Page 321

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 75
8-2-30 P3RLY / P2RLY Assy & P2RLY with Dummy 
8-2-30-1 Tools • Common pilIips screwdrivers
8-2-30-2 Preparations • Shut down the system and switch off the main breaker.
8-2-30-3 Removal procedure 1.) Remove the side right cover. Refer to the  8-2-16 Side Right Cover on page 8-52.
2.) Remove the side left cover. Refer to the  8-2-15 Side Left Cover on page 8-51.
3.) Remove the Front Cover....

Page 322

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 76 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-30-3              Removal procedure  (cont’d)
7.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-30-4 Mounting Procedure Install the new parts in the reverse order of removal.
Table 8-31    Functional Tests
Service Manual
Section
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 323

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 77
8-2-31 Backplane Assy 
8-2-31-1 Tools • Common pilIips screwdrivers
8-2-31-2 Preparations • Shut down the system and switch off the main breaker.
8-2-31-3 Removal procedure 1.) To detached the backplane, all pwas in the cardcage should be removed from cardcage.
2.) Open the Side Left Cover and Side Right Cover. Refer to the  8-2-15 Side Left Cover on page 8-
51  and  8-2-16 Side Right Cover...

Page 324

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 78 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-31-3              Removal procedure  (cont’d)
9.) Unscrew 16 screws from the Backplane Assy.
10.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-31-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-114   Backplane
Table 8-32    Functional Tests
Service Manual Section
Functional Test /...

Page 325

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 79
8-2-32 ACWD Assy 
8-2-32-1 Tools • Common pilIips screwdrivers
8-2-32-2 Preparations • Shut down the system and switch off the main breaker.
8-2-32-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the  8-2-16 Side Right Cover on page 8-52.
2.) Remove the EMI Cover R. Refer to the  8-2-27 EMI Cover R on page 8-69.
3.) Remove the ASIG Board. Refer to the 8-2-35 ASIG Assy on page...

Page 326

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 80 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-32-3              Removal procedure  (cont’d)
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-32-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-33    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 327

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 81
8-2-33 SYSCON PM As sy/SYSCON CM Assy
8-2-33-1 Tools
• Common pilIips screwdrivers
8-2-33-2 Preparations • Shut down the system and switch off the main breaker.
8-2-33-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the  8-2-16 Side Right Cover on page 8-52.
2.) Remove the EMI Cover R. Refer to the  8-2-27 EMI Cover R on page 8-69.
3.) Remove the SYSCONPM(SYSCONCM) assy  in...

Page 328

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 82 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-33-3              Removal procedure  (cont’d)
5.) Mount ASIR ROM to new SYSCONML  Assy. Refer to the Figure 8-119.
6.) Confirm mounting direction and pin number.
NOTE: It is important to swap ASIR  ROM(U23). Because it contains the serial Number and option the 
system. Figure 8-118   Separation of ASIR ROM
Figure 8-119   ASIR ROM mounting direction 

Page 329

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 83
8-2-33-3              Removal procedure  (cont’d)
7.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-33-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-34    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 330

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 84 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-34 PM SOM Assy with  1G/ CM SOM Assy with 512M
8-2-34-1 Tools
• Common pilIips screwdrivers
8-2-34-2 Preparations • Shut down the system and switch off the main breaker.
8-2-34-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the  8-2-16 Side Right Cover on page 8-52.
2.) Remove the EMI Cover R. Refer to the  8-2-27 EMI Cover R on page 8-69.
3.) Remove the SYSCOM PM  / CM...