Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 105
8-2-47-3              Removal procedure  (cont’d)
4.) Unplug the SATA cable from the back plane on the bottom of backplane.
5.) Disconnect the SATA Power Cable and SATA Cable  from HDD Assy after removing 2ea cable ties. 
And unscrew 4screws to remove the HDD Fixture bracket. Figure 8-144   Unplug the SATA Cable
Figure 8-145   Separate  HDD and HDD Fixture Bracket
Cable Tie 

Page 352

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 106 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-47-3              Removal procedure (cont’d)
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-47-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-48    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 353

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 107
8-2-48 ACPC Assy
8-2-48-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-48-2 Preparations • Shut down the system and switch off the main breaker.
8-2-48-3 Removal procedure 1.) Remove the AC Input Box Assy. Refer to the  8-2-49 Transbox Assy 110V/220V on page 8-109 .
2.) Disconnect 5 connectors.
3.) Unscrew 5 screws to remove...

Page 354

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 108 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-48-3              Removal procedure (cont’d)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-48-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-49    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 355

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 109
8-2-49 Transbox Assy 110V/220V
8-2-49-1 Tools • Common pilIips screwdrivers
• Allen/Unbraco wrench
8-2-49-2 Preparations • Shut down the system and switch off the main breaker.
8-2-49-3 Removal procedure • To detached the AC Power assy, All open 4 pcs c onsole cover that are left, right, front, rear cover
1.) Remove side left cover and right side cover. Refer to the  8-2-15 Side Left Cover...

Page 356

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 110 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-49-3              Removal procedure (cont’d)
6.) Unplug 4 connectors inside of AC tr ansbox and pull out the transbox assy.
7.) Perform the following functional tests. If all are successful, include the debrief script provided below. Figure 8-149   Pull out the Transbox assy
Table 8-50    Functional Tests
Service Manual Section
Functional Test / Diagnostic TestDebrief Script
Section...

Page 357

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 111
8-2-49-4 Mounting Procedure Install the new parts in the reverse order of removal.
When installing the Transbox assy,  the transbox chassis should be inserted into the guide rail which is 
under the frame chassis.
Figure 8-150   Insert the Transbox assy 

Page 358

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 112 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-50 Fuse Set 
8-2-50-1 Tools
• Common pilIips screwdrivers
• Allen/Unbraco wrench
• Stubby screwdriver (Flat tip and Cross tip)
8-2-50-2 Preparations • Shut down the system and switch off the main breaker.
8-2-50-3 Removal procedure 1.) Remove the AC Transbox assy. Refer to the 8-2-49 Transbox Assy 110V/220V on page 8-109 .
2.) Replace the 2 fuses located on th e ACPC assy in the...

Page 359

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 113
8-2-51 Read and Write Serial Info
8-2-51-1 How to start serial no read/write program 1.) Plug the service dongle on  the usb port on rear panel.
2.) Turn on the system. 
3.) When the Start Application dialog box appears , press [F2] before it automatically starts 
Echoloader.)
4.) When the Maintenance Access dialog box  pops up, type in the service password.
Figure 8-152   Start Application...

Page 360

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 114 Section 8-2 -  DISASSEMBLY/RE-ASSEMBLY
8-2-51-1              How to start serial no read/write program  (cont’d)
5.) If the win XP screen appears, run the window explor er. Move the cursor to bottom of the screen. If 
the task bas appear, click the explorer icon.
6.) Run the “asn.exe” by double click. This execut ion file is in “C:\Arirang	arget\bin” directory
Figure 8-154   Windows Explorer
Figure 8-155   “asn.exe” 