GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 105 8-2-47-3 Removal procedure (cont’d) 4.) Unplug the SATA cable from the back plane on the bottom of backplane. 5.) Disconnect the SATA Power Cable and SATA Cable from HDD Assy after removing 2ea cable ties. And unscrew 4screws to remove the HDD Fixture bracket. Figure 8-144 Unplug the SATA Cable Figure 8-145 Separate HDD and HDD Fixture Bracket Cable Tie
Page 352
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 106 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-47-3 Removal procedure (cont’d) 6.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-47-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-48 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 353
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 107 8-2-48 ACPC Assy 8-2-48-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-48-2 Preparations • Shut down the system and switch off the main breaker. 8-2-48-3 Removal procedure 1.) Remove the AC Input Box Assy. Refer to the 8-2-49 Transbox Assy 110V/220V on page 8-109 . 2.) Disconnect 5 connectors. 3.) Unscrew 5 screws to remove...
Page 354
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 108 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-48-3 Removal procedure (cont’d) 4.) Perform the following functional tests. If all are successful, include the debrief script provided below. 8-2-48-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-49 Functional Tests Service ManualSection Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual,...
Page 355
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 109 8-2-49 Transbox Assy 110V/220V 8-2-49-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench 8-2-49-2 Preparations • Shut down the system and switch off the main breaker. 8-2-49-3 Removal procedure • To detached the AC Power assy, All open 4 pcs c onsole cover that are left, right, front, rear cover 1.) Remove side left cover and right side cover. Refer to the 8-2-15 Side Left Cover...
Page 356
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 110 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-49-3 Removal procedure (cont’d) 6.) Unplug 4 connectors inside of AC tr ansbox and pull out the transbox assy. 7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Figure 8-149 Pull out the Transbox assy Table 8-50 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section...
Page 357
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 111 8-2-49-4 Mounting Procedure Install the new parts in the reverse order of removal. When installing the Transbox assy, the transbox chassis should be inserted into the guide rail which is under the frame chassis. Figure 8-150 Insert the Transbox assy
Page 358
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 112 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-50 Fuse Set 8-2-50-1 Tools • Common pilIips screwdrivers • Allen/Unbraco wrench • Stubby screwdriver (Flat tip and Cross tip) 8-2-50-2 Preparations • Shut down the system and switch off the main breaker. 8-2-50-3 Removal procedure 1.) Remove the AC Transbox assy. Refer to the 8-2-49 Transbox Assy 110V/220V on page 8-109 . 2.) Replace the 2 fuses located on th e ACPC assy in the...
Page 359
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 113 8-2-51 Read and Write Serial Info 8-2-51-1 How to start serial no read/write program 1.) Plug the service dongle on the usb port on rear panel. 2.) Turn on the system. 3.) When the Start Application dialog box appears , press [F2] before it automatically starts Echoloader.) 4.) When the Maintenance Access dialog box pops up, type in the service password. Figure 8-152 Start Application...
Page 360
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 114 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-51-1 How to start serial no read/write program (cont’d) 5.) If the win XP screen appears, run the window explor er. Move the cursor to bottom of the screen. If the task bas appear, click the explorer icon. 6.) Run the “asn.exe” by double click. This execut ion file is in “C:\Arirang arget\bin” directory Figure 8-154 Windows Explorer Figure 8-155 “asn.exe”