Page 261
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 15 8-2-6-3 Removal procedure (cont’d) 6.) Unscrew 2 screws (5177684, HSH M5x20 WH T) to separate the LCD MONITOR and 1 screw (2159625, PH M4x8 W/SP) to remove the Cable bracket. NOTE: While unscrewing the hinge hexa screw from the arm, keep hold of the monitor to prevent dropping down. 7.) Unscrew 2 screws (2337572, FH M3x6 WHT) to remove the Arm cover L & R. 8.) Unscrew 2...
Page 262
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 16 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-3 Removal procedure (cont’d) 9.) Unscrew 4 screws (2159634, BH M4x10 WHT) to remove the Arm bottom cover. 10.) Unscrew 1 screw (2329677, TAP M4X16) to remove the Lock cover. 11.) Unscrew 2 screws (2159625, PH M4x8 W/SP) to remove the Lower arm guide bracket. Figure 8-24 Removing the Arm bottom cover Figure 8-25 Removing the Lock cover Figure 8-26 Removing the...
Page 263
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 17 8-2-6-3 Removal procedure (cont’d) 12.) Unscrew 2 screws (2159626, PH M4X20) to remove the Arm stopper. 13.) Pull out power cable & DVI cable. Ma ke sure flexible arm not to rotate. 14.) Unscrew 4 screws (2159625, PH M4x8 W/SP) to remove the Lower arm cable bracket and pull out power cable & DVI cable. Figure 8-27 Removing the Arm stopper Figure 8-28 Pulling out power...
Page 264
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 18 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-3 Removal procedure (cont’d) 15.) Unscrew 3 screws (2159634, BH M4x10 WHT) after removing the rubber caps to remove neck front cover, neck rear cover. 16.) Unscrew 3 screws(5327646, HSH M6X16 WHT) to separate the flexible arm neck pipe. Figure 8-30 Unscrewing 3 screws and removing the neck front cover & neck rear cover Figure 8-31 Separating the neck pipe
Page 265
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 19 8-2-6-3 Removal procedure (cont’d) 17.) Unscrew 1 screw (2159625, PH M4x8 W/SP) to remove the OP bracket. 18.) Unscrew 3 screws (2373562, FH M4x10 YEL) to separate the Flexible arm stopper (A+B). Figure 8-32 Removing OP bracket Figure 8-33 Separating the flexible arm stopper
Page 266
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 20 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-3 Removal procedure (cont’d) 19.) Pull out power cable & DVI cable and remove th e oilless bush after separating the neck pipe. Figure 8-34 Pulling out power cable & DVI cable and removing the stopper
Page 267
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 21 8-2-6-3 Removal procedure (cont’d) 20.) Perform the following functional tests. If all ar e successful, include the debrief script provided below. Table 8-7 Functional Tests Service Manual Section Functional Test / Diagnostic TestDebrief Script Section 4-3-1 Power On/Boot Up “Service Manual, Direction 5394141, Rev 1+, Section 8-2-6. Equipment passed all required tests and...
Page 268
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 22 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure First, remove LP5/LA5 Main K/ B Assy and LP5/LA5 Sub K/B Assy 1.) Change the MON PWR CONN LONG cable and remove the cable tie of the DVI cable for flexible arm. Figure 8-35 Changing the MON PWR CONN LONG cable MON PWR CONN LONG cable
Page 269
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8 - 23 8-2-6-4 Assembly procedure (cont’d) 2.) Assemble the LCD locking mechanism. a.) Put the stopper pin inside the spring and then put them inside the spring holder. Fix it with spring pin. b.) Put lock tite 290 on the M16 tap of the spring holder Assy and screw it. Figure 8-36 Assembling the spring holder Assy Figure 8-37 Screwing the spring holder Assy
Page 270
GE HEALTHCARE DIRECTION 5394141, REVISION 1 LOGIQ™ P5 SERVICE MANUAL 8 - 24 Section 8-2 - DISASSEMBLY/RE-ASSEMBLY 8-2-6-4 Assembly procedure (cont’d) b-1) Mold die of AL neck frame modified as seen in the figure below does not need a nut. c.) Insert holder grip as seen the picture below and fix it with 2set screws(2327580, M2.6X4). Figure 8-38 Screwing the spring holder Assy Figure 8-39 Assembling the holder grip