Page 301

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 55
8-2-18 Rear Cover 
8-2-18-1 Tools • Common pilIips screwdrivers
8-2-18-2 Preparations • Shut down the system and switch off the main breaker.
8-2-18-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the  8-2-15 Side Left Cover on page 8-51 and 8-
2-16 Side Right Cover on page 8-52 .
2.) Remove 4 rubber caps from the rear cover of the system.
3.) Unscrew 4 screws (1-4)...

Page 302

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 56 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-18-3              Removal procedure  (cont’d)
5.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-18-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-19    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 303

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 57
8-2-19 Middle Cover
8-2-19-1 Tools • Common pilIips screwdrivers
• Stubby screwdriver (Flat tip and Cross tip)
8-2-19-2 Preparations • Shut down the system and switch off the main breaker.
8-2-19-3 Removal procedure 1.) Remove the Left, and Right side covers. Refer to the  8-2-15 Side Left Cover on page 8-51 and 8-
2-16 Side Right Cover on page 8-52 . And Remove the Rear Cover. Refer to the...

Page 304

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 58 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-19-3              Removal procedure  (cont’d)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-19-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-20    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 305

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 59
8-2-20 Top Cover
8-2-20-1 Tools • Common pilIips screwdrivers
8-2-20-2 Preparations • Shut down the system and switch off the main breaker.
8-2-20-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
2.) Remove the Top Bottom Cover. Refer to the  8-2-21 Top Bottom Cover on page 8-60.
3.) Remove 2 rubber caps (1-2) from the top cover of the...

Page 306

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 60 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-21 Top Bottom Cover
8-2-21-1 Tools • Common pilIips screwdrivers
• Stubby screwdriver (Flat tip and Cross tip)
8-2-21-2 Preparations • Shut down the system and switch off the main breaker.
8-2-21-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
2.) Unscrew 7 screws (1-7) from the bottom of the top cover.
3.) Remove the Top Bottom...

Page 307

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 61
8-2-22 Pole Cover 
8-2-22-1 Tools • Common pilIips screwdrivers
8-2-22-2 Preparations • Shut down the system and switch off the main breaker.
8-2-22-3 Removal procedure 1.) Remove the Dummy Cover L/R. Refer to the  8-2-25 Dummy Cover L / R on page 8-66.
2.) Remove the Top Bottom Cover. Refer to the  8-2-21 Top Bottom Cover on page 8-60.
3.) Remove the Top Cover. Refer to the  8-2-20 Top...

Page 308

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 62 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-23 Neck Front 
8-2-23-1 Tools • Common pilIips screwdrivers
8-2-23-2 Preparations • Shut down the system and switch off the main breaker.
8-2-23-3 Removal procedure 1.) Remove 1 rubber cap.
2.) Unscrew 1 screw and remove the Neck Front. Figure 8-98   rubber cap of neck front cover
Figure 8-99   Open neck front cover 

Page 309

GE HEALTHCARE
DIRECTION 5394141, REVISION  1LOGIQ™ P5 SERVICE MANUAL 
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 63
8-2-23-3              Removal procedure  (cont’d)
3.) Perform the following functional tests. If all ar e successful, include the debrief script provided below.
8-2-23-4 Mounting Procedure Install the new parts in the reverse order of removal. Table 8-24    Functional Tests
Service ManualSection
Functional Test / Diagnostic TestDebrief Script
Section 4-3-1 Power On/Boot Up
“Service Manual,...

Page 310

GE HEALTHCARE
DIRECTION 5394141, REVISION  1 LOGIQ™ P5 SERVICE MANUAL  
8 - 64 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-24 Neck Rear
8-2-24-1 Tools • Common pilIips screwdrivers
8-2-24-2 Preparations • Shut down the system and switch off the main breaker.
8-2-24-3 Removal procedure 1.) Remove 2 rubber caps (1-2).
2.) Unscrew 2 screws and remove the Neck Rear. Figure 8-100   rubber cap of neck rear cover
Figure 8-101   Open neck rear cover
(2)
(1) 