Page 321
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 75 8-2-30 P3RLY / P2RLY Assy & P2RLY with Dummy 8-2-30-1 Tools • Common pilIips screwdrivers 8-2-30-2 Preparations • Shut down the system and switch off the main breaker. 8-2-30-3 Removal procedure 1.) Remove the side right cover. Refer to the 8-2-16 Side Right Cover on page 8-52. 2.) Remove the side left cover. Refer to the 8-2-15 Side Left Cover on page 8-51. 3.) Remove the Front Cover....
Page 322
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 76 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-30-3 Removal procedure (cont’d) 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-30-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-31 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 323
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 77 8-2-31 Backplane Assy 8-2-31-1 Tools • Common pilIips screwdrivers 8-2-31-2 Preparations • Shut down the system and switch off the main breaker. 8-2-31-3 Removal procedure 1.) To detached the backplane, all pwas in the cardcage should be removed from cardcage. 2.) Open the Side Left Cover and Side Right Cover. Refer to the 8-2-15 Side Left Cover on page 8- 51 and 8-2-16 Side Right Cover...
Page 324
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 78 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-31-3 Removal procedure (cont’d) 9.) Unscrew 16 screws from the Backplane Assy. 10.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-31-4 Mounting Procedure Install the new parts in the reverse order of removal. Figure 8-114 Backplane Table 8-32 Functional Tests Service Manual Section Functional Test /...
Page 325
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 79 8-2-32 ACWD Assy 8-2-32-1 Tools • Common pilIips screwdrivers 8-2-32-2 Preparations • Shut down the system and switch off the main breaker. 8-2-32-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 Side Right Cover on page 8-52. 2.) Remove the EMI Cover R. Refer to the 8-2-27 EMI Cover R on page 8-69. 3.) Remove the ASIG Board. Refer to the 8-2-35 ASIG Assy on page...
Page 326
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 80 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-32-3 Removal procedure (cont’d) 5.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-32-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-33 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 327
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 81 8-2-33 SYSCON PM As sy/SYSCON CM Assy 8-2-33-1 Tools • Common pilIips screwdrivers 8-2-33-2 Preparations • Shut down the system and switch off the main breaker. 8-2-33-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 Side Right Cover on page 8-52. 2.) Remove the EMI Cover R. Refer to the 8-2-27 EMI Cover R on page 8-69. 3.) Remove the SYSCONPM(SYSCONCM) assy in...
Page 328
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 82 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-33-3 Removal procedure (cont’d) 5.) Mount ASIR ROM to new SYSCONML Assy. Refer to the Figure 8-119. 6.) Confirm mounting direction and pin number. NOTE: It is important to swap ASIR ROM(U23). Because it contains the serial Number and option the system. Figure 8-118 Separation of ASIR ROM Figure 8-119 ASIR ROM mounting direction
Page 329
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 83 8-2-33-3 Removal procedure (cont’d) 7.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. 8-2-33-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-34 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 330
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 84 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-34 PM SOM Assy with 1G/ CM SOM Assy with 512M 8-2-34-1 Tools • Common pilIips screwdrivers 8-2-34-2 Preparations • Shut down the system and switch off the main breaker. 8-2-34-3 Removal procedure 1.) Remove the Side Right Cover. Refer to the 8-2-16 Side Right Cover on page 8-52. 2.) Remove the EMI Cover R. Refer to the 8-2-27 EMI Cover R on page 8-69. 3.) Remove the SYSCOM PM / CM...