printer set backup
Page 372
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 126 Section 8-3 - Software Loading Procedure 8-3-5-3 Printer setting Back-up (cont’d) 8.) Click the “OK” until all printer windows are closed. 9.) If Sony D895 is used, repeat the step1 to setp4 and write down the Orientation set. Figure 8-174 Printer set backup Figure 8-175 Printer set backup
Page 373
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 127 8-3-5-3 Printer setting Back-up (cont’d) 10.) Click the “Advanced”, then writhe down the Paper size and scaling parameters. 11.) Click “Properties” of the “Density Adjust”. Figure 8-176 Printer set backup ParametersDescriptions Orientation Portrait or Landscape Paper size Paper size Scaling Rate of length and width Table 8-52 Parameters Figure 8-177 Printer set backup
Page 374
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 128 Section 8-3 - Software Loading Procedure 8-3-5-3 Printer setting Back-up (cont’d) 12.) Click “Save” -> select a remo vable disk or other media -> save the printer configuration. 13.) Click the “OK” until all printer windows are closed. 14.) If the printer does not support the configuration save or write option, write down or print out the following parameters. NOTE: The parameters depend on printers, so the...
Page 375
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 129 Section 8-4Software Installation 8-4-1 Installing Base Image Software NOTE: While we do not believe it is necessary, It would not hurt to disconnect the system from the network and remove all transducers. NOTE: All digital peripherals, reporting systems conn ected via the USB ports should be either power off or disconnected prior to the software installation 1.) Power ON the scanner....
Page 376
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 130 Section 8-3 - Software Loading Procedure 8-4-1 Installing Base Image Software (cont’d) 4.) If the BIOS setting value in Primary IDE Slave is ATAPI CDROM , the 1st Boot Device in Boot menu should be Primary Slave. If the 1st Boot Device is not Primary Slave, please change the setting value using Enter key and Arrow key. 5.) If the BIOS setting value in Secondary IDE Master is ATAPI CDROM, the 1st Boot...
Page 377
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 131 8-4-1 Installing Base Image Software (cont’d) 6.) Insert the BT11 Base Image Software Disk into t he CD/DVD ROM drive. 7.) After restarting th e system, the following sc reen will be displayed. Figure 8-183 Boot Device Priority Figure 8-184 Installation Warning
Page 378
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 132 Section 8-3 - Software Loading Procedure 8-4-1 Installing Base Image Software (cont’d) 8.) Press Enter key and then the following screen will be shown. (If you want to abort installation procedures, press CTRL+C.) 9.) Press A key to install the complete disk. Figure 8-185 Base Image Software Inst allation Menu NOTICESelect A when installing the new HDD or upgrading from BT09. (HDD capacity should be...
Page 379
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-3 - Software Loading Procedure 8 - 133 8-4-1 Installing Base Image Software (cont’d) 10.) Approx. twenty(20) minutes later, the following screen appears. Eject the Base Image Software Disk from the CD/DVD ROM drive. 11.) After removing the disk from the drive, Press ALT + CTRL + DEL key. NOTE: After Installing Base Image Software, the “General failure reading drive A. Abort, Retry, Fail?” message can be...
Page 380
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 134 Section 8-3 - Software Loading Procedure 8-4-2 Verifying Base Image Software 8-4-2-1 Partition Checks 1.) Reboot the SYSTEM and the Desktop screen will be shown as blow. 2.) Click Start button and Run menu, then type explorer and click OK 3.) Right-click My Computer icon on the Desktop and click Manage. Figure 8-188 Desktop Screen Figure 8-189 Executing Windows Explorer Figure 8-190 Opening the Manage Window